Dose Load during Angiographic Procedures

DOI: 10.4236/oalib.1107109, PP. 1-13

Subject Areas: Internal Medicine

Keywords: G-Arm Type Angiographic X-Ray System, Dose Load, Interventional Cardiology, Cardiology Hospital

Full-Text   Cite this paper   Add to My Lib

Abstract

On basis of dosimetric measurement, this article compares the radiation doses received by the main interventional cardiologist during the angiographic procedures. The dose load during three procedures is compared—LAD-stenting, LCx-stenting and RCA-stenting. An angiographic X-ray system Philips Allura Xper FD10 (with G-arm) was used. The dose obtained was measured with an X-ray-Gamma Dosimeter 27091. During each procedure, two modes of the equipment are used—radiography and fluoroscopy. The dose measurements were made for the respective projections of each angiographic procedure and for the operation modes of the equipment used during the respective procedure at three measurement points on the operator’s body—head, gonads and feet and at three position of the patient’s table—lowest, highest and zero. The results obtained from the calculations, based on the measured dose values, show maximum dose load in the procedure that uses the radiographic mode of operation for the longest time, namely RCA-stenting. The other two procedures have the same time on radiography and fluoroscopy. From them a higher dose load was obtained for the procedure LAD-stenting. Of the three procedures studied, the lowest dose load is obtained for the procedure LCx-stenting.

Cite this paper

Ivanova, N. and Ivanov, J. (2021). Dose Load during Angiographic Procedures. Open Access Library Journal, 8, e7109. doi: http://dx.doi.org/10.4236/oalib.1107109.

References

[1]  Strauss, K.J. and Seibert, A.J. Angiographic Equipment Selection and Configuration with Authors.
https://radiologykey.com/angiographic-equipment-selection-and-configuration/
[2]  Sadick, V., Reed, W., Collins, L., Sadick, N., Heard, R. and Robinson, J. (2010) Impact of Biplane versus Single-Plane Imaging on Radiation Dose, Contrast Load and Procedural Time in Coronary Angioplasty. British Journal of Radiology, 83, 379-394.
https://doi.org/10.1259/bjr/21696839
[3]  Mavrikou, I., Kottou, S., Tsapaki, V. and Neofotistou, V. (2008) High Patient Doses in Interventional Cardiology Due to Physicians’ Negligence: How Can They Be Prevented? Radiation Protection Dosimetry, 12, 67-70.
https://doi.org/10.1093/rpd/ncn005
[4]  Fornell, D. (2016) 5 Technologies to Reduce Cath Lab Radiation Exposure.
https://www.dicardiology.com/article/5-technologies-reduce-cath-lab-radiation-exposure
[5]  Ivanova, N. (2020) Angiographic X-Ray System with G-Arm. Dose Load of Personnel—Monograph, Medical University, Varna. [Иванова, Н. (2020) Ангиографска рентгенова система с G-рамо. Дозово натоварване на персонала, Медицински университет, Варна, ноември. (In Bulgarian)]
[6]  Röntgen-Gamma Dosimeter 27091, Technical Description and Operating Instructions, 2 September 2008.
http://www.step-sensor.de/media/main/rgd_27091-manual_.pdf

Full-Text


comments powered by Disqus

Contact Us

service@oalib.com

QQ:3279437679

WhatsApp +8615387084133