Performance Analysis and Cross-Calibration of PPC-40 and FC65-P Ionization Chambers in Low-Energy Electron Beams Absolute Dosimetry

DOI: 10.4236/ijmpcero.2025.143008, PP. 100-111

Keywords: LINAC, Low-Energy Electron Beams, Ionization Chambers, Cross-Calibration and Absolute Dosimetry, PDD, EBRT, PPC-40, FC65-P

Full-Text   Cite this paper   Add to My Lib

Abstract:

Algorithm of electron beam dose delivery is dissimilar in comparison to X-ray radiation therapy. Therefore, absolute dosimetry of low energy like (≤9 MeV) is fundamentally effective in electron beam radiation therapy by a linear accelerator. This effectively needs evaluation of cross calibration by PPC-40 and FC65-P ionization chambers for accurate dosimetry to assure quality of treatment with clinical accuracy. Implications of results enunciated in this paper show that salient features are documented. The FC65-P chamber exhibited a deviation of up to 1.04% at shallow depths, indicating limitations for low-energy electron beams. In contrast, the PPC-40 demonstrated superior accuracy in surface dose and shallow-depth measurements due to its parallel-plate design and larger effective volume. The importance of cross-calibration yielded a new correction factor of 1.01, allowing conversion of FC65-P dose measurements to PPC-40 values. This relationship remained stable under controlled conditions. The clinical significance of the selection of ionization chamber for low energy electrons dosimetry is absolutely required for treatment precision and patient safety in electron beam radiation therapy (EBRT).

References

[1]  Connell, P.P. and Hellman, S. (2009) Advances in Radiotherapy and Implications for the Next Century: A Historical Perspective. Cancer Research, 69, 383-392.
https://doi.org/10.1158/0008-5472.can-07-6871
[2]  Ferlay, J., Soerjomataram, I., Ervik, M., Dikshit, R., Eser, S., Mathers, C., et al. (2013) GLO-BOCAN 2012: Estimated Cancer Incidence, Mortality, and Prevalence Worldwide in 2012 v1.0: IARC Cancer Base No. 11. IARC Publications.
[3]  Ding, G.X. (2004) An Investigation of Accelerator Head Scatter and Output Factor in Air. Medical Physics, 31, 2527-2533.
https://doi.org/10.1118/1.1784131
[4]  Baskar, R., Lee, K.A., Yeo, R. and Yeoh, K. (2012) Cancer and Radiation Therapy: Current Advances and Future Directions. International Journal of Medical Sciences, 9, 193-199.
https://doi.org/10.7150/ijms.3635
[5]  Hogstrom, K.R. and Almond, P.R. (2006) Review of Electron Beam Therapy Physics. Physics in Medicine and Biology, 51, R455-R489.
https://doi.org/10.1088/0031-9155/51/13/r25
[6]  Su, Z., Indelicato, D.J., Mailhot, R.B. and Bradley, J.A. (2020) Impact of Different Treatment Techniques for Pediatric Ewing Sarcoma of the Chest Wall: IMRT, 3DCPT, and IMPT with/without Beam Aperture. Journal of Applied Clinical Medical Physics, 21, 100-107.
https://doi.org/10.1002/acm2.12870
[7]  Rahman, M., Shamsuzzaman, M., Sarker, M., Jobber, A., Mia, M., Bairagi, A.K., et al. (2021) Dosimetric Characterization of Medical Linear Accelerator Photon and Electron Beams for the Treatment Accuracy of Cancer Patients. World Journal of Advanced Engineering Technology and Sciences, 3, 41-59.
https://doi.org/10.30574/wjaets.2021.3.1.0046
[8]  IAEA International Atomic Energy Agency (1997) The Use of Plane-Parallel Ionization Chambers in High-Energy Electron and Photon Beams: An International Code of Practice for Dosimetry. Technical Report Series No. 381, IAEA.
[9]  International Atomic Energy Agency (2024) Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water, Technical Reports Series No. 398 rev1. IAEA.
[10]  Burns, D.T., Ding, G.X. and Rogers, D.W.O. (1996) R50 as a Beam Quality Specifier for Selecting Stopping‐Power Ratios and Reference Depths for Electron Dosimetry. Medical Physics, 23, 383-388.
https://doi.org/10.1118/1.597893
[11]  Boag, J.W. and Currant, J. (1980) Current Collection and Ionic Recombination in Small Cylindrical Ionization Chambers Exposed to Pulsed Radiation. The British Journal of Radiology, 53, 471-478.
https://doi.org/10.1259/0007-1285-53-629-471
[12]  Derikum, K. and Roos, M. (1993) Measurement of Saturation Correction Factors of Thimble-Type Ionization Chambers in Pulsed Photon Beams. Physics in Medicine and Biology, 38, 755-763.
https://doi.org/10.1088/0031-9155/38/6/009
[13]  Chowdhury, M.R.I., Rabby, M.F., Ahmed, R., Akter, M. and Rahman, M.M. (2024) A Comparative Study of Photon Beam Percentage Depth Dose of a Recently Installed Varian Vitalbeam Linear Accelerator at TMSS Cancer Center, Bangladesh. International Journal of Radiology & Radiation Therapy, 11, 100-104.
https://doi.org/10.15406/ijrrt.2024.11.00394