Three Years’ Daily Off-Iso Winston-Lutz Test Experience

DOI: 10.4236/ijmpcero.2026.153007, PP. 76-87

Keywords: SRS/SBRT/SIMT, Off-Iso Winston-Lutz Test

Full-Text   Cite this paper   Add to My Lib

Abstract:

Background: Stereotactic radiosurgery (SRS) is one of the main approaches to managing brain metastases. Single iso multi-target (SIMT) treatment emerges as a new branch of SRS. Standard daily clinic quality assurance of SIMT is in the need to assure the treatment with high precision. Purpose: This study reports a three years’ longitudinal experience of performing daily (the day of SRS/SBRT/SIMT treatment) Off-Isocenter Winston-Lutz (OWL) test on one VitalBeam linear accelerator (Varian Medical Systems, Palo Alto, CA). A practical workflow and detailed radiation-mechanical spatial discrepancy information associated with six ball bearings (BBs) inside MMWL phantom (Sun Nuclear, Melbourne, FL) are reported for the first time in routine clinical practice. Method: The daily Off-iso Winston-Lutz (OWL) test was performed on one VitalBeam Linac which is equipped with a 120-leaf Millennium Multileaf Collimator (MLC) and a standard IGRT couch. Imaging capabilities include onboard kV and MV system that supports both 2D radiographic and 3D CBCT acquisitions. The MultiMet phantom and QA software (v2.1) are from Sun Nuclear Inc. The “Excluding 45? Couch Kicks” 10-beam plan was adopted in Aria 16.1 R&V and Eclipse system (v16.1). The plan utilizes eight distinct Gantry-Couch-Collimator (GCC) angle configurations. An institutional workflow of off-iso Winston-Lutz test was established and implemented to ensure high-precision measurements with minimal impact on the daily treatment schedule. Results: The three years (May 2023-May 2026) OWL test were conducted on eight gantry, couch and collimator (GCC) angle sets. A total of 81 valid and formatted reports were collected and analyzed. Initial phantom positioning offset and final radiation-mechanical field center offset are reported. From GCC (0, 0, 0), GCC (0, 90, 0) and GCC (0, 0, 270) three years’ trend, it clearly showed that the further the target is from iso center, the larger radiation-mechanical spatial discrepancy is. From the GCC (90, 0, 90), GCC (180, 0, 90) and GCC (270, 0, 90) three years’ trend, it suggests us the gantry sagging as a contribution factor to affect the discrepancy in this VitalBeam linear accelerator. Conclusions: Three years of daily implementation off-iso Winston-Lutz test experience confirm that the OWL test is a feasible and highly valuable routine QA practice. The rich spatial discrepancy information provides actionable insights for the plan designer to optimize the beam placement and refine isocenter selection for

References

[1]  Bou Dargham, T., Winter, S.F., Batich, K., Gehring, K., Parsons, M.W., Dietrich, J., et al. (2025) Neurocognitive Outcomes in Patients with Brain Metastases: A Systematic Review. The Lancet Oncology, 26, e671-e682.
https://doi.org/10.1016/s1470-2045(25)00525-x
[2]  NRG-CC009: Testing Two Different Radiation Therapy Types to See Which Is Better at Preventing Loss of Memory and Changes in Thinking Ability for People with SCLC That Has Spread to the Brain.
https://www.nrgoncology.org/patients/patient-study-webpages/nrg-cc009-sclc-spread-to-the-brain/
[3]  Sheehan, J.P., Yen, C.P., Lee, C.C. and Loeffler, J.S. (2014) Cranial Stereotactic Radiosurgery: Current Status of the Initial Paradigm Shifter. Journal of Clinical Oncology, 32, 2836-2846.
https://doi.org/10.1200/jco.2013.53.7365
[4]  Meeks, S.L., Mercado, C.E., Popple, R.A., Agazaryan, N., Kaprealian, T., Fiveash, J.B., et al. (2022) Practical Considerations for Single Isocenter LINAC Radiosurgery of Multiple Brain Metastases. Practical Radiation Oncology, 12, 195-199.
https://doi.org/10.1016/j.prro.2021.09.007
[5]  Gao, J.F. and Liu, X.Q. (2016) Off-Isocenter Winston-Lutz Test for Stereotactic Radiosurgery/Stereotactic Body Radiotherapy. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 5, 154-161.
https://doi.org/10.4236/ijmpcero.2016.52017
[6]  Poder, J., Brown, R., Porter, H., Gupta, R. and Ralston, A. (2018) Development of a Dedicated Phantom for Multi-Target Single-Isocentre Stereotactic Radiosurgery End to End Testing. Journal of Applied Clinical Medical Physics, 19, 99-108.
https://doi.org/10.1002/acm2.12452
[7]  Yaqub, M.M. (2018) Design & Delivery of Automated Winston-Lutz Test for ISO-Centric & Off-Axis Delivery Stability Utilizing Truebeam Developer Mode & Electronic Portal Imaging Device. Master’s Thesis, University of Nevada, Las Vegas, 3349 p.
https://doi.org/10.34917/13568805
[8]  Kudrolli, H., Murray, A., Tirpak, L., Matin, A. and Zack, J. (2020) PO-1320: A Machine QA Tool to Verify Targeting Accuracy of Off-Isocenter Metastases. Radiotherapy and Oncology, 152, S696-S697.
https://doi.org/10.1016/s0167-8140(21)01338-4
[9]  Grafil, E.M., De Jean, P., Capaldi, D.P.I., Skinner, L.B., Xing, L. and Yu, A.S. (2024) Evaluation of Artificial Intelligence and Optical Image Recognition Techniques Used in Oneiso, an Off-Axis Winston-Lutz Quality Assurance Phantom. Biomedical Physics & Engineering Express, 11, Article ID: 015056.
https://doi.org/10.1088/2057-1976/ada037
[10]  Gao, J.F. and Anand, D. (2024) Off-ISO Winston-Lutz Test on Seven Linear Accelerators. Journal of Applied Clinical Medical Physics, 25, e14470.
https://doi.org/10.1002/acm2.14470
[11]  Ono, T., Kido, T., Nakamura, M., Iramina, H., Kakino, R. and Mizowaki, T. (2022) Automatic Measurement of Beam-Positioning Accuracy at Off-Isocenter Positions. Journal of Applied Clinical Medical Physics, 24, e13844.
https://doi.org/10.1002/acm2.13844
[12]  Poder, J., Chambers, R., Porter, H., Brown, R., Koprivec, D., Yuen, J., et al. (2025) Quality Assurance of Single Isocentre Multiple Target Stereotactic Radiosurgery: Findings from Long-Term Off-Axis Winston-Lutz Testing and Machine Performance Checks. Journal of Applied Clinical Medical Physics, 26, e70275.
https://doi.org/10.1002/acm2.70275
[13]  Sahgal, A., Ruschin, M., Ma, L., Verbakel, W., Larson, D. and Brown, P.D. (2017) Stereotactic Radiosurgery Alone for Multiple Brain Metastases? A Review of Clinical and Technical Issues. Neuro-Oncology, 19, ii2-ii15.
https://doi.org/10.1093/neuonc/nox001
[14]  Du, W.L., Gao, S., Wang, X. and Kudchadker, R.J. (2012) Quantifying the Gantry Sag on Linear Accelerators and Introducing an MLC-Based Compensation Strategy. Medical Physics, 39, 2156-2162.
https://doi.org/10.1118/1.3697528

Full-Text

Contact Us

service@oalib.com

QQ:3279437679

WhatsApp +8615387084133