Stripping of a Newly Deployed Coronary Stent during Manual Thrombectomy: A Cautionary Case

DOI: 10.4236/crcm.2026.155026, PP. 195-200

Keywords: Percutaneous Coronary Intervention, Aspiration Thrombectomy, Stent Stripping

Full-Text   Cite this paper   Add to My Lib

Abstract:

A 54-year-old male with hypertension and hyperlipidemia presented with an anterior ST-elevation myocardial infarction (STEMI) due to a mid-left anterior descending (LAD) artery occlusion. Following the deployment of a drug-eluting stent (DES), persistent distal thrombus prompted manual aspiration thrombectomy. During withdrawal of the Pronto? V4 catheter, the device entangled and stripped the newly deployed stent, pulling it into the guide catheter. The complication was managed by retrieving the deformed stent and successfully placing two rescue stents. This case underscores the risks of aspiration thrombectomy through fresh stents which carries a significant risk of mechanical stripping. Extreme caution during advancement and withdrawal of these catheters can mitigate the risk.

References

[1]  Williams, P.D., Mamas, M.A., Morgan, K.P., El-Omar, M., Clarke, B., Bainbridge, A., et al. (2012) Longitudinal Stent Deformation: A Retrospective Analysis of Frequency and Mechanisms. EuroIntervention, 8, 267-274.
https://doi.org/10.4244/eijv8i2a41
[2]  S?awek-Szmyt, S., Araszkiewicz, A., Grygier, M. and Lesiak, M. (2025) Continuous Mechanical Aspiration Thrombectomy for STEMI with High Thrombus Burden. Cardiology Journal, 32, 407-410.
https://doi.org/10.5603/cj.102528
[3]  Vemula, P., Kalavakunta, J.K., Abela, G.S. and Karve, M. (2014) A Rare and Serious Unforeseen Complication of Cutting Balloon Angioplasty. Case Reports in Cardiology, 2014, Article ID: 246784.
https://doi.org/10.1155/2014/246784
[4]  Baz, J.A., Jiménez, V.A., Sepúlveda, J., Chantada, D., Díaz, C.M., Juan-Salvadores, P., et al. (2015) Unintentional Extraction of an Endothelized Coronary Stent with an Aspiration Catheter during Primary Percutaneous Coronary Intervention. Circulation: Cardiovascular Interventions, 8, e002114.
https://doi.org/10.1161/circinterventions.114.002114
[5]  Yilmaztepe, M., Gurlertop, H.Y., Kacmaz, Y. and Aksoy, Y. (2016) Inadvertent Extraction of a Previously Deployed Stent. Journal of Cardiology Cases, 14, 38-39.
https://doi.org/10.1016/j.jccase.2016.03.016
[6]  Seo, K., Yang, H., Lim, H. and Yoon, M. (2022) Stent Dislodgement Force of Drug-Eluting Coronary Stents: A Bench Test. Cardiovascular Diagnosis and Therapy, 12, 370-377.
https://doi.org/10.21037/cdt-22-49
[7]  Chichareon, P., Katagiri, Y., Asano, T., Takahashi, K., Kogame, N., Modolo, R., et al. (2019) Mechanical Properties and Performances of Contemporary Drug-Eluting Stent: Focus on the Metallic Backbone. Expert Review of Medical Devices, 16, 211-228.
https://doi.org/10.1080/17434440.2019.1573142
[8]  Jolly, S.S., Cairns, J.A., Yusuf, S., Meeks, B., Pogue, J., Rokoss, M.J., et al. (2015) Randomized Trial of Primary PCI with or without Routine Manual Thrombectomy. New England Journal of Medicine, 372, 1389-1398.
https://doi.org/10.1056/nejmoa1415098