Calibration of HDR 192Ir Source Air-Kerma Strength During Time Linearity Assessments

DOI: 10.4236/ijmpcero.2026.151002, PP. 14-27

Keywords: Brachytherapy, HDR Source, Calibration, Determination, Air-Kerma Strength, Well Chamber, Sweet Spot

Full-Text   Cite this paper   Add to My Lib

Abstract:

A common method of a new HDR source calibration is to measure a maximum ionization current (Imax) using a well ionization chamber, in which the 192Ir source travels along the chamber axis in the spatial domain. The source’s air-kerma strength (SK) can be determined from the measured Imax. In this work, an approach for routine calibration of the air-kerma strength of an HDR 192Ir source in the temporal domain using integrated charge measurements with a well ionization chamber is presented. The charge Q (t) is measured as a function of time (t) for different dwell times at a sweet spot position in the well ionization chamber. The air-kerma strength is obtained from the slope (dQ/dt) of charge linearity measurement derived from the relationship between total ionization charge and time. This paper compares this method with the commonly used calibration approach based on measuring the maximum ionization current Imax in the well chamber. The agreement between these two methods of calibrating the 192Ir source over the past several years is good and differences are less than 1%, the key finding of this study. The effect of the sweet spot range in the chamber is measured and the transit time is analyzed. New dQ/dt data over the past few years are presented. This paper provides detailed data on the sweet spot regions, transit time, and sensitivities for two types of well-type ionization chambers, and discusses the application in HDR 192Ir source exchanges. This dQ/dt method provides additional support to 192Ir sources calibration and verification.

References

[1]  Nath, R., Anderson, L., Jones, D., Ling, C., et al. (1987) Specification of Brachytherapy Source Strength. American Association of Physicists in Medicine Report No. 21. AAPM Task Group.
[2]  Kutcher, G.J., Coia, L., Gillin, M., Hanson, W.F., Leibel, S., Morton, R.J., et al. (1994) Comprehensive QA for Radiation Oncology: Report of AAPM Radiation Therapy Committee Task Group 40. Medical Physics, 21, 581-618.
https://doi.org/10.1118/1.597316
[3]  International Atomic Energy Agency (IAEA) (1999) Calibration of Brachytherapy Sources—Guidelines on Standardized Procedures for the Calibration of Brachytherapy Sources at Secondary Standard Dosimetry Laboratories (SSDLs) and Hospitals. IAEA-TECDOC-1079.
[4]  International Atomic Energy Agency (IAEA) (2002) Calibration of Photon and Beta ray Sources Used in Brachytherapy—Guidelines on Standardized Procedures at Secondary Standards Dosimetry Laboratories (SSDLs) and Hospitals. IAEA-TECDOC-1274.
[5]  Nath, R., Anderson, L.L., Meli, J.A., Olch, A.J., Stitt, J.A. and Williamson, J.F. (1997) Code of Practice for Brachytherapy Physics: Report of the AAPM Radiation Therapy Committee Task Group No. 56. Medical Physics, 24, 1557-1598.
https://doi.org/10.1118/1.597966
[6]  International Atomic Energy Agency (IAEA) (2007) Appendix to IAEA Calibration Certificate: Well Type Chamber and Electrometer Calibration Procedure at the IAEA Dosimetry Laboratory. IAEA.
[7]  Podgorsak, M.B., DeWerd, L.A. and Paliwal, B.R. (1993) The Half‐Life of High Dose Rate Ir‐192 Sources. Medical Physics, 20, 1257-1259.
https://doi.org/10.1118/1.596976
[8]  Sander, T. and Nutbrown, R.F. (2006) The NPL Air-Kerma Primary Standard TH100C for High Dose Rate 192Ir Brachytherapy Sources, NPL Report DQL-RD 004: National Physics Laboratory (NPL).
[9]  Bidmead, A.M., Sander, T., Locks, S.M., Lee, C.D., Aird, E.G.A., Nutbrown, R.F., et al. (2010) The IPEM Code of Practice Based on the NPL Air-Kerma Standard. Physics in Medicine and Biology, 55, 3145-3159.
https://doi.org/10.1088/0031-9155/55/11/011
[10]  Awunor, O.A., Lecomber, A.R., Richmond, N. and Walker, C. (2011) A Practical Implementation of the 2010 IPEM High Dose Rate Brachytherapy Code of Practice for the Calibration of 192Ir Sources. Physics in Medicine and Biology, 56, 5397-5410.
https://doi.org/10.1088/0031-9155/56/16/020
[11]  Goetsch, S.J., Attix, F.H., Pearson, D.W. and Thomadsen, B.R. (1991) Calibration of 192Ir High-Dose-Rate Afterloading Systems. Medical Physics, 18, 462-467.
https://doi.org/10.1118/1.596649
[12]  Goetsch, S.J., Attix, F.H., Dewerd, L.A. and Thomadsen, B.R. (1992) A New Re-Entrant Ionization Chamber for the Calibration of Iridium-192 High-Dose-Rate Sources. International Journal of Radiation Oncology, Biology, Physics, 24, 167-170.
https://doi.org/10.1016/0360-3016(92)91037-n
[13]  Dewerd, L.A. and Thomadsen, B.R. (1995) Source Strength Standards and Calibration of HDR/PDR Sources. In: Williamson, J., Thomadsen, B.R. and Nath, R., Brachytherapy Physics, Med Phys Publishing, 541-555.
[14]  Rasmussen, B.E., Davis, S.D., Schmidt, C.R., Micka, J.A. and DeWerd, L.A. (2011) Comparison of Air-Kerma Strength Determinations for HDR 192Ir Sources. Medical Physics, 38, 6721-6729.
https://doi.org/10.1118/1.3656683
[15]  Podgorsak, M.B., Dewerd, L.A., Thomadsen, B.R. and Paliwal, B.R. (1992) Thermal and Scatter Effects on the Radiation Sensitivity of Well Chambers Used for High Dose Rate Ir192 Calibrations. Medical Physics, 19, 1311-1314.
https://doi.org/10.1118/1.596764
[16]  DeWerd, L.A., Ibbott, G.S., Meigooni, A.S., Mitch, M.G., Rivard, M.J., Stump, K.E., et al. (2011) A Dosimetric Uncertainty Analysis for Photon-Emitting Brachytherapy Sources: Report of AAPM Task Group No. 138 and GEC-ESTRO. Medical Physics, 38, 782-801.
https://doi.org/10.1118/1.3533720
[17]  Griffin, S.L., DeWerd, L.A., Micka, J.A. and Bohm, T.D. (2005) The Effect of Ambient Pressure on Well Chamber Response: Experimental Results with Empirical Correction Factors. Medical Physics, 32, 700-709.
https://doi.org/10.1118/1.1861157
[18]  Bjerke, H., Dewerd, L.A., Seuntjens, J.P. and Bidmead, M. (2013) Source Calibration. In: Venselaar, J., Baltas, D., Meigooni, A. and Hoskin, P., Comprehensive Brachytherapy: Physical and Clinical Aspects, CRC/Taylor & Francis Group, 61-74.
[19]  International Atomic Energy Agency (IAEA) (2023) IAEA TRS 492; Dosimetry in Brachytherapy—An International Code of Practice for Secondary Standards Dosimetry Laboratories and Hospitals. IAEA.
[20]  Kubo, H.D., Glasgow, G.P., Pethel, T.D., Thomadsen, B.R. and Williamson, J.F. (1998) High Dose-Rate Brachytherapy Treatment Delivery: Report of the AAPM Radiation Therapy Committee Group No. 59. Medical Physics, 25, 375-403.
https://doi.org/10.1118/1.598232
[21]  Glasgow, G.P. (1999) Brachytherapy. In: Van Dyk, J., The Modern Technology of Radiation Oncology: A Compendium for Medical Physicists and Radiation Oncologists, Medical Physics Publishing, 695-752.
[22]  Butler, W.M., Bice, W.S., DeWerd, L.A., Hevezi, J.M., Huq, M.S., Ibbott, G.S., et al. (2008) Third-Party Brachytherapy Source Calibrations and Physicist Responsibilities: Report of the AAPM Low Energy Brachytherapy Source Calibration Working Group. Medical Physics, 35, 3860-3865.
https://doi.org/10.1118/1.2959723
[23]  Hayes, E. (2012) Introduction of IPEM Code of Practice for Determination of Reference Air Kerma Rate for HDR 192Ir Brachytherapy Sources Based on the NPL Air Kerma Standard. Physica Medica, 28, 336.
https://doi.org/10.1016/j.ejmp.2012.06.019
[24]  Venselaar, J. and Perez-Calatayud, J. (2004) European Guidelines for Quality Assurance in Radiotherapy, ESTRO Booklet No. 8, a Practical Guide to Quality Control of Brachytherapy Equipment. ESTRO.
[25]  Standard Imaging Inc (2007) Specifications of IVB 1000 Well Type Chamber 1300-20, 9/07. REF 90009.
[26]  Standard Imaging Inc (2007) Specifications of HDR 1000 Plus Well Type Chamber. REF 90008.
[27]  Standard Imaging Inc (2022) HDR 1000 Plus Well Chamber User Manual, REF 90008. DOC #80026-25.

Full-Text

Contact Us

service@oalib.com

QQ:3279437679

WhatsApp +8615387084133