Performance Evaluation of Ionization Chambers and Microdiamond Detectors in Small Field Penumbra Measurements

DOI: 10.4236/ijmpcero.2025.143007, PP. 89-99

Keywords: Detectors, Small-Field, Beam-Profile, Penumbra

Full-Text   Cite this paper   Add to My Lib

Abstract:

In small-field radiotherapy, accurate characterization of the penumbra is critical for precise dose delivery, particularly in treatments like stereotactic radiosurgery (SRS), where steep dose gradients are required to spare surrounding healthy tissues. The objective of this study was to quantify the penumbra region in small radiation fields to improve dose delivery accuracy and minimize unnecessary exposure. Three detectors—the Semiflex 3D ion chamber, PinPoint 3D ion chamber, and microDiamond detector—were used to measure beam profiles for field sizes ranging from 1 × 1 cm2 to 4 × 4 cm2 using 6 MV photon energy. These measured profiles were compared against data from the Monaco treatment planning system (TPS) to evaluate consistency and detector performance. For larger field sizes (4 × 4 cm2 and 3 × 3 cm2), the Semiflex 3D detector showed the smallest deviations from TPS data, with errors of 14.5% and 14.3%, respectively. The PinPoint 3D and microDiamond detectors exhibited larger deviations, with microDiamond showing the greatest discrepancies (30.7% and 34.5%). In the 2 × 2 cm2 field, Semiflex maintained a deviation of 14.9%, while PinPoint 3D and microDiamond showed errors of 28.1% and 38.4%. For the smallest field (1 × 1 cm2), all detectors underestimated penumbra width: 11.3% (Semiflex 3D), 27.1% (PinPoint 3D), and 32.7% (microDiamond). Despite strong overall correlations between measured and planned profiles, noticeable deviations were observed, particularly in smaller fields. The Semiflex 3D, with the largest collecting volume, showed the best agreement with TPS data across all field sizes. This study highlights the importance of detector selection in small-field dosimetry and underscores the need for accurate penumbra quantification to ensure optimal treatment outcomes.

References

[1]  Malicki, J. (2012) The Importance of Accurate Treatment Planning, Delivery, and Dose Verification. Reports of Practical Oncology & Radiotherapy, 17, 63-65.
https://doi.org/10.1016/j.rpor.2012.02.001
[2]  Cmrecak, F., Andrasek, I., Mekic, M.S., Ravlic, M. and Beketic-Oreskovic, L. (2019) Modern Radiotherapy Techniques. Croatian Journal of Oncology, 47, 2-3.
[3]  Friedman, W.A. and Bova, F.J. (1989) The University of Florida Radiosurgery System. Surgical Neurology, 32, 334-342.
https://doi.org/10.1016/0090-3019(89)90135-3
[4]  Podgorsak, E.B., Bruce Pike, G., Pla, M., Olivier, A. and Souhami, L. (1990) Radiosurgery with Photon Beams: Physical Aspects and Adequacy of Linear Accelerators. Radiotherapy and Oncology, 17, 349-358.
https://doi.org/10.1016/0167-8140(90)90008-k
[5]  Wilcox, E.E. and Daskalov, G.M. (2008) Accuracy of Dose Measurements and Calculations within and Beyond Heterogeneous Tissues for Photon Fields Smaller than Produced by Cyberknife. Medical Physics, 35, 2259-2266.
https://doi.org/10.1118/1.2912179
[6]  Sharpe, M.B., Jaffray, D.A., Battista, J.J. and Munro, P. (1995) Extrafocal Radiation: A Unified Approach to the Prediction of Beam Penumbra and Output Factors for Megavoltage X-Ray Beams. Medical Physics, 22, 2065-2074.
https://doi.org/10.1118/1.597648
[7]  Bova, F.J., Buatti, J.M., Friedman, W.A., Mendenhall, W.M., Yang, C. and Liu, C. (1997) The University of Florida Frameless High-Precision Stereotactic Radiotherapy System. International Journal of Radiation Oncology, Biology, Physics, 38, 875-882.
https://doi.org/10.1016/s0360-3016(97)00055-2
[8]  Schulte, R.W., Fargo, R.A., Meinass, H.J., Slater, J.D. and Slater, J.M. (2000) Analysis of Head Motion Prior to and during Proton Beam Therapy. International Journal of Radiation Oncology, Biology, Physics, 47, 1105-1110.
https://doi.org/10.1016/s0360-3016(00)00551-4
[9]  Hunt, M.A., Desobry, G.E., Fowble, B. and Coia, L.R. (1997) Effect of Low-Density Lateral Interfaces on Soft-Tissue Doses. International Journal of Radiation Oncology, Biology, Physics, 37, 475-482.
https://doi.org/10.1016/s0360-3016(96)00499-3
[10]  Sharpe, M.B., Miller, B.M. and Wong, J.W. (2000) Compensation of X-Ray Beam Penumbra in Conformal Radiotherapy. Medical Physics, 27, 1739-1745.
https://doi.org/10.1118/1.1287283
[11]  O’Malley, L., Pignol, J., Beachey, D.J., Keller, B.M., Presutti, J. and Sharpe, M. (2006) Improvement of Radiological Penumbra Using Intermediate Energy Photons (IEP) for Stereotactic Radiosurgery. Physics in Medicine and Biology, 51, 2537-2548.
https://doi.org/10.1088/0031-9155/51/10/012
[12]  PTW (2017) Semiflex 3D Ion Chamber 31021.
https://www.ptwdosimetry.com/en/products/semiflex-3d-ion-chamber-31021
[13]  PTW (2018) PinPoint 3D Ion Chamber.
https://www.ptwdosimetry.com/en/products/pinpoint-3d-ion-chamber
[14]  PTW (2013) microDiamond.
https://www.ptwdosimetry.com/en/products/microdiamond#c781
[15]  Hidaytalla, L.A. (1994) Development of Isodose Curves for a 6MV X-Ray Beam. Master’s Thesis, University of Khartoum.

Full-Text

Contact Us

service@oalib.com

QQ:3279437679

WhatsApp +8615387084133