Patient-Specific QA Using Multidimensional Detectors in Single Isocenter Multitarget Stereotactic Radiotherapy for Multiple Brain Metastasis

DOI: 10.4236/ijmpcero.2025.142002, PP. 15-32

Keywords: Single Isocenter Multitarget, Dynamic Conformal Arc, Brain Metastasis, Patient-Specific Quality Assurance

Full-Text   Cite this paper   Add to My Lib

Abstract:

Purpose: This study used ArcCHECK (AC) to investigate the connection between target- and treatment plan-dependent parameters and the gamma passing rate (GPR) to determine the characteristics of patient-specific quality assurance (PSQA) in single isocenter multitarget (SIMT) stereotactic radiotherapy (SRT). Methods: Twenty-four SIMT SRT treatment plans (92 targets) were evaluated. All treatment plans were developed using dynamic conformal arc therapy (DCAT), and the equivalent multi-leaf collimator (MLC) field size, distance from the isocenter (IC), and monitor/Gy were determined. The absolute dose error from the treatment planning system (TPS) was calculated using 92 targets and an ionization chamber detector. Gamma analysis was performed with AC, a multidimensional detector, and SNC patient software. Threshold values of 10% and 20%, absolute and relative dosimetry, global mode, and dose difference (DD)/distance-to-agreement (DTA) criteria of 3%/2 mm, 2%/2 mm, 2%/1 mm, and 1%/1 mm were employed. Differences in GPR were assessed for each condition. The correlation of GPR with treatment planning parameters and ionization chamber detector errors was investigated. Results: With stricter DD (2%/2 - 2%/1 mm), the GPR fell by an average of 18.6% (86.4% - 67.8%). The average equivalent MLC field size was moderately correlated with GPR and weakly correlated with the maximum error of the ionization chamber detector and MU/Gy, but not with the max target-IC distance. In the case of small targets, the dose in the center was relatively acceptable, but the lower dose range outside of the target was less responsive, frequently resulting in failures. Conclusions: In SIMT SRT, we investigated in detail the differences in GPR based on the AC’s gamma analysis conditions, as well as the relationship between the GPR and treatment planning parameters. The GPR was found to be significantly reduced when the average equivalent MLC field size was small, i.e., there were many small target sizes.

References

[1]  Brown, P.D., Jaeckle, K., Ballman, K.V., Farace, E., Cerhan, J.H., Anderson, S.K., et al. (2016) Effect of Radiosurgery Alone vs Radiosurgery with Whole Brain Radiation Therapy on Cognitive Function in Patients with 1 to 3 Brain Metastases: A Randomized Clinical Trial. JAMA, 316, 401-409.
https://doi.org/10.1001/jama.2016.9839
[2]  Palmer, J.D., Sebastian, N.T., Chu, J., DiCostanzo, D., Bell, E.H., Grecula, J., et al. (2020) Single-Isocenter Multitarget Stereotactic Radiosurgery Is Safe and Effective in the Treatment of Multiple Brain Metastases. Advances in Radiation Oncology, 5, 70-76.
https://doi.org/10.1016/j.adro.2019.08.013
[3]  Yamamoto, M., Serizawa, T., Higuchi, Y., Sato, Y., Kawagishi, J., Yamanaka, K., et al. (2017) A Multi-Institutional Prospective Observational Study of Stereotactic Radiosurgery for Patients with Multiple Brain Metastases (JLGK0901 Study Update): Irradiation-Related Complications and Long-Term Maintenance of Mini-Mental State Examination Scores. International Journal of Radiation OncologyBiologyPhysics, 99, 31-40.
https://doi.org/10.1016/j.ijrobp.2017.04.037
[4]  Clark, G.M., Popple, R.A., Young, P.E. and Fiveash, J.B. (2010) Feasibility of Single-Isocenter Volumetric Modulated Arc Radiosurgery for Treatment of Multiple Brain Metastases. International Journal of Radiation OncologyBiologyPhysics, 76, 296-302.
https://doi.org/10.1016/j.ijrobp.2009.05.029
[5]  Roper, J., Chanyavanich, V., Betzel, G., Switchenko, J. and Dhabaan, A. (2015) Single-Isocenter Multiple-Target Stereotactic Radiosurgery: Risk of Compromised Coverage. International Journal of Radiation OncologyBiologyPhysics, 93, 540-546.
https://doi.org/10.1016/j.ijrobp.2015.07.2262
[6]  Stanhope, C., Chang, Z., Wang, Z., Yin, F., Kim, G., Salama, J.K., et al. (2016) Physics Considerations for Single-Isocenter, Volumetric Modulated Arc Radiosurgery for Treatment of Multiple Intracranial Targets. Practical Radiation Oncology, 6, 207-213.
https://doi.org/10.1016/j.prro.2015.10.010
[7]  Feygelman, V., Walker, L., Chinnaiyan, P. and Forster, K. (2008) Simulation of Intrafraction Motion and Overall Geometric Accuracy of a Frameless Intracranial Radiosurgery Process. Journal of Applied Clinical Medical Physics, 9, 68-86.
https://doi.org/10.1120/jacmp.v9i4.2828
[8]  Saglam, Y., Bolukbasi, Y., Atasoy, A.I., Karakose, F., Alpan, V., Selek, U., et al. (2020) Quality Assurance in Stereotactic Radiosurgery and Stereotactic Body Radiotherapy. Advances in Research, 21, 22-33.
https://doi.org/10.9734/air/2020/v21i730216
[9]  Benedict, S.H., Cai, J., Libby, B., Lovelock, M., Schlesinger, D., Sheng, K., et al. (2010) SRT and SBRT: Current Practices for QA Dosimetry and 3D. Journal of Physics: Conference Series, 250, Article 012057.
https://doi.org/10.1088/1742-6596/250/1/012057
[10]  Halvorsen, P.H., Cirino, E., Das, I.J., Garrett, J.A., Yang, J., Yin, F., et al. (2017) AAPM‐RSS Medical Physics Practice Guideline 9.a. for SRS‐SBRT. Journal of Applied Clinical Medical Physics, 18, 10-21.
https://doi.org/10.1002/acm2.12146
[11]  Dunn, L., Tamborriello, A., Subramanian, B., Xu, X. and Ruruku, T.T. (2024) Assessing the Sensitivity and Suitability of a Range of Detectors for SIMT PSQA. Journal of Applied Clinical Medical Physics, 25, e14343.
https://doi.org/10.1002/acm2.14343
[12]  Farhood, B., Parwaie, W., Refahi, S. and Ardekani, M. (2018) Different Dosimeters/Detectors Used in Small-Field Dosimetry: Pros and Cons. Journal of Medical Signals & Sensors, 8, 195-203.
https://doi.org/10.4103/jmss.jmss_3_18
[13]  Pappas, E., Maris, T.G., Zacharopoulou, F., Papadakis, A., Manolopoulos, S., Green, S., et al. (2008) Small SRS Photon Field Profile Dosimetry Performed Using a Pinpoint Air Ion Chamber, a Diamond Detector, a Novel Silicon‐Diode Array (DOSI), and Polymer Gel Dosimetry. Analysis and Intercomparison. Medical Physics, 35, 4640-4648.
https://doi.org/10.1118/1.2977829
[14]  Ezzell, G.A. (2017) The Spatial Accuracy of Two Frameless, Linear Accelerator‐Based Systems for Single‐Isocenter, Multitarget Cranial Radiosurgery. Journal of Applied Clinical Medical Physics, 18, 37-43.
https://doi.org/10.1002/acm2.12044
[15]  Cullom, E.T., Xia, Y., Chuang, K., Gude, Z.W., Zlateva, Y., Adamson, J.D., et al. (2021) Single Isocenter SRS Using CAVMAT Offers Improved Robustness to Commissioning and Treatment Delivery Uncertainty Compared to Vmat. Journal of Applied Clinical Medical Physics, 22, 36-43.
https://doi.org/10.1002/acm2.13248
[16]  Goodall, S.K., Dunn, L., Dunning, J., Muñoz, L., Rowshanfarzad, P. and Ebert, M.A. (2022) Matched Linac Stereotactic Radiotherapy: An Assessment of Delivery Similarity and Distributive Patient‐Specific Quality Assurance Feasibility. Journal of Applied Clinical Medical Physics, 23, e13652.
https://doi.org/10.1002/acm2.13652
[17]  Salari, E., Parsai, E.I., Shvydka, D. and Sperling, N.N. (2021) Evaluation of Parameters Affecting Gamma Passing Rate in Patient‐Specific Qas for Multiple Brain Lesions IMRS Treatments Using Ray‐Station Treatment Planning System. Journal of Applied Clinical Medical Physics, 23, e13467.
https://doi.org/10.1002/acm2.13467
[18]  Chang, J., Wernicke, G. and Pannullo, S.C. (2018). Restricted Single Isocenter for Multiple Targets Dynamic Conformal Arc (RSIMT DCA) Technique for Brain Stereotactic Radiosurgery (SRS) Planning. Journal of Radiosurgery and SBRT, 5, 145-156.
https://pmc.ncbi.nlm.nih.gov/articles/PMC5893455/
[19]  McCulloch, J., Pawlowski, J., Kirby, N., Rasmussen, K., Shi, Z., Myers, P., et al. (2020) Patient‐Specific Dose Quality Assurance of Single‐Isocenter Multiple Brain Metastasis Stereotactic Radiosurgery Using PTW Octavius 4D. Journal of Applied Clinical Medical Physics, 21, 107-115.
https://doi.org/10.1002/acm2.12979
[20]  Junis, I., Yousif, Y., Stensmyr, R. and Barber, J. (2024) Comprehensive Characterisation of the IBA Myqa SRS for SRS and SBRT Patient Specific Quality Assurance. Physical and Engineering Sciences in Medicine, 47, 327-337.
https://doi.org/10.1007/s13246-023-01370-0
[21]  Knill, C., Sandhu, R., Halford, R., Snyder, M. and Seymour, Z. (2021) Commissioning Cranial Single‐Isocenter Multi‐Target Radiosurgery for the Versa Hd. Journal of Applied Clinical Medical Physics, 22, 108-114.
https://doi.org/10.1002/acm2.13223
[22]  Hoopingarner, S., Oldham, M., Adamson, J. and Lakrad, K. (2023) Investigation of a 3D Dosimetry System Utilizing Radio‐Chromic Gel and Telecentric Optical‐CT Readout. Medical Physics, 50, 5734-5744.
https://doi.org/10.1002/mp.16632
[23]  Calvo-Ortega, J., Greer, P.B., Moragues-Femenía, S., Pozo-Massó, M. and Casals-Farran, J. (2022) Verification of Stereotactic Radiosurgery Plans for Multiple Brain Metastases Using a Virtual Phantom-Based Procedure. Reports of Practical Oncology and Radiotherapy, 27, 449-457.
https://doi.org/10.5603/rpor.a2022.0042
[24]  Dunn, L. and Jolly, D. (2018) Automated Data Mining of a Plan‐Check Database and Example Application. Journal of Applied Clinical Medical Physics, 19, 739-748.
https://doi.org/10.1002/acm2.12396
[25]  Jolly, D., Dunn, L. and Kenny, J. (2016) A Clinical Database to Assess Action Levels and Tolerances for the Ongoing Use of Mobius3D. Journal of Applied Clinical Medical Physics, 18, 59-65.
https://doi.org/10.1002/acm2.12009
[26]  Han, Z., Hacker, F., Killoran, J., Kukluk, J., Aizer, A. and Zygmanski, P. (2019) Optimization of MLC Parameters for TPS Calculation and Dosimetric Verification: Application to Single Isocenter Radiosurgery of Multiple Brain Lesions Using Vmat. Biomedical Physics & Engineering Express, 6, Article 015004.
https://doi.org/10.1088/2057-1976/ab57ce
[27]  Adamson, J., Erickson, B.G., Wang, C., Cui, Y., Alber, M., Kirkpatrick, J., et al. (2023) Generation, Validation, and Benchmarking of a Commercial Independent Monte Carlo Calculation Beam Model for Multi-Target SRS. Zeitschrift für Medizinische Physik.
https://doi.org/10.1016/j.zemedi.2023.08.004
[28]  Erickson, B., Cui, Y., Alber, M., Wang, C., Fang Yin, F., Kirkpatrick, J., et al. (2024) Independent Monte Carlo Dose Calculation Identifies Single Isocenter Multi‐Target Radiosurgery Targets Most Likely to Fail Pre‐Treatment Measurement. Journal of Applied Clinical Medical Physics, 25, e14290.
https://doi.org/10.1002/acm2.14290
[29]  Palmans, H., Andreo, P., Huq, M.S., Seuntjens, J., Christaki, K.E. and Meghzifene, A. (2018) Dosimetry of Small Static Fields Used in External Photon Beam Radiotherapy: Summary of TRS‐483, the IAEA-AAPM International Code of Practice for Reference and Relative Dose Determination. Medical Physics, 45, e1123-e1145.
https://doi.org/10.1002/mp.13208
[30]  Lee, Y. C., & Kim, Y. (2021). A Patient-Specific QA Comparison between 2D and 3D Diode Arrays for Single-Lesion SRS and SBRT Treatments. Journal of Radiosurgery and SBRT, 7, 295-307.
https://pmc.ncbi.nlm.nih.gov/articles/PMC8492049/
[31]  Benedict, S.H., Yenice, K.M., Followill, D., Galvin, J.M., Hinson, W., Kavanagh, B., et al. (2010) Stereotactic Body Radiation Therapy: The Report of AAPM Task Group 101. Medical Physics, 37, 4078-4101.
https://doi.org/10.1118/1.3438081
[32]  Xia, Y., Adamson, J., Zlateva, Y., & Giles, W. (2020). Application of TG-218 Action Limits to SRS and SBRT Pre-Treatment Patient Specific QA. Journal of Radiosurgery and SBRT, 7, 135-147.
https://pmc.ncbi.nlm.nih.gov/articles/PMC7717087/
[33]  James, S., Al‐Basheer, A., Elder, E., Huh, C., Ackerman, C., Barrett, J., et al. (2023) Evaluation of Commercial Devices for Patient Specific QA of Stereotactic Radiotherapy Plans. Journal of Applied Clinical Medical Physics, 24, e14009.
https://doi.org/10.1002/acm2.14009
[34]  Ramachandran, P., Tajaldeen, A., Taylor, D., Wanigaratne, D., Roozen, K. and Geso, M. (2017) Evaluation and Performance of Arccheck and Film Using Gamma Criteria in Pre-Treatment Quality Assurance of Stereotactic Ablative Radiotherapy. Journal of Medical Physics, 42, 251-257.
https://doi.org/10.4103/jmp.jmp_132_16
[35]  Babu, P., Selvan, K., Padma, G., Revathy, M., Nambi Raj, N. and Senthilnathan, K. (2019) Dosimetric Effect of Rotational Setup Errors in Single-Isocenter Volumetric-Modulated Arc Therapy of Multiple Brain Metastases. Journal of Medical Physics, 44, 84-90.
https://doi.org/10.4103/jmp.jmp_103_18
[36]  Liang, B., Liu, B., Zhou, F., Yin, F. and Wu, Q. (2016) Comparisons of Volumetric Modulated Arc Therapy (VMAT) Quality Assurance (QA) Systems: Sensitivity Analysis to Machine Errors. Radiation Oncology, 11, Article No. 146.
https://doi.org/10.1186/s13014-016-0725-4
[37]  Mark, T. (2022). Error Detection Thresholds of Diode Detector Arrays in SRT Patient Specific QA: ESTRO 2022, 6-10 May 2022, PO-1555.
https://www.estro.org/Congresses/ESTRO2022/664/dosimetry/11161/errordetectionthresholdsofdiodedetectorarraysinsrt
[38]  Li, G., Bai, S., Chen, N., Henderson, L., Wu, K., Xiao, J., et al. (2013) Evaluation of the Sensitivity of Two 3D Diode Array Dosimetry Systems to Setup Error for Quality Assurance (QA) of Volumetric‐Modulated Arc Therapy (VMAT). Journal of Applied Clinical Medical Physics, 14, 13-24.
https://doi.org/10.1120/jacmp.v14i5.3828
[39]  Miften, M., Olch, A., Mihailidis, D., Moran, J., Pawlicki, T., Molineu, A., et al. (2018) Tolerance Limits and Methodologies for IMRT Measurement‐Based Verification QA: Recommendations of AAPM Task Group No. 218. Medical Physics, 45, e53-e83.
https://doi.org/10.1002/mp.12810

Full-Text

Contact Us

service@oalib.com

QQ:3279437679

WhatsApp +8615387084133