Functional Confirmation Using a Medical X-Ray System of a Semiconductor Survey Meter

DOI: 10.4236/ojrad.2024.141001, PP. 1-13

Keywords: Semiconductor Survey Meter, Functional Confirmation, Medical X-Ray System, Calibration Factor, Time Constant

Full-Text   Cite this paper   Add to My Lib

Abstract:

In recent years, semiconductor survey meters have been developed and are in increasing demand worldwide. This study determined if it is possible to use the X-ray system installed in each medical facility to calculate the time constant of a semiconductor survey meter and confirm the meter’s function. An additional filter was attached to the medical X-ray system to satisfy the standards of N-60 to N-120, more copper plates were added as needed, and the first and second half-value layers were calculated to enable comparisons of the facility’s X-ray system quality with the N-60 to N-120 quality values. Next, we used a medical X-ray system to measure the leakage dose and calculate the time constant of the survey meter. The functionality of the meter was then checked and compared with the energy characteristics of the meter. The experimental results showed that it was possible to use a medical X-ray system to reproduce the N-60 to N-120 radiation quality values and to calculate the time constant from the measured results, assuming actual leakage dosimetry for that radiation quality. We also found that the calibration factor was equivalent to that of the energy characteristics of the survey meter.

References

[1]  Ishii, H., Saturai, K., Uesugi, N., Kato, M., Sannohe, M., Miyata, K., Inaba, Y. and Chida, K. (2018) Fundamental Characteristics of a Semiconductor Survey Meter. Japanese Society of Radiation Safety Management, 17, 2-8.
[2]  Saturai, K., Ishii, H., Haga, Y., Kaga, Y., Sato, H., Honda, T., Inaba, Y. and Chida, K. (2018) Performance Evaluation of New Survey Meter Capable of the Energy Measurement in Diagnostic Radiology. Japanese Society of Radiation Safety Management, 17, 114-120.
[3]  Kobayashi, R., Chida, K., Inaba, Y., Haga, Y., Kaga, Y. and Zuguchi, M. (2015) Fundamental Study of a Radiation Dose Measuring Tool for Diagnostic X-Ray Apparatus. Bulletin of School of Health Sciences Tohoku University, 24, 39-44.
[4]  Hourdakis, C.J., Boziari, A. and Manetou, A. (2010) Performance Evaluation of Diagnostic Radiology Dosimeters in Clinical and Calibration X-Ray Beams. Health Physics, 98, 704-716.
https://doi.org/10.1097/HP.0b013e3181d18d45
[5]  Terasaki, K., Fujibuchi, T., Murazaki, H., Kuramoto, T., Umezu, Y., Ishigaki, Y. and Matsumoto, Y. (2017) Evaluation of Basic Characteristics of a Semiconductor Detector for Personal Radiation Dose Monitoring. Radiological Physics and Technology, 10, 189-194.
https://doi.org/10.1007/s12194-016-0384-z
[6]  Fujibuchi, T., Iimori, T., Masuda, Y., Uchida, Y., Isobe, T. and Sakae, T. (2010) Evaluation of a Real-Time Semiconductor Dosimeter and Measurement of Finger Dose in Nuclear Medicine Departments. Radiological Physics and Technology, 3, 53-57.
https://doi.org/10.1007/s12194-009-0076-z
[7]  Posar, J.A., Davis, J., Brace, O., Sellin, P., Griffith, M.J., Dhez, O., Wilkinson, D., Lerch, M.L.F., Rosenfeld, A. and Petasecca, M. (2020) Characterization of a Plastic Dosimeter Based on Organic Semiconductor Photodiodes and Scintillator. Physics and Imaging in Radiation Oncology, 14, 48-52.
https://doi.org/10.1016/j.phro.2020.05.007
[8]  Edited by Japanese Standards Association (2018) JIS Z 4511 Calibration Methods for Irradiation Dosimeters and Dose Equivalent Meters. Japanese Standards Association, Tokyo.
[9]  ISO (2019) X and Gamma Reference Radiation for Calibrating Dosemeters and Dose Rate Meters and for Determining Their Response as a Function of Photon Energy. Part 1: Radiation Characteristics and Production Methods. ISO 4037-1.
[10]  ISO (2019) X and Gamma Reference Radiation for Calibrating Dosemeters and Dose Rate Meters and for Determining Their Response as a Function of Photon Energy. Part 2: Dosimetry for Radiation Protection over the Energy Ranges 8 KeV to 1,3 MeV and 4 MeV to 9 MeV. ISO 4037-2.
[11]  ISO (1999) X and Gamma Reference Radiation for Calibrating Dosemeters and Dose Rate Meters and for Determining Their Response as a Function of Photon Energy—Part 3: Calibration of Area and Personal Dosemeters and the Measurement of Their Response as a Function of Energy and Angle of Incidence. ISO 4037-3.
[12]  ISO (2019) X and Gamma Reference Radiation for Calibrating Dosemeters and Dose Rate Meters and for Determining Their Response as a Function of Photon Energy. Part 4: Calibration of Area and Personal Dosemeters in Low Energy X Reference Radiation Fields. ISO 4037-4.
[13]  Negishi, T., Asada, Y., Ochiai, K., Kato, Y., Koyama, S., Sakamoto, H., Sato, H., Shoji, T., Sekimoto, M., Noto, K. and Minamoto, T. (2013) Standard Methods for Measuring Absorbed Dose in the Diagnostic X-Ray Field. Ohmsha, Ltd., Tokyo.
[14]  Nishitani, G., Yamada, K., Maekoshi, H., Kato, Y., Sato, H., Araki, F., Koyama, S., Koshida, K., Horii, H. and Onishi, H. (2013) Radiometry. Japanese Society of Radiological Technology, Kyoto.
[15]  Nakamura, M., Watanabe, Y., Yasuda, N., Shimazaki, H., Kanamori, I., Nakagiri, Y. and Onogi, M. (2007) Medical Radiation Measurement. Iryokagakusha Co., Ltd., Tokyo.
[16]  (1995) ISO/IEC Guide 98-3:2008, Uncertainty of Measurement-Part 3: Guide to the Expression of Uncertainty in Measurement (GUM).
[17]  (2012) Guide for Uncertainty Estimation, Accreditation Center of National Institute of Technology and Evaluation (JCSS), Accreditation-Division-JCG217S11-01.
[18]  Fluke Biomedical: RAYSAFE 452 Users Manual.
[19]  Aoyagi, T., Ogura, I., Abe, S. and Shimizu, E. (2008) Radiological Instrumentation (I). Corona Publishing Co., Ltd., Tokyo.
[20]  National Institute of Standards and Technology (2024) X-Ray Form Factor, Attenuation, and Scattering Tables.
https://physics.nist.gov/PhysRefData/FFast/html/form.html

Full-Text

Contact Us

service@oalib.com

QQ:3279437679

WhatsApp +8615387084133