Purpose: To present a protocol of a dual-field rotational
(DFR) total skin electron therapy (TSET) and to provide an assessment of
clinical implementation, dosimetry properties, and skin dose evaluation. Methods
and Materials: The DFR-TSET combined the Stanford 6-field and McGill
rotational methods. Dual 6 MeV electron beams in high dose total skin electron
mode were used for DFR-TSET on a commercial linac. Beam profiles and dosimetric
properties were measured using solid phantoms. The dose rate at expanded
source-to-surface distance (SSD) was a combination of static rate and
rotational rate. Invivo dosimetry of patient skin was performed on patients’ skin using film, metal oxide semiconductor
field-effect transistors (MOSFET), and optically stimulated luminescent dosimeters (OSLD). Results: Dual field
rotational total skin electron therapy exhibited good (≤±10%) uniformity in the beam profiles in the
vertical direction at an extended SSD of 332cm with a gantry angulation of ±20˚ deviated from the horizontal direction. In-vivo measurements confirmed acceptable uniformity of the patients’ total body
surfaces and revealed anatomically self-blocked or shielded areas where underdosing
occurred. Conclusions: The clinical implementation of DFR-TSET
effectively utilizes the special mode on a linac. This technique provides short
beam-on times, uniform dose distribution, large treatment field, and reduced
dose of x-ray contamination to the patients. In-vivo measurements indicate satisfactory delivery and dose
uniformity of the prescribed dose. Electron boost fields are recommended at
normal SSDs to address underdosed areas.
References
[1]
Karzmark, C.J., Loevinger, R., Steele, R.E. and Weissbluth, M. (1960) A Technique for Large-Field, Superficial Electron Therapy. Radiology, 74, 633-644. https://doi.org/10.1148/74.4.633
[2]
Karzmark, C.J. (1964) Large-Field Superficial Electron Therapy with Linear Accelerators. The British Journal of Radiology, 37, 302-305. https://doi.org/10.1259/0007-1285-37-436-302
[3]
Karzmark, C.J., Anderson, J., Fessenden, F., Svensson, G., et al. (1988) Total Skin Electron Therapy (Technique and Dosimetry. AAPM Report #23, TG-30). American Institute of Physics, New York. https://doi.org/10.37206/22
[4]
Das, I.J., Copeland, J.F. and Bushe, H.S. (1994) Spatial Distribution of Bremsstrahlung in a Dual Electron Beam Used in Total Skin Electron Treatments. Medical Physics, 21, 1733-1738. https://doi.org/10.1118/1.597215
[5]
El-Khatib, E., Hussein, S., Nikolic, M., et al. (1995) Variation of Electron Beam Uniformity with Beam Angulation and Scatterer Position for Total Skin Irradiation with the Stanford Technique. International Journal of Radiation Oncology, Biology, Physics, 33, 469-474. https://doi.org/10.1016/0360-3016(95)00112-C
[6]
Podgorsak, E.B., Pla, C., Pla, M., Lefebvre, P.Y. and Heese, R. (1983) Physical Aspects of a Rotational Total Skin Electron Irradiation. Medical Physics, 10, 159-168. https://doi.org/10.1118/1.595296
[7]
Pla, C., Heese, R., Pla, M. and Podgorsak, E.B. (1984) Calculation of Surface Dose in Rotational Total Skin Electron Irradiation. Medical Physics, 10, 539-546. https://doi.org/10.1118/1.595524
[8]
Podgorsak, E.B. and Podgorsak, M.B. (1999) Total Skin Electron Irradiation. The Modern Technology of Radiation Oncology, In Van Dyk, J., Ed., Medical Physics Publishing, Madison, 663.
[9]
Williams, P.C., Hunter, R.D. and Jackson, S.M. (1979) Whole Body Electron Therapy in Mycosis Fungoides—A Successful Translational Technique Achieved by Modification of an Established Linear Accelerator. The British Journal of Radiology, 52, 302-307. https://doi.org/10.1259/0007-1285-52-616-302
[10]
Nien, H.H., Hsieh, C.H., Shueng, P.W. and Tien, H.J. (2023) Total Skin Treatment with Helical Arc Radiotherapy. International Journal of Molecular Sciences, 24, Article No. 4492. https://doi.org/10.3390/ijms24054492
[11]
Harrison, C., Young, J., Navi, D., Riaz, N., Lingala, B., Kim, Y. and Hoppe, R. (2011) Revisiting Low-Dose Total Skin Electron Beam Therapy in Mycosis Fungoides. International Journal of Radiation Oncology, Biology, Physics, 81, e651-e657. https://doi.org/10.1016/j.ijrobp.2011.01.023
[12]
Miao, T., Zhang, R., Jermyn, M., Bruza, P., Zhu, T., Pogue, B.W., Gladstone, D.J. and Williams, B.B. (2022) Computational Dose Visualization & Comparison in Total Skin Electron Treatment Suggests Superior Coverage by the Rotational versus the Stanford Technique. Journal of Medical Imaging and Radiation Sciences, 53, 612-622. https://doi.org/10.1016/j.jmir.2022.08.006
[13]
Diamantopoulos, S., Platoni, K., Dilvoi, M. and Nazos, I. (2011) Clinical Implementation of Total Skin Electron Beam (TSEB) Therapy: A Review of the Relevant Literature. Physica Medica, 27, 62-68. https://doi.org/10.1016/j.ejmp.2010.09.001
[14]
Piotrowski, T., Milecki, P., Skórska, M. and Fundowicz, D. (2013) Total Skin Electron Irradiation Techniques: A Review. Advances in Dermatology and Allergology, 30, 50-55. https://doi.org/10.5114/pdia.2013.33379
[15]
Chowdhary, M., Chhabra, A.M., Kharod, S. and Marwaha, G. (2016) Total Skin Electron Beam Therapy in the Treatment of Mycosis Fungoides: A Review of Conventional and Low-Dose Regimens. Clinical Lymphoma, Myeloma and Leukemia, 16, 662-671. https://doi.org/10.1016/j.clml.2016.08.019
[16]
Piotrowski, T. and Malicki, J. (2006) The Rotary Dual Technique for Total Skin Irradiation in the Treatment of Mycosis Fungoides. Reports of Practical Oncology and Radiotherapy, 11, 29-37. https://doi.org/10.1016/S1507-1367(06)71047-1
[17]
Canu, D., Pham-Ledard, A., Ouhabrache, N. and Beylot-Barry, M. (2022) Changes in Total Skin Electron Beam Therapy Modalities for Mycosis Fungoides: A Single-Centre Study. Annales de Dermatologie et de Vénéréologie, 149, 180-184. https://doi.org/10.1016/j.annder.2022.02.006
[18]
Smits, K., Quint, K.D., Vermeer, M.H., Daniëls, L.A., Willemze, R., Jansen, P.M., Jansen, W.P.A. and Neelis, K.J. (2022) Total Skin Electron Beam Therapy for Cutaneous T-Cell Lymphomas in the Netherlands: A Retrospective Analysis of Treatment Outcomes and Selection for High or Low Dose Schedule. Clinical and Translational Radiation Oncology, 33, 77-82. https://doi.org/10.1016/j.ctro.2021.12.001
[19]
Morris, S., Scarisbrick, J., Frew, J., Irwin, C., Grieve, R., Humber, C., Kuciejewska, A., Bayne, S., Weatherhead, S., Child, F., Wain, M. and Whittaker, S. (2017) The Results of Low-Dose Total Skin Electron Beam Radiation Therapy (TSEB) in Patients with Mycosis Fungoides from the UK Cutaneous Lymphoma Group. International Journal of Radiation Oncology, Biology, Physics, 99, 627-633. https://doi.org/10.1016/j.ijrobp.2017.05.052
[20]
Xie, Y., Petroccia, H., Maity, A., Miao, T., Zhu, Y., Bruza, P., Pogue, B.W., Plastaras, J.P., Dong, L. and Zhu, T.C. (2020) Cherenkov Imaging for Total Skin Electron Therapy (TSET). Medical Physics, 47, 201-212. https://doi.org/10.1002/mp.13881
[21]
Ding, G.X., Osmundson, E.C., Shinohara, E., Newman, N.B., Price, M. and Kirschner, A.N. (2021) Monte Carlo Study on Dose Distributions from Total Skin Electron Irradiation Therapy (TSET). Physics in Medicine & Biology, 66, Article ID: 075010. https://doi.org/10.1088/1361-6560/abedd7
[22]
Ding, G.X. (2021) Stopping-Power Ratios for Electron Beams Used in Total Skin Electron Therapy. Medical Physics, 48, 5472-5478. https://doi.org/10.1002/mp.15121
[23]
Elsayad, K., Weishaupt, C., Moustakis, C., Danzer, M.F., Müller, E.C., et al. (2023) Ultrahypofractionated Low-Dose Total Skin Electron Beam in Advanced-Stage Mycosis Fungoides and Sézary Syndrome. International Journal of Radiation Oncology, Biology, Physics, 117, 164-170. https://doi.org/10.1016/j.ijrobp.2023.02.052
[24]
Elmahmoud, Z., Gunther, J.R. and Christopherson, K. (2023) A Simple Solution to Create a Custom Scalp-Sparing Helmet to Prevent Alopecia in Patients Undergoing Total Skin Electron Beam Therapy for Cutaneous T Cell Lymphoma. Clinical and Translational Radiation Oncology, 38, 53-56. https://doi.org/10.1016/j.ctro.2022.10.005
[25]
Newman, N.B., Patel, C.G., Ding, G.X., Zic, J.A., Zwerner, J., Osmundson, E.C. and Kirschner, A.N. (2021) Prospective Observational Trial of Low-Dose Skin Electron Beam Therapy in Mycosis Fungoides Using a Rotational Technique. Journal of the American Academy of Dermatology, 85, 121-127. https://doi.org/10.1016/j.jaad.2020.12.023
[26]
Kumar, P.P., Good, R.R., Jones, E.O., McAnulty, B.E. and Reeves, M.A. (1987) Dual-Field Rotational Technique for Total-Skin Electron-Beam Therapy. American Journal of Clinical Oncology, 10, 344-354. https://doi.org/10.1097/00000421-198708000-00018
[27]
Heumann, T.R., Esiashvili, N., Parker, S., Switchenko, J.M., Dhabbaan, A., Goodman, M., Lechowicz, M.J., Flowers, C.R. and Khan, M.K. (2015) Total Skin Electron Therapy for Cutaneous T-Cell Lymphoma Using a Modern Dual-Field Rotational Technique. International Journal of Radiation Oncology, Biology, Physics, 92, 183-191. https://doi.org/10.1016/j.ijrobp.2014.11.033
[28]
Peters, V.G. and Jaywant, S.M. (1995) Implementation of Total Skin Electron Therapy Using an Optional High Dose Rate Mode on a Conventional Linear Accelerator. Medical Dosimetry, 20, 99-104. https://doi.org/10.1016/0958-3947(95)00013-M
[29]
Chen, Z., Agostinelli, A.G., Wilson, L.D. and Nath, R. (2004) Matching Dosimetry Characteristics of a Dual-Field Stanford Technique to a Customized Single-Field Stanford Technique for Total Skin Electron Therapy. International Journal of Radiation Oncology, Biology, Physics, 59, 872-885. https://doi.org/10.1016/j.ijrobp.2004.02.046
[30]
Varian Medical System (2012) Clinac Technical Reference Guide. Version 9.1, PN 100054117-03, 101-102.
[31]
Fraass, B.A., Roberson, P.L. and Glatstein, E. (1983) Whole-Skin Electron Treatment: Patient Skin Dose Distribution. Radiology, 146, 811-814. https://doi.org/10.1148/radiology.146.3.6828698
[32]
Smeda, M.I., Johnston, D.O., Salzman, F.A., Trump, J.G. and Wright, K.A. (1962) Ten-Year Experience with Low Megavolt Electron Therapy. The American Journal of Roentgenology Radium Therapy and Nuclear Medicine, 88, 215-228.
[33]
Butson, M.J., Rozenfeld, A., Mathur, J.N., Carolan, M., Wong, T.Y. and Metcalfe, P.E. (1996) A New Radiotherapy Surface Dose Detector: The MOSFET. Medical Physics, 23, 655-658. https://doi.org/10.1118/1.597702
[34]
Qin, S., Chen, T., Wang, L., Tu, Y., Yue, N. and Zhou, J. (2014) Angular Dependence of the MOSFET Dosimeter and Its Impact on in Vivo Surface Dose Measurement in Breast Cancer Treatment. Technology in Cancer Research & Treatment, 13, 345-352. https://doi.org/10.7785/tcrt.2012.500382
[35]
Consorti, R., Petrucci, A., Fortunato, F., Soriani, A., Marzi, S., Iaccarino, G., Landoni, V. and Benassi, M. (2005) In Vivo Dosimetry with MOSFETs: Dosimetric Characterization and First Clinical Results in Intraoperative Radiotherapy. International Journal of Radiation Oncology, Biology, Physics, 63, 952-960. https://doi.org/10.1016/j.ijrobp.2005.02.049
[36]
Jursinic, P.A. (2010) Changes in Optically Stimulated Luminescent Dosimeter (OSLD) Dosimetric Characteristics with Accumulated Dose. Medical Physics, 37, 132-140. https://doi.org/10.1118/1.3267489
Khan, F.M. and Gibbons, J.P. (2014) The Physics of Radiation Therapy. 5th Edition, Lippincott Williams & Wilkins, Philadelphia.
[39]
Attix, F.H. (1986) Introduction to Radiological Physics and Radiation Dosimetry. Wiley, New York. https://doi.org/10.1002/9783527617135
[40]
Gaza, R., McKeveer, S.W.S., Akselrod, M.S., Akselrod, U.A.T., Yoder, C., Andersen, C.E., Aznar, M.C., Marckmann, C.J. and Bøtter-Jensen, L. (2004) A Fiber Dosimetry Method Based on OSL from Al2O3:C: for Radiotherapy Applications. Radiation Measurements, 38, 809-812. https://doi.org/10.1016/j.radmeas.2003.12.004
[41]
Mijnheer, B., Beddar, S., Izewska, J. and Reft, C. (2013) In Vivo Dosimetry in External Beam Radiotherapy. Medical Physics, 40, Article ID: 070903. https://doi.org/10.1118/1.4811216