|
眼球参数对iCare IC100眼压计和Goldmann压平式眼压计测量结果的影响
|
Abstract:
目的:探讨中央角膜厚度、角膜曲率、角膜散光、白到白、前房深度、晶体厚度和眼轴长度对iCare IC100眼压计与Goldmann压平式眼压计(Goldmann applanation tonometry, GAT)测量结果的影响。方法:前瞻性研究。选取47例(73眼)门诊患者作为研究对象。采用IOL Master700测量患者中央角膜厚度、角膜曲率、角膜散光、白到白、前房深度、晶体厚度和眼轴长度,使用iCare IC100眼压计和GAT对所有患者进行眼压测量。采用线性相关分析比较iCare IC100和GAT的测量结果以及iCare IC100和GAT与中央角膜厚度、角膜曲率、角膜散光、白到白、前房深度、晶体厚度和眼轴长度的相关性,并进一步采用多重线性回归法分析相关性好的眼球参数与两种眼压计测量结果的关系。结果:使用iCare IC100和GAT测量眼压的平均值分别为(19.01 ± 7.68) mmHg和(18.99 ± 7.58) mmHg (r = 0.942, P = 0.000)。中央角膜厚度与iCare IC100和GAT测量值均明显相关(r = 0.527, P = 0.000; r = 0.485, P = 0.000)。中央角膜厚度每增加1 μm,iCare IC100测量值增加0.094 mmHg,GAT测量值增加0.085 mmHg。中央角膜厚度对iCare IC100测量值的影响较其对于GAT测量值的影响大。角膜散光与iCare IC100和GAT的测量值均明显相关(r = 0.246, P = 0.036; r = 0.309, P = 0.008)。角膜散光每增加1 D,iCare IC100测量值增加2.045 mmHg,GAT测量值增加2.662 mmHg。角膜散光对iCare IC100测量值的影响较其对于GAT测量值的影响小。眼轴每增加1 mm,iCare IC100测量值增加0.593 mmHg,GAT测量值增加0.489 mmHg。眼轴对iCare ic100测量值的影响较其对于GAT测量值的影响大。角膜曲率、白到白、前房深度、晶体厚度和眼轴长度与iCare IC100和GAT的测量值均无显著相关性(P均 > 0.05)。结论:中央角膜厚度、角膜散光和眼轴长度对iCare IC100和GAT的测量值均有影响,中央角膜厚度和眼轴长度对iCare IC100测量值的影响较其对于GAT的影响大,角膜散光对iCare IC100测量值的影响较其对于GAT的影响小。
Objective: To investigate the effects of central corneal thickness, corneal curvature, corneal astigmatism, white to white, anterior chamber depth, lens thickness and axial length on the measurement results of iCare IC100 tonometer and Goldmann applanation tonometer (GAT). Methods: Prospective study. Forty-seven outpatients (73 eyes) were selected as study subjects. Central corneal thickness, corneal curvature, corneal astigmatism, white to white, anterior chamber depth, lens thickness and axial length were measured using the IOL Master700, and intraocular pressure measurements were performed in all patients using the iCare IC100 tonometer and GAT. Linear correlation analysis was used to compare the measurement results of iCare IC100 and GAT and the correlation between iCare IC100 and GAT and central corneal thickness, corneal curvature, corneal astigmatism, white to white, anterior chamber depth, lens thickness and eye axis length, and the relationship between the correlated parameters and the two tonometer measurements was further analyzed by multiple linear regression. Results: The mean values of iCare IC100 and GAT were (19.01 ± 7.68) mmHg and (18.99 ± 7.58) mmHg (r = 0.942, P = 0.000), respectively. Central corneal thickness was significantly correlated with iCare IC100 and GAT measurements (r = 0.527, P = 0.000; r = 0.485, P = 0.000). For every 1 μm increase in central corneal thickness, the iCare IC100 measurement increased by 0.094 mmHg and the GAT measurement increased by 0.085 mmHg. The
[1] | Soh, Z.D., Yu, M., Betzler, B.K., et al. (2021) The Global Extent of Undetected Glaucoma in Adults: A Systematic Review and Meta-Analysis. Ophthalmology, 128, 1393-1404. https://doi.org/10.1016/j.ophtha.2021.04.009 |
[2] | Vodencarevic, A.N., Jusufovic, V., Terzic, S. and Halilba?i?, M. (2016) Comparison of Intraocular Pressure Measurements Obtained by Rebound, Noncontact, and Goldmann Applanation Tonometry in Children. American Journal of Ophthalmology, 163, 192. https://doi.org/10.1016/j.ajo.2015.12.012 |
[3] | Tonnu, P.A., Ho, T., et al. (2005) The Influence of Central Corneal Thickness and Age on Intraocular Pressure Measured by Pneumotonometry, Non-Contact Tonometry, the Tono-Pen XL, and Goldmann Applanation Tonometry. British Journal of Ophthalmology, 89, 851-854. https://doi.org/10.1136/bjo.2004.056622 |
[4] | Wangsupadilok, B. and Horatanaruang, O. (2011) The Impact of Central Corneal Thickness on Intraocular Pressure Measured by Non-Contact Tonometry. Journal of the Medical Association of Thailand, 94, 574-578. |
[5] | Gupta, V., Sony, P., Agarwal, H.C., et al. (2006) Inter-Instrument Agreement and Influence of Central Cornealthickness on Measurements with Goldmann, Pneumotonometer and Noncontact Tonometer in Glaucomatous Eyes. Indian Journal of Ophthalmology, 54, 261-265. https://doi.org/10.4103/0301-4738.27952 |
[6] | Mansoori, T. and Balakrishna, N. (2017) Effect of Central Corneal Thickness on Intraocular Pressure and Comparison of Topcon CT-80 Non-Contact Tonometry with Goldmann Applanation Tonometry. Clinical & Experimental Optometry, 101, 206-212. https://doi.org/10.1111/cxo.12620 |
[7] | Whitacre, M.M. and Stein, R. (1993) Sources of Error with Use of Goldmann-Type Tonometers. Survey of Ophthalmology, 38, 1-30. https://doi.org/10.1016/0039-6257(93)90053-A |
[8] | Baudouin, C. and Gastaud, P. (1994) Influence of Topical Anesthesia on Tonometric Values of Intraocular Pressure. Ophthalmologica, 208, 309-313. https://doi.org/10.1159/000310527 |
[9] | Shunsuke, N. (2018) Icare Rebound Tonometers: Review of Their Characteristics and Ease of Use. Clinical Ophthalmology, 12, 1245-1253. https://doi.org/10.2147/OPTH.S163092 |
[10] | 郑伊苹, 赵平, 刘瑞, 等. iCare iC100与Corvis ST测量眼压的重复性和一致性比较[J]. 中华眼视光学与视觉科学杂志, 2021, 23(10): 745-751. https://doi.org/10.3760/cma.j.cn115909-20210222-00069 |
[11] | Gao, F., Liu, X., Zhao, Q. and Pan, Y.Z. (2017) Comparison of the iCare Rebound Tonometer and the Goldmann Applanation Tonometer. Experimental and Therapeutic Medicine, 13, 1912-1916. https://doi.org/10.3892/etm.2017.4164 |
[12] | Kohlhaas, M., Boehm, A.G., Spoerl, E., et al. (2006) Effect of Central Corneal Thickness, Corneal Curvature, and Axial Length on Applanation Tonometry. Archives of Ophthalmology, 124, 471-476.
https://doi.org/10.1001/archopht.124.4.471 |
[13] | Shah, S., Chatterjee, A., Mathai, M., et al. (1999) Relationship between Corneal Thickness and Measured Intraocular Pressure in a General Ophthalmology Clinic. Ophthalmology, 106, 2154-2160.
https://doi.org/10.1016/S0161-6420(99)90498-0 |
[14] | Harada, Y., Hirose, N., Kubota, T. and Tawara, A. (2008) The Influence of Central Corneal Thickness and Corneal Curvature Radius on the Intraocular Pressure as Measured by Different Tonometers: Noncontact and Goldmann Applanation Tonometers. Journal of Glaucoma, 17, 619-625. https://doi.org/10.1097/IJG.0b013e3181634f0f |
[15] | Martinez-de-la-Casa, J.M., Garcia-Feijoo, J., Vico, E., et al. (2006) Effect of Corneal Thickness on Dynamic Contour, Rebound, and Goldmann Tonometry. Ophthalmology, 113, 2156-2162. https://doi.org/10.1016/j.ophtha.2006.06.016 |
[16] | 张扬, 赵家良, 卞爱玲, 等. 中央角膜厚度、角膜曲率对Goldmann压平眼压计和非接触眼压计测量结果的影响[J]. 中华眼科杂志, 2009, 45(8): 713-718. |
[17] | 冯震. Icare回弹式眼压计、非接触式眼压计、Goldmann压平式眼压计的测量结果比较[D]: [硕士学位为论文]. 石家庄: 河北医科大学, 2012. https://doi.org/10.7666/d.y2105415 |
[18] | 赵姣, 税丹, 毛科, 等. 两种眼压计测量成人高度近视眼压比较及其与角膜生物力学关系的研究[J]. 中国科技期刊数据库医药, 2022(9): 19-22. |
[19] | 李阳明. 三种压平式眼压计的一致性分析[D]: [硕士学位论文]. 温州: 温州医科大学, 2011.
https://doi.org/10.7666/d.y1946068 |