Empirical Formalism for the Phantom Scatter Factor of Small Fields: Using Different Density Media

DOI: 10.4236/ijmpcero.2022.111004, PP. 36-47

Keywords: Small Field Dosimetry, Anthropomorphic Study, Phantom Scatter Factor, Dosimetry

Full-Text   Cite this paper   Add to My Lib

Abstract:

We present the Empirical Formula (EF) to calculate the phantom scatter factor, Sp, of small radiation fields under charge particle dis-equilibrium conditions. The Empirical Formula (EF) was verified by examining the calculated data with experimentally measured data utilizing the anthropomorphic phantom in twelve different combinations of beam entry and point location, where the value for Sp per tissue composition was within 3% in 8/12 cases, 5% in 1/12 cases, and 10% in 3/12 cases. Our results showed a good agreement with experimental data to less than 1% when the ion chamber was surrounded by the homogeneous tissue, whether lung, soft tissue, or bone. Indicating that the prediction of the equation is valid, and it can be reliably used for phantom scatter factor calculation for different homogeneous media under charge particle dis- equilibrium conditions.

References

[1]  Ding, G.X., Duggan, D.M. and Coffey, C.W. (2006) Commissioning Stereotactic Radiosurgery Beams Using Both Experimental and Theoretical Methods. Physics in Medicine & Biology, 51, 2549-2566.
https://doi.org/10.1088/0031-9155/51/10/013
[2]  Sauer, O.A. and Wilbert, J. (2007) Measurement of Output Factors for Small Photon Beams. Medical Physics, 34, 1983-1988.
https://doi.org/10.1118/1.2734383
[3]  Klein, D.M., et al. (2010) Measuring Output Factors of Small Fields Formed by Collimator Jaws and Multileaf Collimator Using Plastic Scintillation Detectors. Medical Physics, 37, 5541-5549.
https://doi.org/10.1118/1.3488981
[4]  Benmakhlouf, H., Sempau, J. and Andreo, P. (2014) Output Correction Factors for Nine Small Field Detectors in 6 MV Radiation Therapy Photon Beams: A PENELOPE Monte Carlo Study. Medical Physics, 41, Article ID: 041711.
https://doi.org/10.1118/1.4868695
[5]  Francescon, P., Cora, S. and Cavedon, C. (2008) Total Scatter Factors of Small Beams: A Multidetector and Monte Carlo Study. Medical Physics, 35, 504-513.
https://doi.org/10.1118/1.2828195
[6]  Back, M., et al. (2013) Dosimetric Improvements Utilising Intensity Modulated Radiation Therapy for Patients with Glioblastoma Multiforme. Journal of Cancer Therapy, 4, 18-24.
https://doi.org/10.4236/jct.2013.411A003
[7]  Huq, M.S., et al. (2018) A Dosimetric Evaluation of the IAEA-AAPM TRS483 Code of Practice for Dosimetry of Small Static Fields Used in Conventional Linac Beams and Comparison with IAEA TRS-398, AAPM TG51, and TG51 Addendum Protocols. Medical Physics, 45, 4257-4273.
https://doi.org/10.1002/mp.13092
[8]  Zhu, T.C., et al. (2009) Report of AAPM Therapy Physics Committee Task Group 74: In-Air Output Ratio, for Megavoltage Photon Beams. Medical Physics, 36, 5261-5291.
https://doi.org/10.1118/1.3227367
[9]  Masanga, W., et al. (2016) Determination of Small field Output Factors in 6 and 10 MV Flattening Filter Free Photon Beams Using Various Detectors. Journal of Physics: Conference Series, 694, Article ID: 012027.
https://doi.org/10.1088/1742-6596/694/1/012027
[10]  Ozdemir, H., Tuncel, N. and Kizildağ, A. (2017) The Impact of Small Field’s Off-Set on Output Factor in Intensity Modulated Radiation Treatment. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 6, 433-444.
https://doi.org/10.4236/ijmpcero.2017.64039
[11]  Haider, T.K. and El-Khatib, E.E. (1994) Measurements of Phantom Scatter Factors for Small Field Sizes in High Energy X Rays. Medical Physics, 21, 663-666.
https://doi.org/10.1118/1.597319
[12]  Charles, P.H., et al. (2014) A Practical and Theoretical Definition of very Small Field Size for Radiotherapy Output Factor Measurements. Medical Physics, 41, Article ID: 041707.
https://doi.org/10.1118/1.4868461
[13]  Klingman, E. (2021) A Self-Linking Field Formalism. Journal of Modern Physics, 12, 440-452.
https://doi.org/10.4236/jmp.2021.124031
[14]  Das, I.J., Ding, G.X. and Ahnesjo, A. (2008) Small Fields: Nonequilibrium Radiation Dosimetry. Medical Physics, 35, 206-215.
https://doi.org/10.1118/1.2815356
[15]  Bassinet, C., et al. (2013) Small Fields Output Factors Measurements and Correction Factors Determination for Several Detectors for a CyberKnife(R) and Linear Accelerators Equipped with MicroMLC and Circular Cones. Medical Physics, 40, Article ID: 071725.
https://doi.org/10.1118/1.4811139
[16]  Nasir, M., et al. (2017) Measurement and Analysis of PDDs Profile and Output Factors for Small Field Sizes by cc13 and Micro-Chamber cc01. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 6, 36-56.
https://doi.org/10.4236/ijmpcero.2017.61005
[17]  Islam, M.R., Watabe, H. and Andreoli, S. (2020) Measurement and Comparison of Output Factors Using Two Detectors for NOVAC7 IntraOperative Radiotherapy Accelerator. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 9, 52-61.
https://doi.org/10.4236/ijmpcero.2020.92006
[18]  Hrbacek, J., Lang, S. and Klock, S. (2011) Commissioning of Photon Beams of a Flattening Filter-Free Linear Accelerator and the Accuracy of Beam Modeling Using an Anisotropic Analytical Algorithm. International Journal of Radiation Oncology, Biology, Physics, 80, 1228-1237.
https://doi.org/10.1016/j.ijrobp.2010.09.050
[19]  Laub, W.U. and Wong, T. (2003) The Volume Effect of Detectors in the Dosimetry of Small Fields Used in IMRT. Medical Physics, 30, 341-347.
https://doi.org/10.1118/1.1544678

Full-Text

Contact Us

service@oalib.com

QQ:3279437679

WhatsApp +8615387084133