Design and Manufacture an Elastic Neutron Scattering Spectrometer at the Dalat Nuclear Reactor

DOI: 10.4236/jasmi.2021.113003, PP. 23-28

Keywords: Neutron Scattering, Small-Angle Neutron Scattering (SANS), Elastic Neu-tron Cross Section

Full-Text   Cite this paper   Add to My Lib

Abstract:

The objective of this study is to design an elastic neutron scattering system according to the angle with a sample using thermal neutron beam at the Dalat Nuclear Reactor (DNR). The system is used for research and training in the field of material structure analysis by neutron scattering and diffraction techniques. It is designed on the basis of inheriting the neutron measurement spectrometer systems at the DNR and the scattered neutron measurement systems in the world. The measuring system, which was installed at the horizontal channel 4 of the DNR, consists of 5-helium-3 detectors and a fully electronic system to record the scatter counts and a mechanical system with the possibility of rotating at 15˚-75˚ angles. The constructed system is tested for evaluation of the accuracy, stability and reliability of the mechanical and electronic systems of moving detectors by angles.

References

[1]  Anh, T.T., Son, P.N., Tan, V.H., Khang, P.D. and Hoa, P.C. (2014) Characteristics of Filtered Neutron Beam Energy Spectra at Dalat Reactor. World Journal of Nuclear Science and Technology, 4, 96-102.
https://doi.org/10.4236/wjnst.2014.42015
[2]  Iijima, T., Mukaiyama, T. and Shirakata, K. (1971) Measurement of Fast Neutron Spectrum with Helium-3 Proportional Counter. Journal of Nuclear Science and Technology, 8, 192-200.
https://doi.org/10.1080/18811248.1971.9735315
[3]  Ibaraki, M., Baba, M., Miura, T., Hirasawa, Y., Nauchi, Y., Nakashima, H., Meigo, S.-I., Iwamoto, O. and Tanaka, S. (2000) Experimental Method for Neutron Elastic Scattering Cross-Section Measurement in 40-90 MeV Region at TIARA. Nuclear Instruments and Methods in Physics Research A, 446, 536-544.
https://doi.org/10.1016/S0168-9002(99)01274-7
[4]  Stokes, R. (2005) Helium 3 Detector 207 Formation. General Electric Company.
[5]  Canberra (2007) Model 2006 Proportional Counter Preamplifier Datasheet.
[6]  Canberra (2010) 3002D High Voltage Power Supply Datasheet.
[7]  Canberra (2010) 2022 Amplifier Datasheet.
[8]  Canberra (2017) Canbera Multiport II Datasheet.
[9]  Microchip Technology, Inc. (2007) Pic16F886 Datasheet.