In order to solve the lack
of relevant evaluation research on the accuracy of HMP155A humidity sensor
calibration results in the past, this paper designs the corresponding
experimental scheme, and obtains the corresponding calibration results
according to the experimental scheme; Then the measurement uncertainty of the
indication error in the calibration results is evaluated by GUM, and the
corresponding extended uncertainty U95 is obtained. Finally,
according to the requirements of JJF1094-2016 characteristic evaluation of
measuring instruments, combined with the calibration results and the actual
situation of U95, the conformity of the indication error of
calibration is determined. The result is that each calibration point of the
sensor meets the requirements of conformity determination and is within the
qualified range. This research effectively makes up for the blank of the
previous research on the conformity determination of the indication error of
the calibration results and has strong theoretical and practical significance.
References
[1]
(2008) ISO/IEC Guide 98-3: 2008, Uncertainty of Measurement—Part 3: Guide to the Expression of Uncertainty in Measurement (GUM).
[2]
Lu, T. and Chen, C. (2007) Uncertainty Evaluation of Humidity Sensors Calibrated by Saturated Salt Solutions. Measurement, 40, 591-599. https://doi.org/10.1016/j.measurement.2006.09.012
[3]
Zhang, C., Zhang, W., Webb, D.J., et al. (2010) Optical Fibre Temperature and Humidity Sensor. Electronics Letters, 46, 643-644. https://doi.org/10.1049/el.2010.0879
[4]
Yu, Y., et al. (2021) Multi-Channel Impedance Measurement System for Polymer Humidity Sensors. Measurement, 176, Article No. 109113. https://doi.org/10.1016/j.measurement.2021.109113
[5]
Sharma, H., Vaidya, U. and Ganapathysubramanian, B. (2019) A Transfer Operator Methodology for Optimal Sensor Placement Accounting for Uncertainty. Building and Environment, 155, 334-349. https://doi.org/10.1016/j.buildenv.2019.03.054
[6]
Martins, L.L., Ribeiro, A.S., e Sousa. J.A., et al. (2012) Measurement Uncertainty of Dew-Point Temperature in a Two-Pressure Humidity Generator. International Journal of Thermophysics, 33, 1568-1582. https://doi.org/10.1007/s10765-011-1005-z
[7]
Hans, I., Burgdorf, M., Buehler, S.A., et al. (2019) An Uncertainty Quantified Fundamental Climate Data Record for Microwave Humidity Sounders. Remote Sensing, 11, Article No. 548. https://doi.org/10.3390/rs11050548
[8]
Yang, Z., Zhang, H., Wang, Q., et al. (2021) Uncertainty Evaluation of Wind Speed Sensor in Automatic Weather Station. IOP Conference Series: Earth and Environmental Science, 768, Article No. 012008. https://doi.org/10.1088/1755-1315/768/1/012008
[9]
Wei, M., Zeng, Y., Wen, C., et al. (2019) Comparison of MCM and GUM Method for Evaluating Measurement Uncertainty of Wind Speed by Pitot Tube. MAPAN, 34, 345-355. https://doi.org/10.1007/s12647-019-00339-3
[10]
Wei, M., Zeng, Y., Zou, L., et al. (2019) Analysis of the Influence of Water-Vapor Correction Term on the Measurement Uncertainty of Wind Speed. MAPAN, 34, 333-343. https://doi.org/10.1007/s12647-019-00338-4