Determination of the effective radiation dose of the patient in computed tomographyKeywords: Bilgisayarl? tomografi,TLD,X ???nlar?,ASIR Abstract: In recent years, the use of computed tomography (CT) has been recognized by all authorities as an indispensable imaging technology for diagnosing patients. With the introduction of multislice CT devices that were very fast, the patients could be examined without the need for anesthesia. This has led to an excessive increase in the use of CT in pediatric patients. In adults, a large part of the excessive increase is observed in asymptomatic patients such as CT colonoscopy, cardiac angiography and whole body scan. The excessive use of CT increases the risk of exposure to radiation. In this study, in order to reveal the actual radiation dose that the patient was exposed to in the CT examinations, a comparison was made between the value obtained from the CT device and the radiation dose exposed to by the patient. Thermoluminescence (TLD) detectors were placed on 25 patients. It was found that there was a 4% difference between the radiation dose read from the CT and the radiation dose from the TLD crystals
|