全部 标题 作者
关键词 摘要

OALib Journal期刊
ISSN: 2333-9721
费用:99美元

查看量下载量

相关文章

更多...
-  2014 

Automixer: equipment for the reduction of risks associated with inadequate oxygen supply

Keywords: Blender, fuzzy logic, newborn, pulse oxymetry, oxygen saturation

Full-Text   Cite this paper   Add to My Lib

Abstract:

In this paper, the AUTOMIXER, a medical support equipment for the reduction of risks associated with the inadequate supply of oxygen, is presented. The equipment automatically regulates the Inspired Oxygen Fraction (FIO2 concentration), which is supplied to the patient according to user parameters, measures of oxygen saturation and patient's status. The AUTOMIXER is based on a fuzzy control system, which was implemented in a microcontroller using a LookUp Table. The operation of the device was validated using simulated data and actual patients. The results show that the AUTOMIXER is an important tool for decreasing the incidence of pathologies related to unsuitable handling (excess or lack) of oxygen

Full-Text

Contact Us

service@oalib.com

QQ:3279437679

WhatsApp +8615387084133