Objective: To evaluate feasibility of MRI in patients with non-pacemaker (PM)/Implantable cardioverter defibrillator (ICD) metallic devices and abandoned leads. Background: Relative safety of MRI performed using specified protocol has been established in MR non-conditional PM/ICDs. With limited safety data, many non-PM/ICD metallic devices and abandoned leads continue to be a contraindication for MRI. Methods: We retrospectively analyzed consecutive patients with extra-cardiac devices, non-programmable cardiac devices, and abandoned leads, who underwent MRI (GE 1.5 Tesla, WI) at a single tertiary care center over a span of 13 years. Scan protocol was designed to maintain specific absorption rate (SAR) < 4.0 W/kg and scan time < 60 minutes. Results: The cohort comprised 127 MRI exams representing 94 patients, with 13 patients having two or more scans. The devices consisted of: 23 vagal nerve stimulators (VNS), 22 implantable loop recorders, 16 spinal stimulators, 5 peripheral nerve stimulators, 3 bladder stimulators, 2 deep brain stimulators, 1 gastric stimulator, 1 bone stimulator, 1 WATCHMAN device, 22 abandoned PM/lCD leads and 1 VNS lead. There was no immediate (peri-MRI exam) morbidity or mortality. Patients did not report any discomfort, palpitations, heating, or sensation of device migration during the exam. Local follow-up data was available in 65% (100% for thoracic imaging) with a mean of 190 ± 475 days (median 13 days). No device malfunction was reported during follow-up. Conclusions: With appropriate precautions, MRI is feasible in patients with extracardiac devices, nonprogrammable cardiac devices, and abandoned leads.
References
[1]
Damadian, R., Goldsmith, M. and Minkoff, L. (1977) NMR in Cancer: XVI. FONAR Image of the Live Human Body. Physiological Chemistry and Physics, 9, 97-100, 108.
[2]
(2008) Sentinel Event Alert, Issue 38: Preventing Accidents and Injuries in the MRI Suite.
[3]
Expert Panel on MRS, Kanal, E., Barkovich, A.J., Bell, C., Borgstede, J.P., Bradley, W.G., et al. (2013) ACR Guidance Document on MR Safe Practices: 2013. Journal of Magnetic Resonance Imaging, 37, 501-530.
https://doi.org/10.1002/jmri.24011
[4]
Davis, P.L., Crooks, L., Arakawa, M., McRee, R., Kaufman, L. and Margulis, A.R. (1981) Potential Hazards in NMR Imaging: Heating Effects of Changing Magnetic Fields and RF Fields on Small Metallic Implants. American Journal of Roentgenology, 137, 857-860. https://doi.org/10.2214/ajr.137.4.857
[5]
Zheutlin, J.D., Thompson, J.T. and Shofner, R.S. (1987) The Safety of Magnetic Resonance Imaging with Intraorbital Metallic Objects after Retinal Reattachment or Trauma. American Journal of Ophthalmology, 103, 831.
https://doi.org/10.1016/S0002-9394(14)74404-6
[6]
Shellock, F.G. (2002) Magnetic Resonance Safety Update 2002: Implants and Devices. Journal of Magnetic Resonance Imaging, 16, 485-496. https://doi.org/10.1002/jmri.10196
[7]
Faris, O.P. and Shein, M.J. (2005) Government Viewpoint: U.S. Food & Drug Administration: Pacemakers, ICDs and MRI. Pacing and Clinical Electrophysiology, 28, 268-269.
https://doi.org/10.1111/j.1540-8159.2005.50035.x
[8]
(2011) FDA Approves First Pacemaker System Safe for Magnetic Resonance Imaging. Traditional Cardiac Devices Have Been at Risk of Malfunctioning or Overheating with MRIs. Heart Advisor, 14, 1, 11.
[9]
Nazarian, S., Hansford, R., Roguin, A., Goldsher, D., Zviman, M.M., Lardo, A.C., et al. (2011) A Prospective Evaluation of a Protocol for Magnetic Resonance Imaging of Patients with Implanted Cardiac Devices. Annals of Internal Medicine, 155, 415-424. https://doi.org/10.7326/0003-4819-155-7-201110040-00004
[10]
Russo, R.J., Costa, H.S., Silva, P.D., Anderson, J.L., Arshad, A., Biederman, R.W., et al. (2017) Assessing the Risks Associated with MRI in Patients with a Pacemaker or Defibrillator. The New England Journal of Medicine, 376, 755-764. https://doi.org/10.1056/NEJMoa1603265
[11]
Russo, R.J. (2013) Determining the Risks of Clinically Indicated Nonthoracic Magnetic Resonance Imaging at 1.5 T for Patients with Pacemakers and Implantable Cardioverter-Defibrillators: Rationale and Design of the MagnaSafe Registry. American Heart Journal, 165, 266-272. https://doi.org/10.1016/j.ahj.2012.12.004
[12]
Nazarian, S., Hansford, R., Rahsepar, A.A., Weltin, V., McVeigh, D., Gucuk Ipek, E., et al. (2017) Safety of Magnetic Resonance Imaging in Patients with Cardiac Devices. The New England Journal of Medicine, 377, 2555-2564. https://doi.org/10.1056/NEJMoa1604267
[13]
Samar, H., Yamrozik, J.A., Williams, R.B., Doyle, M., Shah, M., Bonnet, C.A., et al. (2017) Diagnostic Value of MRI in Patients with Implanted Pacemakers and Implantable Cardioverter-Defibrillators across a Cross Population. Does the Benefit Justify the Risk? A Proof of Concept Study. JACC: Clinical Electrophysiology, 3, 991-1002. https://doi.org/10.1016/j.jacep.2017.03.009
[14]
Guinand, A., Noble, S., Frei, A., Renard, J., Tramer, M.R. and Burri, H. (2016) Extra-Cardiac Stimulators: What Do Cardiologists Need to Know? Europace, 18, 1299-1307. https://doi.org/10.1093/europace/euv453
[15]
Indik, J.H., Gimbel, J.R., Abe, H., Alkmim-Teixeira, R., Birgersdotter-Green, U., Clarke, G.D., et al. (2017) HRS Expert Consensus Statement on Magnetic Resonance Imaging and Radiation Exposure in Patients with Cardiovascular Implantable Electronic Devices. Heart Rhythm, 14, e97-e153.
https://doi.org/10.1016/j.hrthm.2017.04.025
[16]
Centers for Medicare & Medicaid Services (2018) Proposed Decision Memo for Magnetic Resonance Imaging (MRI) (CAG-00399R4).
[17]
Roguin, A., Zviman, M.M., Meininger, G.R., Rodrigues, E.R., Dickfeld, T.M., Bluemke, D.A., Lardo, A., Berger, R.D., Calkins, H. and Halperin, H.R. (2004) Modern Pacemaker and Implantable Cardioverter/Defibrillator Systems Can Be Magnetic Resonance Imaging Safe: In Vitro and in Vivo Assessment of Safety and Function at 1.5 T. Circulation, 110, 475-482. https://doi.org/10.1161/01.CIR.0000137121.28722.33
[18]
Poh, P.G., Liew, C., Yeo, C., Chong, L.R., Tan, A. and Poh, A. (2017) Cardiovascular Implantable Electronic Devices: A Review of the Dangers and Difficulties in MR Scanning and Attempts to Improve Safety. Insights Imaging, 8, 405-418. https://doi.org/10.1007/s13244-017-0556-3
[19]
Gimbel, J.R. and Wilkoff, B.L. (2003) Artefact Mimicking Tachycardia during Magnetic Resonance Imaging in a Patient with an Implantable Loop Recorder. Heart, 89, e10. https://doi.org/10.1136/heart.89.3.e10
[20]
Gimbel, J.R., Zarghami, J., Machado, C. and Wilkoff, B.L. (2005) Safe Scanning, But Frequent Artifacts Mimicking Bradycardia and Tachycardia during Magnetic Resonance Imaging (MRI) in Patients with an Implantable Loop Recorder (ILR). Annals of Noninvasive Electrocardiology, 10, 404-408.
https://doi.org/10.1111/j.1542-474X.2005.00056.x
[21]
Hong, S.N., Rahimi, A., Kissinger, K.V., Pedrosa, I., Manning, W.J. and O’Halloran, T.D. (2010) Cardiac Magnetic Resonance Imaging and the WATCHMAN Device. Journal of the American College of Cardiology, 55, 2785. https://doi.org/10.1016/j.jacc.2010.03.027
[22]
Langman, D.A., Goldberg, I.B., Finn, J.P. and Ennis, D.B. (2011) Pacemaker Lead Tip Heating in Abandoned and Pacemaker-Attached Leads at 1.5 Tesla MRI. Journal of Magnetic Resonance Imaging, 33, 426-431.
https://doi.org/10.1002/jmri.22463
[23]
Kim, J., Hwang, J., Choi, J.H., Choi, H.I., Kim, M.S., Jung, S.H., et al. (2017) Frequency and Clinical Impact of Retained Implantable Cardioverter Defibrillator Lead Materials in Heart Transplant Recipients. PLoS ONE, 12, e0176925. https://doi.org/10.1371/journal.pone.0176925
[24]
Martin, A., Voss, J., Shannon, D., Ruygrok, P. and Lever, N. (2014) Frequency and Sequelae of Retained Implanted Cardiac Device Material Post Heart Transplantation. Pacing and Clinical Electrophysiology, 37, 242-248. https://doi.org/10.1111/pace.12274
[25]
Mattei, E., Gentili, G., Censi, F., Triventi, M. and Calcagnini, G. (2015) Impact of Capped and Uncapped Abandoned Leads on the Heating of an MR-Conditional Pacemaker Implant. Magnetic Resonance in Medicine, 73, 390-400. https://doi.org/10.1002/mrm.25106
[26]
Henderson, J.M., Tkach, J., Phillips, M., Baker, K., Shellock, F.G. and Rezai, A.R. (2005) Permanent Neurological Deficit Related to Magnetic Resonance Imaging in a Patient with Implanted Deep Brain Stimulation Electrodes for Parkinson’s Disease: Case Report. Neurosurgery, 57, E1063.
https://doi.org/10.1227/01.NEU.0000180810.16964.3E
[27]
Blaschke, F., Lacour, P., Walter, T., Wutzler, A., Huemer, M., Parwani, A., et al. (2016) Cardiovascular Magnetic Resonance Imaging in Patients with an Implantable Loop Recorder. Annals of Noninvasive Electrocardiology, 21, 319-324. https://doi.org/10.1111/anec.12333
[28]
Higgins, J.V., Gard, J.J., Sheldon, S.H., Espinosa, R.E., Wood, C.P., Felmlee, J.P., et al. (2014) Safety and Outcomes of Magnetic Resonance Imaging in Patients with Abandoned Pacemaker and Defibrillator Leads. Pacing and Clinical Electrophysiology, 37, 1284-1290. https://doi.org/10.1111/pace.12419
[29]
Strach, K., Naehle, C.P., Muhlsteffen, A., Hinz, M., Bernstein, A., Thomas, D., et al. (2010) Low-Field Magnetic Resonance Imaging: Increased Safety for Pacemaker Patients? Europace, 12, 952-960.
https://doi.org/10.1093/europace/euq081
[30]
Hartnell, G.G., Spence, L., Hughes, L.A., Cohen, M.C., Saouaf, R. and Buff, B. (1997) Safety of MR Imaging in Patients Who Have Retained Metallic Materials after Cardiac Surgery. American Journal of Roentgenology, 168, 1157-1159. https://doi.org/10.2214/ajr.168.5.9129404
[31]
Murphy, K.J., Cohan, R.H. and Ellis, J.H. (1999) MR Imaging in Patients with Epicardial Pacemaker Wires. American Journal of Roentgenology, 172, 727-728. https://doi.org/10.2214/ajr.172.3.10063869
[32]
Kanal, E. (1998) Safety of MR Imaging in Patients with Retained Epicardial Pacer Wires. American Journal of Roentgenology, 170, 213-214. https://doi.org/10.2214/ajr.170.1.9423633
[33]
de Jonge, J.C., Melis, G.I., Gebbink, T.A., de Kort, G.A. and Leijten, F.S. (2014) Safety of a Dedicated Brain MRI Protocol in Patients with a Vagus Nerve Stimulator. Epilepsia, 55, e112-e115. https://doi.org/10.1111/epi.12774
[34]
Bohning, D.E., Lomarev, M.P., Denslow, S., Nahas, Z., Shastri, A. and George, M.S. (2001) Feasibility of Vagus Nerve Stimulation-Synchronized Blood Oxygenation Level-Dependent Functional MRI. Investigative Radiology, 36, 470-479. https://doi.org/10.1097/00004424-200108000-00006
[35]
Laible, M., Mohlenbruch, M., Horstmann, S., Pfaff, J., Geis, N.A., Pleger, S., et al. (2017) Peri-Procedural Silent Cerebral Infarcts after Left Atrial Appendage Occlusion. European Journal of Neurology, 24, 53-57.
https://doi.org/10.1111/ene.13129
[36]
Mohrs, O.K., Wunderlich, N., Petersen, S.E., Pottmeyer, A. and Kauczor, H.U. (2011) Contrast-Enhanced CMR in Patients after Percutaneous Closure of the Left Atrial Appendage: A Pilot Study. Journal of Cardiovascular Magnetic Resonance, 13, 33. https://doi.org/10.1186/1532-429X-13-33
[37]
Padmanabhan, D., Kella, D.K., Mehta, R., Kapa, S., Deshmukh, A., Mulpuru, S., Jaffe, A.S., Felmlee, J.P., Jondal, M.L., Dalzell, C.M., Asirvatham, S.J., Cha, Y.M., Watson, R.E. and Friedman, P.A. (2018) Safety of Magnetic Resonance Imaging in Patients with Legacy Pacemakers and Defibrillators and Abandoned Leads. Heart Rhythm, 15, 228-233. https://doi.org/10.1016/j.hrthm.2017.10.022
[38]
Chouliaras, G., Berdoukas, V., Ladis, V., Kattamis, A., Chatziliami, A., Fragodimitri, C., et al. (2011) Impact of Magnetic Resonance Imaging on Cardiac Mortality in Thalassemia Major. Journal of Magnetic Resonance Imaging, 34, 56-59. https://doi.org/10.1002/jmri.22621
[39]
Centers for Medicare & Medicaid Services (2011) Decision Memo for Magnetic Resonance Imaging (MRI) (CAG-00399R2).
[40]
Kramer, D.B. and Kesselheim, A.S. (2017) Coverage of Magnetic Resonance Imaging for Patients with Cardiac Devices: Improving the Coverage with Evidence Development Program. JAMA Cardiology, 2, 711-712.
https://doi.org/10.1001/jamacardio.2017.1674
[41]
Acha, M.R., Keaney, J.J., Lubitz, S.A., Milan, D.J., Mansour, M., Heist, K.E., et al. (2015) Increased Perforation Risk with an MRI-Conditional Pacing Lead: A Single-Center Study. Pacing and Clinical Electrophysiology, 38, 334-342. https://doi.org/10.1111/pace.12550
[42]
Elmouchi, D.A., Rosema, S., Vanoosterhout, S.M., Khan, M., Davis, A.T., Gauri, A.J., et al. (2014) Cardiac Perforation and Lead Dislodgement after Implantation of a MR-Conditional Pacing Lead: A Single-Center Experience. Pacing and Clinical Electrophysiology, 37, 4-10. https://doi.org/10.1111/pace.12293