On Factors Affecting Subharmonic

DOI: 10.1177/0161734618812083

Keywords: ultrasound,acoustic power,subharmonic pressure estimation,optimization,hematocrit,contrast agent concentration,subharmonic imaging

Full-Text   Cite this paper   Add to My Lib

Abstract:

Subharmonic-aided pressure estimation (SHAPE) estimates hydrostatic pressure using the inverse relationship with subharmonic amplitude variations of ultrasound contrast agents (UCAs). We studied the impact of varying incident acoustic outputs (IAO), UCA concentration, and hematocrit on SHAPE. A Logiq 9 scanner with a 4C curvilinear probe (GE, Milwaukee, Wisconsin) was used with Sonazoid (GE Healthcare, Oslo, Norway) transmitting at 2.5 MHz and receiving at 1.25 MHz. An improved IAO selection algorithm provided improved correlations (r from ?0.85 to ?0.95 vs. ?0.39 to ?0.98). There was no significant change in SHAPE gradient as the pressure increased from 10 to 40 mmHg and hematocrit concentration was tripled from 1.8 to 4.5 mL/L (Δ0.00-0.01 dB, p = 0.18), and as UCA concentration was increased from 0.2 to 1.2 mL/L (Δ0.02-0.05 dB, p = 0.75). The results for the correlation between the SHAPE gradient and hematocrit values for patients (N = 100) in an ongoing clinical trial were also calculated showing a poor correlation value of 0.14. Overall, the SHAPE gradient is independent of hematocrit and UCA concentration. An improved algorithm for IAO selection will make SHAPE more accurate