-  2019 

Endovascular Aneurysm Sealing of a Collapsed and Thrombosed Aortic Stent

DOI: 10.1177/1526602818811915

Keywords: chimney graft,endograft,endovascular aneurysm repair,endovascular aneurysm sealing,iliac limb thrombosis,stent-graft collapse,stent-graft,thrombosis

Full-Text   Cite this paper   Add to My Lib

Abstract:

Purpose: To report the use of the Nellix endovascular aneurysm sealing (EVAS) system in the management of proximal stent-graft collapse associated with thrombosis following endovascular aneurysm repair (EVAR). Case Report: A 76-year-old man was admitted for proximal collapse of an aortic extension following bifurcated AFX stent-graft implantation associated with chimney grafts in both renal arteries and the superior mesenteric artery 1 month prior. Imaging identified thrombosis of the aortic stent-graft and the iliac limbs. A Nellix EVAS was placed into the AFX stent-graft to recanalize the aneurysm lumen and address the aortic thrombosis. There was no endoleak, and the renovisceral chimney stent-grafts remained patent over a follow-up of 25 months. Conclusion: While further studies are required to generalize its use, EVAS appears to be feasible in the management of aortic stent-graft collapse