全部 标题 作者
关键词 摘要

OALib Journal期刊
ISSN: 2333-9721
费用:99美元

查看量下载量

相关文章

更多...
-  2018 

Reentry Devices for Lamella Neofenestration During Endovascular Aortic Repair of Chronic Type B Aortic Dissection

DOI: 10.1177/1526602818759756

Keywords: chronic aortic dissection,endovascular repair,false lumen,fenestration,intravascular ultrasound,lamella,neofenestration,reentry devices

Full-Text   Cite this paper   Add to My Lib

Abstract:

Purpose: To present a technique that creates a neofenestration in the dissecting lamella of chronic aortic dissections using standard or intravenous ultrasound (IVUS)–guided reentry devices. Technique: To create a neofenestration, a standard or IVUS-guided reentry device is deployed in either the true or false aortic lumen while a balloon is simultaneously inflated in the opposite lumen to stabilize the lamella, provide needle counterforce, and prevent displacement of the lamella away from the reentry needle. Once the lamella is perforated and a guidewire passed across the 2 lumens, progressive dilation of the neofenestration is performed to establish persistent communication. Conclusion: The balloon-supported lamella neofenestration technique involving off-label application of two currently available reentry devices can be applied to endovascular treatment chronic aortic dissections. While the technique will definitely not become a standard of practice, it provides a safe, effective, and readily available option for endovascular management of those patients with complex chronic aortic dissections requiring bailout maneuvers or elective treatment