Scintillator Pixel Detectors for Measurement of Compton Scattering

DOI: https://doi.org/10.3390/condmat4010024

Full-Text   Cite this paper   Add to My Lib

Abstract:

The Compton scattering of gamma rays is commonly detected using two detector layers, the first for detection of the recoil electron and the second for the scattered gamma. We have assembled detector modules consisting of scintillation pixels, which are able to detect and reconstruct the Compton scattering of gammas with only one readout layer. This substantially reduces the number of electronic channels and opens the possibility to construct cost-efficient Compton scattering detectors for various applications such as medical imaging, environment monitoring, or fundamental research. A module consists of a 4 × 4 matrix of lutetium fine silicate scintillators and is read out by a matching silicon photomultiplier array. Two modules have been tested with a 22 Na source in coincidence mode, and the performance in the detection of 511 keV gamma Compton scattering has been evaluated. The results show that Compton events can be clearly distinguished with a mean energy resolution of 12.2% ± 0.7% in a module and a coincidence time resolution of 0.56 ± 0.02 ns between the two modules. View Full-Tex

Full-Text

Contact Us

service@oalib.com

QQ:3279437679