- 2017
Management of Elevated Therapeutic Impedances on Deep Brain Stimulation LeadsDOI: 10.7916/D8BR94MV Keywords: Therapeutic impedance, lead integrity impedance, deep brain stimulation, open circuit, short circuit, constant current Abstract: A 64-year-old male with a history of essential tremor with bilateral thalamic ventralis intermedius deep brain stimulation implants had elevated therapeutic impedance values despite normal lead integrity impedances and good response to stimulation
|