Experimental Evaluation of Perfluorocarbon Aerosol Generation with Two Novel Nebulizer Prototypes

DOI: 10.3390/pharmaceutics11010019

Keywords: drug delivery, nebulizer, aerosol, aerodynamic particle sizer, particle size distribution

Full-Text   Cite this paper   Add to My Lib

Abstract:

The potential of non-invasive ventilation procedures and new minimally invasive techniques has resulted in the research of alternative approaches as the aerosolization for the treatment of respiratory distress syndrome (RDS). The aim of this work was to design two nebulizer prototypes and to evaluate them studying the particle size distribution of the inhaled droplets generated with distilled water and two perfluorocarbons (PFCs). Different experiments were performed with driving pressures of 1–3 bar for each compound. An Aerodynamic Particle Sizer was used to measure the aerodynamic diameter (Da), the mass median aerodynamic diameter (MMAD) and the geometric standard deviation (GSD). The results showed that both prototypes produced heterodisperse aerosols with Da mean values in all cases below 5 μm. The initial experiments with distilled water showed MMAD values lower than 9 μm and up to 15 μm with prototype 1 and prototype 2, respectively. Regarding the PFCs, relatively uniform MMAD values close to 12 μm were achieved. The air delivery with outer lumens of prototype 1 presented more suitable mass distribution for the generation and delivery of a uniform aerosol than the two half-circular ring geometry proposed in the prototype 2