-  2016 

3种根测仪测量根管工作长度准确性的比较
An evaluation of three electronic apex locators on measuring root canal working length in vivo

DOI: 10.11778/j.jdxb.2016.06.012

Keywords: propexII,raypex5,raypex6,根管工作长度
propexII
,raypex5,raypex6,root canal working length

Full-Text   Cite this paper   Add to My Lib

Abstract:

摘要 目的:比较研究propexII根管长度测量仪、raypex5根管长度测量仪和raypex6根管长度测量仪测量根管工作长度的准确性. 方法:选牙体牙髓科就诊患者的165个根管随机分成propexII组、raypex5组和 raypex6组,每组各55个根管.患牙开髓后疏通根管,以上3组分别使用propexII、raypex5和 raypex6根管长度测量仪测量根管工作长度.然后,每个根管按照各自测量的工作长度用Protaper器械根管预备至F2,选取主牙胶尖插入根管达工作长度,拍摄X线片,通过PACS软件量取牙胶根尖到牙根根尖的距离,测量3次,取平均值,并进行单因素方差统计分析. 结果:在本实验研究中,propexII组恰填率为96.36%,raypex5组恰填率为94.54%,raypex6组恰填率为96.36%.3组牙胶根尖到牙根根尖的距离无统计学差异(P=0.593>0.05). 结论:PropexII、Raypex5 和Raypex6根测仪测定根管工作长度均具有较高的准确率,3者之间无显著差别.

References

[1]  Society of Cariology and Endodontology of Chinese Stomatological Association. National root canal therapy technical standards and quality control standards[J].West China Journal of Stomatoloy,2004.22(5):379-380.
[2]  CHAKRAVARTHY P K V. An In Vitro Comparison of Propex II Apex Locator to Standard Radiographic Method[J]. Iran Endod J, 2013, 8(3): 114-117.
[3]  段艳霞,李月恒,杨正艳,等. Propex Ⅱ根尖定位仪不同根管冲洗液下测量根管再治疗工作长度准确性的体外研究[J]. 广东医学, 2015(6): 888-890.
[4]  DUAN Y X,LI Y H,YANG Z Y,et al.In Vitro Study on Accuracy of Propex Ⅱ Apex Locator in Measuring Root Canal Re-treatment Length under Different Root Canal Irrigants[J].Guangdong Medical Journal,2015(6):888-890.
[5]  XI P,LI Q,ZHU X F,et al.Effect of blood in root canal on measurement accuracy of apex locator Raypex5[J].Beijing Journal of Stomatology,2011,19(3):147-148.
[6]  唐宇英, 陈方淳. 根尖狭窄破坏情况下两种根尖定位仪定位准确性的体外研究[J]. 激光杂志, 2011, 32(6): 68-69.
[7]  樊文明. 牙体牙髓病学[M]. 4版:北京:人民卫生出版社,2012:278-279.
[8]  FAN W M. Operative Dentistry and Endodontics[M].4 edition.Beijing: People‘s Health Publishing House,2012:278-279.
[9]  CERTOSIMO F J,MILOS M F, WALKER T.Endodontic working length determination——where does it end?[J]. Gen Dent, 1999, 47(3): 281-286.
[10]  乔玉华,曾靖华,黄世光.128例冷侧压与热牙胶垂直加压充填术的疗效比较[J].暨南大学学报(自然科学与医学版),2005.26(4):559-561.
[11]  QIAO Y H,ZENG J H,HUANG S G.Comparison of the effect of 128 cases of cold lateral pressure and vertical pressure filling with thermal gutta-percha[J].Journal of Jinan University Natural Science &Medicine Editon,2005.26(4):559-561.
[12]  曾素娟,方 颖,刘亚蕊,等.牙科数字成像系统的准确性分析[J]. 暨南大学学报(自然科学与医学版),2013.34(2):239-242.
[13]  ZENG S J,FANG Y,LIU Y W,et al.Accuracy Analysis of Dental Digital Imaging System[J]. Journal of Jinan University Natural Science &Medicine Editon,2013.34(2):239-242.
[14]  SOUZA R A. The importance of apical patency and cleaning of the apical foramen on root canal preparation[J]. Braz Dent J, 2006, 17(1): 6-9.
[15]  GONZALEZ S J A, DURAN-SINDREU F,DE N S, et al. Centring ability and apical transportation after overinstrumentation with ProTaper Universal and ProFile Vortex instruments[J]. Int Endod J, 2012, 45(6): 542-551.
[16]  NEKOOFAR M H, GHANDI M M,HAYES S J, et al. The fundamental operating principles of electronic root canal length measurement devices[J]. Int Endod J, 2006, 39(8): 595-609.
[17]  MCDONALD N J, HOVLAND E J. An evaluation of the Apex Locator Endocater[J]. JEndod, 1990,16(1):5-8.
[18]  ELAYOUT A, WEIGER R, LOST C. Frequency of overinstrumentation with an acceptable radiographic working length[J].J Endod, 2001, 27(1): 49-52.
[19]  王 凯, 周 莉,毛跃红,等. Raypex5根尖定位仪测量根管长度的准确性[J]. 上海医学, 2008, 31(12): 880-881.
[20]  中华口腔医学会牙体牙髓病专业委员会制定.全国根管治疗技术规范和质量控制标准[J]. 华西口腔医学杂志, 2004, 22(5):379-380.
[21]  JUNG I Y, YOON B H,LEE S J, et al. Comparison of the reliability of “0.5” and “APEX” mark measurements in two frequency-based electronic apex locators[J]. J Endod, 2011, 37(1): 49-52.
[22]  WANG K,ZHOU L,MAO Y H,et al.Raypex5 Apex Locator in measuring root canal length:an assessment of accuracy[J].Shanghai Med J,2008,31(12):880-881.
[23]  SINGH D, TYAGI S P,GUPTA S, et al. Comparative evaluation of adequacy of final working length after using Raypex5 or radiography: an in vivo study[J]. J Indian Soc Pedod Prev Dent, 2015, 33(3): 208-212.
[24]  王晓彦, 耿发云,李红文,等.根管冲洗液及血液对根尖定位仪测量准确性的影响[J]. 广东牙病防治, 2014, 22(10): 529-531.
[25]  WANG X Y,GENG F Y,LI H W,et al.Effects of root canal irrigation fluid and blood on measurement accuracy of apical locator[J].Journal of Dental Prevention &Treatment,2014,22(10):529-531.
[26]  忻 平, 李 青,朱晓芳,等. 根管内血液对根尖定位仪Raypex5测量准确性的影响[J]. 北京口腔医学, 2011, 19(3): 147-148.
[27]  TANG Y Y,CHEN F C.A study in vitro on accuracy of two kinds of electronic apex locators in canals with destroyed apical constrictions[J].Laser Journal,2011,32(6):68-69.
[28]  STOBER E K, DURAN-SINDREU F,MERCADE M, et al. An evaluation of root ZX and iPex apex locators: an in vivo study[J]. J Endod, 2011. 37(5): 608-610.
[29]  RICUCCI D, LANGELAND K. Apical limit of root canal instrumentation and obturation, part 2. A histological study[J]. Int Endod J, 1998, 31(6): 394-409.