IT与ISQ能反映种植体初期稳定性吗
Do IT and ISQ Reflect Primary Stability of Dental Implants?

DOI: 10.13701/j.cnki.kqyxyj.2017.02.008

Keywords: 种植体,植入扭矩,种植体稳定系数,微动,初期稳定性,
Implants
, Insertion torque, Implant stability quotient, Micromotions ,Primary stability

Full-Text   Cite this paper   Add to My Lib

Abstract:

摘要 目的:研究不同植入扭矩(insertion torque,IT)及其种植体稳定系数(implant stability quotient,ISQ )值的种植体抵抗力作用下微动的能力,分析IT与ISQ能否反映种植体初期稳定性。方法:将24颗种植体以设定扭矩植入牛肋骨中并测量相应的ISQ值。安放好基台及全瓷冠后,进行水平向及垂直向加力,记录微动150 μm时的力。结果:在每组样本中,IT和ISQ均与种植体初期稳定性不相关(P>0.05);在同一样本中,IT和ISQ可与种植体初期稳定性相关。 结论:IT与ISQ反映种植体初期稳定性是有条件的;IT与ISQ不一定能反映种植体初期稳定性

References

[1]  Takeshi T, Hideaki T, Yasunori A, et al. Primary Stability of a Hybrid Implant Compared with Tapered and Cylindrical Implants in an Ex Vivo Model [J]. Clin Implant Dent Relat Res, 2015, 17(5)∶950-956
[2]  Bayarchimeg D, Namgoong H, Kim B K, et al. Evaluation of the correlation between insertion torque and primary stability of dental implants using a block bone test [J]. J Periodontal Implant Sci, 2013, 43(1)∶30-36
[3]  韩雪莲,刘宗伟,李岩涛.种植牙即刻负重的生物力学的三维有限元分析[J].华西口腔医学杂志,2011,29(2)∶121-124
[4]  Li W, Lin D, Rungsiyakull C, et al. Finite element based bone remodeling and resonance frequency analysis for osseointegration assessment of dental implants [J].Finite Elem. Anal Des, 2011, 47(8)∶898-905
[5]  Paolo T,Teocrito C,Macro C,et al. Implant Stability Quotient(ISQ) vs dierect in vitro measurement of primary stability (micromotion):effect of bone density and insertion torque[J]. J Osteol Biomat,2010,1(1)∶141-151
[6]  Brizuelavelasco A, lvarezArenal , Gilmur F J, et al. Relationship Between Insertion Torque and Resonance Frequency Measurements, Performed by Resonance Frequency Analysis, in Micromobility of Dental Implants: An In Vitro Study[J]. Implant Dent, 2015, 24(5)∶607-611
[7]  邵现红, 王钰莹, 朱汤梅,等.种植体扭矩与初级稳定性相关性的研究[J]. 现代口腔医学杂志,2014(2)∶69-71
[8]  Marco D, Giuseppe D, Adriano P. Implants inserted with low insertion torque values for intraoral welded full-arch prosthesis: 1-year follow-up [J]. Clin Implant Dent Relat Res, 2012, 14 Suppl 1(1)∶39-45
[9]  游嘉,周乐峰,彭伟,等.面向即刻种植的种植体的初期稳定性研究[J].中国生物医学工程学报,2016,35(3)∶324-329
[10]  Atieh MA, Alsabeeha NH, Payne A G. Can resonance frequency analysis predict failure risk of immediately loaded implants [J]. Int J Prosthodont, 2012, 25(4)∶326-339
[11]  Torroella-Saura G, Mareque-Bueno J, Cabratosa-Termes J, et al. Effect of implant design in immediate loading. A randomized, controlled, split-mouth, prospective clinical trial [J]. Clin Oral Implants Res,2015,26(3)∶240-244
[12]  Winter W, Klein D, Karl M. Micromotion of Dental Implants: Basic Mechanical Considerations[J]. J Med Eng, 2013, 2013∶265412
[13]  Favre P, Henderson AD. Prediction of stemless humeral implant micromotion during upper limb activities[J].Clin Biochem, 2016, 36(1)∶46-51
[14]  Trisi P, Berardi D, Paolantonio M, et al. Primary stability, insertion torque, and bone density of conical implants with internal hexagon: is there a relationship [J]. J Craniofac Surg, 2013, 24(3)∶841-844
[15]  Gómez-Polo M, Ortega R, Gómez-Polo C, et al. Does Length, Diameter, or Bone Quality Affect Primary and Secondary Stability in Self-Tapping Dental Implants [J]. J Oral Maxillofac Surg, 2016, 74(7)∶1344-1353
[16]  Howashi M, Tsukiyama Y, Ayukawa Y, et al. Relationship between the CT Value and Cortical Bone Thickness at Implant Recipient Sites and Primary Implant Stability with Comparison of Different Implant Types [J]. Clin Implant Dent Relat Res, 2014,18(1)∶107-116
[17]  Mathieu V, Vayron R, Richard G, et al. Biomechanical determinants of the stability of dental implants: Influence of the bone-implant interface properties [J]. J Biomech, 2014, 47(1)∶3-13
[18]  Wang S, Liu G R, Hoang K C, et al. Identifiable range of osseointegration of dental implants through resonance frequency analysis [J]. Med Eng Phys, 2010, 32(10)∶1094-1106

Full-Text

Contact Us

service@oalib.com

QQ:3279437679

WhatsApp +8615387084133