Purpose: The purposes of this study were to estimate
accumulated kV X-ray imaging dose throughout dynamic tumor tracking (DTT)
irradiationby Vero4DRT system and to address an analytical skin dose
formula for well-balanced kV X-ray imaging
conditions between skin dose and image noise.Method:First, skin dose was measured using kV X-ray tube,
chamber, and water-equivalent phantoms.
Next, imaging dose for six patients in DTT treatment was computed using
log files. Subsequently, scattered dose ratiowas calculated by amount of ionization in front of flat panel
detector (FPD) for fields with size of maximum and the chamber
for 0-200 mm-thickness phantoms and
tube voltage of 60, 80, 100, 120 kV, respectively. Furthermore, image noise was computed from
FPD images. Results: The skin dose was greater by a factor of 1.4-1.6 than those in Synergy XVI system.The image noise in FPD, ?was expressed asN
= 0.045×(1/QFPDen)0.479, where QFPDen denotes amount of ionization
in front
of FPD. Then, skin dose,D (N, t, v) was formulated as (0.045/N)(1/0.479)/QFPDen/mAs
(t, v) ×D/mAs (v),whereQFPDen/mAs
(t, v) andD/mAs
(v) denote amount of ionization in front of FPDand skin dose per mAs, respectively. Using the formulae, it
References
[1]
Mukumoto, N., Nakamura, M., Sawada, A., Takahashi, K., Mizowaki, T., Kokubo, M. and Hiraoka, M. (2011) Geometric Accuracy of the X-Ray Image-Based Dynamic Tracking for a Four-Dimensional Image-Guided Radiotherapy System with Gimbals Mechanism of MHI-TM2000 (Vero). 2011 Joint AAPM/COMP Meeting, Vancouver, 31 July-4 August 2011, 3477. https://doi.org/10.1118/1.3611919
[2]
Mukumoto, N., Nakamura, M., Sawada, A., Takahashi, K., Miyabe, Y., Takayama, K., et al. (2010) Verification of the Tracking Accuracy for a Four-Dimensional Image-Guided Radiotherapy System with a Gimbaled X-Ray Head, MHI-TM2000. Proceedings of 10th Asia-Oceania Congress of Medial Physics, Taipei, 15-17 October 2010, 88.
[3]
Depuydt, T., Poels, K., Van Gompel, G., Buls, N., Gevaert, T., Duchateau, M., et al. (2011) Low Dose Based Detection of Implanted Marker Position on the VERO System for Real-Time Tumor Tracking. 2011 Joint AAPM/COMP Meeting, Vancouver, 31 July-4 August 2011, 3605.
[4]
Cardella, J., Faulkner, K., Hoperwell, J., Nakamura, H., Rehani, M., Rosenstein, M., et al. (2000) Avoidance of Radiation Injuries from Medical Interventional Procedures. ICRP Publication 85, Annals of the ICRP. SAGE Publications Ltd., Thousand Oaks, 30.
[5]
Murphy, M.J., Balter, J., Balter, S., BenComo, J.A., Das, I.J., Jiang, S.B., et al. (2007) The Management of Imaging Dose during Image-Guided Radiotherapy: Report of AAPM Task Group 75. Medical Physics, 34, 4041-4063. https://doi.org/10.1118/1.2775667
[6]
Siedband, M., Baiter, B., Morgan, T., Bratemann, L., Properzio, W., Britt, W., et al. (1977) Basic Quality Control in Diagnostic Radiology. AAPM Report No. 4. https://aapm.org/pubs/reports/rpt_04.pdf
[7]
Balter, S., Fletcher, D.W., Kuan, H.M., Miller, D., Richter, D., Seissl, H., et al. (2002) Techniques to Estimate Radiation Dose to Skin during Fluoroscopically Guided Procedures. Skin Dose Measurements AAPM, July 2002. http://citeseerx.ist.psu.edu/viewdoc/download?doi=10.1.1.483.6023&rep=rep1&type=pdf
[8]
Steven, W.S. (1997) The Scientist and Engineer’s Guide to Digital Signal Processing. California Technical Publishing, San Diego.
[9]
Kamino, Y., Miura, S., Kokubo, M., Yamashita, I., Hirai, E., Hiraoka, M. and Ishikawa, J. (2007) Development of an Ultrasmall C-Band Linear Accelerator Guide for a Four-Dimensional Image-Guided Radiotherapy System with a Gimbaled X-Ray Head. Medical Physics, 34, 1797-1808. https://doi.org/10.1118/1.2723878
[10]
Selzer, S.M. and Hubbell, J.H. (1995) Table and Graphs of Photon Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients for Photon Energies 1 keV to 20MeV for Elements Z = 1 to 92 and Some Dosimetric Materials. JSRT Japanese Society of Radiological Technology Press, Kyoto.
[11]
Islam, M.K., Purdie, T.G., Norrlinger, B.D., Alasti, H., Moseley, D.J., Sharpe, M.B., et al. (2006) Patient Dose from Kilovoltage Cone Beam Computed Tomography Imaging in Radiation Therapy. Medical Physics, 33, 1573-1582. https://doi.org/10.1118/1.2198169
[12]
Nakai, T., Itou, T., Kubo, K., Sueoka, M., Tanabe, H., Furukawa, H. and Kokubo, M. (2011) Measurement of Imaging Dose toward Pursuit Irradiation; A Preliminary Study. 67th Japanese Society of Radiological Technology, Yokohama, 7-10 April 2011, 257.
[13]
Hirai, E., Tsukuda, K., Kamino, Y., Miura, S., Takayama, K. and Aoi, T. (2009) MHI’s First Radiotherapy Machine for Overseas to Begin Treatment at Brussels University Hospital (UZ Brussel). Mitsubishi Heavy Industries Technical Review, 46, 29-32.