Investigation of Well-Balanced kV X-Ray Imaging Conditions between Skin Dose and Image Noise for Dynamic Tumor Tracking Irradiation

DOI: 10.4236/ijmpcero.2017.64037, PP. 410-420

Keywords: Dynamic Tumor Tracking Irradiation, Skin Dose, Vero4DRT

Full-Text   Cite this paper   Add to My Lib

Abstract:

Purpose: The purposes of this study were to estimate accumulated kV X-ray imaging dose throughout dynamic tumor tracking (DTT) irradiation by Vero 4DRT system and to address an analytical skin dose formula for well-balanced kV X-ray imaging conditions between skin dose and image noise. Method: First, skin dose was measured using kV X-ray tube, chamber, and water-equivalent phantoms. Next, imaging dose for six patients in DTT treatment was computed using log files. Subsequently, scattered dose ratio was calculated by amount of ionization in front of flat panel detector (FPD) for fields with size of maximum and the chamber for 0 - 200 mm-thickness phantoms and tube voltage of 60, 80, 100, 120 kV, respectively. Furthermore, image noise was computed from FPD images. Results: The skin dose was greater by a factor of 1.4 - 1.6 than those in Synergy XVI system. The image noise in FPD, ?was expressed as N = 0.045×(1/QFPDen)0.479, where QFPDen denotes amount of ionization in front of FPD. Then, skin dose, D (N, t, v) was formulated as (0.045/N)(1/0.479)/QFPDen/mAs (t, v) ×D/mAs (v), where QFPDen/mAs (t, v) and D/mAs (v) denote amount of ionization in front of FPD and skin dose per mAs, respectively. Using the formulae, it

References

[1]  Mukumoto, N., Nakamura, M., Sawada, A., Takahashi, K., Mizowaki, T., Kokubo, M. and Hiraoka, M. (2011) Geometric Accuracy of the X-Ray Image-Based Dynamic Tracking for a Four-Dimensional Image-Guided Radiotherapy System with Gimbals Mechanism of MHI-TM2000 (Vero). 2011 Joint AAPM/COMP Meeting, Vancouver, 31 July-4 August 2011, 3477.
https://doi.org/10.1118/1.3611919
[2]  Mukumoto, N., Nakamura, M., Sawada, A., Takahashi, K., Miyabe, Y., Takayama, K., et al. (2010) Verification of the Tracking Accuracy for a Four-Dimensional Image-Guided Radiotherapy System with a Gimbaled X-Ray Head, MHI-TM2000. Proceedings of 10th Asia-Oceania Congress of Medial Physics, Taipei, 15-17 October 2010, 88.
[3]  Depuydt, T., Poels, K., Van Gompel, G., Buls, N., Gevaert, T., Duchateau, M., et al. (2011) Low Dose Based Detection of Implanted Marker Position on the VERO System for Real-Time Tumor Tracking. 2011 Joint AAPM/COMP Meeting, Vancouver, 31 July-4 August 2011, 3605.
[4]  Cardella, J., Faulkner, K., Hoperwell, J., Nakamura, H., Rehani, M., Rosenstein, M., et al. (2000) Avoidance of Radiation Injuries from Medical Interventional Procedures. ICRP Publication 85, Annals of the ICRP. SAGE Publications Ltd., Thousand Oaks, 30.
[5]  Murphy, M.J., Balter, J., Balter, S., BenComo, J.A., Das, I.J., Jiang, S.B., et al. (2007) The Management of Imaging Dose during Image-Guided Radiotherapy: Report of AAPM Task Group 75. Medical Physics, 34, 4041-4063.
https://doi.org/10.1118/1.2775667
[6]  Siedband, M., Baiter, B., Morgan, T., Bratemann, L., Properzio, W., Britt, W., et al. (1977) Basic Quality Control in Diagnostic Radiology. AAPM Report No. 4.
https://aapm.org/pubs/reports/rpt_04.pdf
[7]  Balter, S., Fletcher, D.W., Kuan, H.M., Miller, D., Richter, D., Seissl, H., et al. (2002) Techniques to Estimate Radiation Dose to Skin during Fluoroscopically Guided Procedures. Skin Dose Measurements AAPM, July 2002.
http://citeseerx.ist.psu.edu/viewdoc/download?doi=10.1.1.483.6023&rep=rep1&type=pdf
[8]  Steven, W.S. (1997) The Scientist and Engineer’s Guide to Digital Signal Processing. California Technical Publishing, San Diego.
[9]  Kamino, Y., Miura, S., Kokubo, M., Yamashita, I., Hirai, E., Hiraoka, M. and Ishikawa, J. (2007) Development of an Ultrasmall C-Band Linear Accelerator Guide for a Four-Dimensional Image-Guided Radiotherapy System with a Gimbaled X-Ray Head. Medical Physics, 34, 1797-1808.
https://doi.org/10.1118/1.2723878
[10]  Selzer, S.M. and Hubbell, J.H. (1995) Table and Graphs of Photon Mass Attenuation Coefficients and Mass Energy-Absorption Coefficients for Photon Energies 1 keV to 20MeV for Elements Z = 1 to 92 and Some Dosimetric Materials. JSRT Japanese Society of Radiological Technology Press, Kyoto.
[11]  Islam, M.K., Purdie, T.G., Norrlinger, B.D., Alasti, H., Moseley, D.J., Sharpe, M.B., et al. (2006) Patient Dose from Kilovoltage Cone Beam Computed Tomography Imaging in Radiation Therapy. Medical Physics, 33, 1573-1582.
https://doi.org/10.1118/1.2198169
[12]  Nakai, T., Itou, T., Kubo, K., Sueoka, M., Tanabe, H., Furukawa, H. and Kokubo, M. (2011) Measurement of Imaging Dose toward Pursuit Irradiation; A Preliminary Study. 67th Japanese Society of Radiological Technology, Yokohama, 7-10 April 2011, 257.
[13]  Hirai, E., Tsukuda, K., Kamino, Y., Miura, S., Takayama, K. and Aoi, T. (2009) MHI’s First Radiotherapy Machine for Overseas to Begin Treatment at Brussels University Hospital (UZ Brussel). Mitsubishi Heavy Industries Technical Review, 46, 29-32.

Full-Text

Contact Us

service@oalib.com

QQ:3279437679

WhatsApp +8615387084133