Objective: The
standard practice in fractionated radiation of using of visual alignment of
tattoos with weekly portal imaging for radiation setup can miss setup errors. We
previously reported on the feasibility of using an automated infrared-guided
positioning system (iGPS) for daily radiation setup. In this study, we
prospectively evaluate whether use of iGPS can improve daily setup errors for
patients with thoracic malignancies. Methods: Multiple external infrared
markers were placed on patients undergoing thoracic radiation at the time of
simulation. Patients were immobilized using vacuum immobilization bag and
wing-board. Patients were aligned for treatment using only the iGPS system. Daily
portal images were then taken, and shifts in patient position were recorded. Differences
between isocenter position using iGPS versus daily portal imaging were
calculated. Results: Data were collected for 698 treatment sessions for
27 patients. We found that in 94.0%, 96.4%, and 93.7% of treatment sessions,
isocenter position differed between iGPS and daily portal imaging by ≤5mm in the left/right(L/R), anterior/posterior(A/P),
and superior/inferior(S/I) directions, respectively. Isocenter position
differed by 5-10mm in
5.7% (L/R), 3.6% (A/P), and 5.9% (S/I) of treatment sessions, and by >10mm in 0.3% (L/R), 0.0% (A/P), and 0.4% (S/I) of
treatment sessions. Three-dimensional shifts were also calculated, with
differences in isocenter position as follows: 94.9% ≤5mm, 5.1% 5-10mm,
References
[1]
Erridge, S.C., Seppenwoolde, Y., Muller, S.H., et al. (2003) Portal Imaging to Assess Set-Up Errors, Tumor Motion and Tumor Shrinkage during Conformal Radio-therapy of Non-Small Cell Lung Cancer. Radiotherapy and Oncology, 66, 75-85. https://doi.org/10.1016/S0167-8140(02)00287-6
[2]
Seppenwoolde, Y., Shirato, H., Kitamura, K., et al. (2002) Precise and Real-Time Measurement of 3D Tumor Motion in Lung Due to Breathing and Heartbeat, Measured during Radiotherapy. International Journal of Radiation Oncology Biology Physics, 53, 822-834. https://doi.org/10.1016/S0360-3016(02)02803-1
[3]
Cho, B.C., Van Herk, M., Mijnheer, B.J., et al. (2002) The Effect of Set-Up Uncertainties, Contour Changes, and Tissue Inhomogeneities on Target Dose-Volume Histograms. Medical Physics, 29, 2305-2318. https://doi.org/10.1118/1.1508800
[4]
Engelsman, M., Damen, E.M., De Jaeger, K., et al. (2001) The Effect of Breathing and Set-Up Errors on the Cumulative Dose to a Lung Tumor. Radiotherapy and Oncology, 60, 95-105. https://doi.org/10.1016/S0167-8140(01)00349-8
[5]
Schwarz, M., Van der Geer, J., Van Herk, M., et al. (2006) Impact of Geometrical Uncertainties On 3D CRT and IMRT Dose Distributions for Lung Cancer Treatment. International Journal of Radiation Oncology Biology Physics, 65, 1260-1269. https://doi.org/10.1016/j.ijrobp.2006.03.035
[6]
Hurkmans, C.W., Remeijer, P., Lebesque, J.V., et al. (2001) Set-Up Verification Using Portal Imaging; Review of Current Clinical Practice. Radiotherapy and Oncology, 58, 105-120. https://doi.org/10.1016/S0167-8140(00)00260-7
[7]
Kutcher, G.J., Coia, L., Gillin, M., et al. (1994) Comprehensive QA for Radiation Oncology: Report of AAPM Radiation Therapy Committee Task Group 40. Medical Physics, 21, 581-618. https://doi.org/10.1118/1.597316
[8]
Rodrigus, P., Van den Weyngaert, D. and Van den Bogaert, W. (1987) The Value of Treatment Portal Films in Radiotherapy for Bronchial Carcinoma. Radiotherapy and Oncology, 9, 27-31. https://doi.org/10.1016/S0167-8140(87)80216-5
[9]
Dawson, L.A. and Jaffray, D.A. (2007) Advances in Image-Guided Radiation Therapy. Journal of Clinical Oncology, 25, 938-946. https://doi.org/10.1200/JCO.2006.09.9515
[10]
Brahme, A., Nyman, P. and Skatt, B. (2008) 4D Laser Camera for Accurate Patient Positioning, Collision Avoidance, Image Fusion and Adaptive Approaches during Diagnostic and Therapeutic Procedures. Medical Physics, 35, 1670-1681. https://doi.org/10.1118/1.2889720
[11]
Lyatskaya, Y., James, S., Killoran, J.H., et al. (2008) Infrared-Guided Patient Setup for Lung Cancer Patients. International Journal of Radiation Oncology Biology Physics, 71, 1124-1133. https://doi.org/10.1016/j.ijrobp.2007.11.050
[12]
Meeks, S.L., Tome, W.A., Willoughby, T.R., et al. (2005) Optically Guided Patient Positioning Techniques. Seminars in Radiation Oncology, 15, 192-201. https://doi.org/10.1016/j.semradonc.2005.01.004
[13]
Soete, G., Van de Steene, J., Verellen, D., et al. (2002) Initial Clinical Experience with Infrared-Reflecting Skin Markers in the Positioning of Patients Treated by Conformal Radiotherapy for Prostate Cancer. International Journal of Radiation Oncology Biology Physics, 52, 694-698. https://doi.org/10.1016/S0360-3016(01)02642-6
[14]
Baroni, G., Ferrigno, G. and Pedotti, A. (1998) Implementation and Application of Real-Time Motion Analysis Based on Passive Markers. Medical & Biological Engineering & Computing, 36, 693-703. https://doi.org/10.1007/BF02518871
[15]
Herfarth, K.K., Debus, J., Lohr, F., et al. (2000) Extracranial Stereotactic Radiation Therapy: Set-Up Accuracy of Patients Treated for Liver Metastases. International Journal of Radiation Oncology Biology Physics, 46, 329-335. https://doi.org/10.1016/S0360-3016(99)00413-7
[16]
Lyatskaya, Y., Lu, H.M. and Chin, L. (2006) Performance and Characteristics of an IR Localizing System for Radiation Therapy. Journal of Applied Clinical Medical Physics, 7, 18-37. https://doi.org/10.1120/jacmp.v7i2.2190
[17]
Kudchadker, R.J., Lee, A.K., Yu, Z.H., et al. (2009) Effectiveness of Using Fewer Implanted Fiducial Markers for Prostate Target Alignment. International Journal of Radiation Oncology Biology Physics, 74, 1283-1289. https://doi.org/10.1016/j.ijrobp.2009.02.033
[18]
Halperin, R., Roa, W., Field, M., et al. (1999) Setup Reproducibility in Radiation Therapy for Lung Cancer: A Comparison between T-Bar and Expanded Foam Immobilization Devices. International Journal of Radiation Oncology Biology Physics, 43, 211-216. https://doi.org/10.1016/S0360-3016(98)00354-X
[19]
Van Herk, M. (2004) Errors and Margins in Radiotherapy. Seminars in Radiation Oncology, 14, 52-64. https://doi.org/10.1053/j.semradonc.2003.10.003
[20]
Padilla, L., Kang, H., Washington, M., et al. (2014) Assessment of Interfractional Variation of the Breast Surface Following Conventional Patient Positioning for Whole-Breast Radiotherapy. Journal of Applied Clinical Medical Physics, 15, 4921. https://doi.org/10.1120/jacmp.v15i5.4921
[21]
Shah, A.P., Dvorak, T., Curry, M.S., et al. (2013) Clinical Evaluation of Inter-fractional Variations for Whole Breast Radiotherapy Using 3-Dimensional Surface Imaging. Practical Radiation Oncology, 3, 16-25. https://doi.org/10.1016/j.prro.2012.03.002
[22]
Alderliesten, T., Sonke, J.J., Betgen, A., et al. (2013) Accuracy Evaluation of a 3-Dimensional Surface Imaging System for Guidance in Deep-Inspiration Breath-Hold Radiation Therapy. International Journal of Radiation Oncology Biology Physics, 85, 536-542. https://doi.org/10.1016/j.ijrobp.2012.04.004
[23]
Tang, X., Zagar, T.M., Bair, E., et al. (2014) Clinical Experience with 3-Dimensional Surface Matching-Based Deep Inspiration Breath Hold for Left-Sided Breast Cancer Radiation Therapy. Practical Radiation Oncology, 4, e151-e158. https://doi.org/10.1016/j.prro.2013.05.004
[24]
Gierga, D.P., Turcotte, J.C., Sharp, G.C., et al. (2012) A Voluntary Breath-Hold Treatment Technique for the Left Breast with Unfavorable Cardiac Anatomy Using Surface Imaging. International Journal of Radiation Oncology Biology Physics, 84, e663-e668. https://doi.org/10.1016/j.ijrobp.2012.07.2379
[25]
Smadi, M.P., Esmaili, T.A. and Nankali, S. (2016) Investigation of the Optimum Location of External Markers for Patient Setup Accuracy Enhancement at External Beam Radiotherapy. Journal of Applied Clinical Medical Physics, 17, 6265.