Quality control of Gamma Camera with SPECT System is
highly valuable for assurance performance characteristic. We report the
performance characteristic of gamma camera by intrinsic calibration and
verification measurement. The study has been done using the data from Siemens
Symbia S Series gamma camera by using a point source 99mTc
at the Institute of Nuclear Medicine & Allied Sciences (INMAS), Khulna,
Bangladesh. From intrinsic calibration and verification flood series, the
integral uniformity for the central field of view (CFOV) has been found in
between 4.01% and 2.88% and for the useful field of view (UFOV) has been in
between 4.77% and 4.30%. The differential uniformity for the CFOV has been in
between 1.53% and 2.04% and for the UFOV has been in between 2.32% and 2.77%.
According to Operating Instruction Symbia System S Series manual, uniformity
can compensate for values exceeding 10%, however while integral uniformity
exceed 7%, have to contract Siemens customer service representative. In conclusion, these results show
that the intrinsic uniformity of the gamma camera under this condition is
within an acceptable range; thus the gamma camera working in INMAS is performed
well.
References
[1]
TiP-TVTM Training in Partnership Program Supplement for Imaging Professionals (2004) NM/PET: Advances in Gamma Camera Technology.
[2]
International Atomic Energy Agency (2009) Quality Assurance for SPECT Systems. IAEA Human Health Series No. 6. International Atomic Energy Agency, Vienna.
[3]
Gar-Elnabi, M.E.M., Ali, W.M., Omer, M.A.A., Sam, A.K. and Edam, G.A. (2015) Development and Assessment of Quality Control Phantom for Linearity and Uniformity. Open Journal of Radiology, 5, 59-65. https://doi.org/10.4236/ojrad.2015.52010
[4]
National Electrical Manufacturers Association (NEMA) (2001) Performance Measurements of Scintillation Cameras. Global Engineering Document. NEMA Standards Publication NU, Rosslyn, Virginia.
[5]
Ejeh, J.E., Adedapo, K.S., Akinlade, B.I. and Osifo, B.O.A. (2011) Gamma Camera Intrinsic Uniformity in an Unstable Power Supply Environment. Hellenic Journal of Nuclear Medicine, 14, 146-148. http://nuclmed.web.auth.gr/magazine/eng/may11/7.pdf
[6]
Graham, S.L. (1995) Quality Control for SPECT Systems, the AAPM/RSNA Physics Tutorials for Residents. Imaging and Therapeutic Technology, 15, 1471-1481.
[7]
Nichols K.J., et al. (2007) ASNC Imaging Guidelines for Nuclear Cardiology Procedures. Instrumentation Quality Assurance and Performance. Journal of Nuclear Cardiology, 14, 61-78. https://doi.org/10.1016/j.nuclcard.2007.09.024
[8]
Graham, S.L., Levin, C.S. and Muehllehner, G. (2003) Anger Scintillation Camera. In: Sandler, M.P., Coleman, R.E., Patton, J.A., ThWackers, F.J. and Gottschalk, A., Eds., Diagnostic Nuclear Medicine, Lippincott Williams and Wilkins Philadelphia, Philadelphia, 31-42.
[9]
Cherry, S.R., Sorenson, J.A. and Phelps, M.E. (2003) Physics in Nuclear Medicine. Elsevier Science, Philadelphia, 250-262.
[10]
O’Connor, M.K. and Vermeersch, C. (1991) Critical Examination of the Uniformity Requirements for Single Photon Emission Computed Tomography. Medical Physics, 18, 190-197. https://doi.org/10.1118/1.596706
[11]
Young, K.C., Kouris, K., Awdeh, M. and Abdel-Dayem, H.M. (1990) Reproducibility and Action Levels for Gamma Camera Uniformity. Nuclear Medicine Communications, 11, 95-101. https://doi.org/10.1097/00006231-199002000-00003
[12]
Syngo MI Applications (2009) Operating Instruction Symbia System S Series. https://www.usa.siemens.com/healthcare