基于量子级联激光器光腔衰荡光谱技术的痕量氨气检测

Keywords: 光腔衰荡光谱,量子级联激光器,氨气,线宽

Full-Text   Cite this paper   Add to My Lib

Abstract:

采用基于红外脉冲量子级联激光器的高灵敏光腔衰荡光谱(Cavityring-downspectroscopy)技术建立了氨气在1027cm-1(9.7μm)附近的吸收光谱痕量检测实验装置.使用该装置对人体口腔呼出气体以及超净实验室环境中氨气成份进行了测量和分析,氨气的检测灵敏度达到10ppb.模拟分析了激光器光谱线宽对测量结果和系统检测灵敏度的影响,随着激光器光谱线宽的增加,测量结果偏低,系统检测灵敏度下降.为了减小光源线宽对测量结果的影响,通过修正曲线对测量数据进行修正,得到了较好的结果.

References

[1]  【1】OKEEFE A, DEACON D A G. Cavity ring-down optical spectrometer for absorption measurement using pulsed laser sources[J]. Rev. Sci. Instrum.,1988,59(12):25442561.
[2]  【2】ROMANNI D, KACHANOV A A, SADEGHI N, et al. CW cavity ring down spectroscopy[J]. Chem. Phys. Lett.,1997,264:316322.
[3]  【3】DEBECKER I, MOHAMED A K, ROMANINI D. High-speed cavity ringdown spectroscopy with increased spectral resolution by simultaneous laser and cavity tuning[J]. Opt. Express,2005,13(8):29062915.
[4]  【4】SCHOCKER A, HOINGHAUS K K, BROCKHINKE A. Quantitative determination of combustion intermediates with cavity ring-down spectroscopy: systematic study in propene flames near the soot-formation limit[J].Appl. Opt.,2005,44(31):66606672.
[5]  【5】RICHMAN B A, KACHANOV A A, PALDUS B A. Novel detection of aerosols: combined cavity ring-down and fluorescence spectroscopy[J]. Opt. Express,2005,13(9):33763387.
[6]  【6】WANG Chu-Ji, SAHAY P. Breath Analysis Using Laser Spectroscopic Technique: Breath Biomarkers, Spectral Fingerints, and Detection Limits[J]. Sensors,2009,9:82308262.
[7]  【7】NARASIMHAN L R, GOODMAN W, PATEL C K N. Correlation of breath ammonia with blood urea nitrogen and creatinine during hemodialysis[J]. PANS.2001,98(8):46174621.
[8]  【8】ROTHMAN S, JACQUEMART D, BARBE A, et al. The HITRAN 2004 molecular spectroscopic database[J]. J. Quantitative Spectroscopy& Radiative Transfer.2005,96:139204.
[9]  【9】ZHU Fang-Qiang, JIANG Kai-Li, HE Qiu-Hong, et al. A Study of the Laser Linewidth Effects in Cavity Ring-down Absorption Measurements[J]. Chinese Journal of Chemical Physics(朱方强,姜开利,何秋荣等.腔衰荡吸收测量中光源线宽影响研究.化学物理学报),1998,11(1):2630.
[10]  【10】ZALICKI P, ZARE R N. Cavity ring-down spectroscopy for quantitative absorption measurements[J].J. Chem. Phys.,1995,102(7):27082717.
[11]  【11】PUSHKARSKY M B, WEBBER M E, PATEL C K N. Ultra-sensitive ambient ammonia detection using CO2-laser-based photoacoustic Spectroscopy[J]. Appl. Phys.B,2003,77:381385