新型光电子源离子迁移谱在检测四氯化碳中的应用

, PP. 138-145

Keywords: 光谱学,离子迁移谱,光电效应,光电子源,四氯化碳

Full-Text   Cite this paper   Add to My Lib

Abstract:

紫外光电离源是离子迁移谱中最重要的电离源之一,常被用来电离有机物,产生正离子。利用紫外光照射金属表面产生电子作为离子源,建立了一套光电子源离子迁移谱仪,由于卤化物具有较强的吸附电子能力,该装置可以应用于卤化物的检测。使用CCl4样品,得到了该装置的离子峰强度、半峰宽、分辨率随迁移管电场、离子门脉宽、离子门电压的变化关系,并进行了简单的理论分析。兼顾离子峰强度与分辨率,得到了优化的实验参数,在该参数下测量了CCl4的检测限达到4ppb,线性范围跨越2个数量级。

References

[1]  Ewing R G, Atkinson D A, Eiceman G A, et al. A critical review of ion mobility spectrometry for the detection of explosives and explosive related compounds [J]. Talanta, 2001, 54(3): 515-529.
[2]  Lawrence A H. Characterization of benzodiazepine drugs by ion mobility spectrometry [J]. Anal. Chem., 1989, 61(4): 343-349.
[3]  Bell S E, Ewing R G, Eiceman G A, et al. Atmospheric pressure chemical ionization of alkanes, alkenes, and cycloalkanes [J]. J. Am. Soc. Mass Spectrom., 1994, 5(3): 177-185.
[4]  Karpas Z, Pollevoy Y, Melloul S. Determination of bromine in air by ion mobility spectrometry [J]. Anal. Chim. Acta., 1991, 249(2): 503-507.
[5]  Tang X, Bruce J E, Hill H H. Characterizing electrospray ionization using atmospheric pressure ion mobility spectrometry [J]. Anal. Chem., 2006, 78(22): 7751-7760.
[6]  Khayamian T, Jafari M T. Design for electrospray ionization-ion mobility spectrometry [J]. Anal. Chem., 2007, 79(8): 3199-3205.
[7]  Wu C, Siems W F, Hill H H. Secondary electrospray ionization ion mobility spectrometry/mass spectrometry of illicit drugs [J]. Anal. Chem., 2000, 72(2): 396-403.
[8]  Tam M, Hill H H. Secondary electrospray ionization-ion mobility spectrometry for explosive vapor detection [J]. Anal. Chem., 2004, 76(10): 2741-2747.
[9]  Lubinan D M, Kronlck M N. Multiwavelength-selective ionization of organic compounds in an ion mobility spectrometer [J]. Anal. Chem., 1983, 55(6): 867-873.
[10]  Han Haiyan, Huang Guodong, Jing Sunping, et al. Determination of alcohol compounds using corona discharge ion mobility spectrometry [J]. J. Environ. Sci-China, 2007, 19(6): 751-755.
[11]  Han Haiyan, Wang Hongmei, Jiang Haie, et al. Corona Discharge Ion Mobility Spectrometry of Ten Alcohols [J]. Chinese J. Chem. Phys., 2009, 22(6): 605-610.
[12]  Sielemann S, Baumbach J I, Schmidt H, et al. Quantitative analysis of benzene, toluene, and m-xylene with the use of a UV-ion mobility spectrometer [J]. Field Anal. Chem. Technol., 2000, 4(4): 157-169.
[13]  Li Hu, Niu Wenqi, Wang Hongmei, et al. parameters and performance analysis of photoionization ion mobility spectrometry (UV-IMS) [J]. Chinese Journal of Quantum Electronics(量子电子学报), 2012, 29(1): 8-14 (in Chinese).
[14]  Baumbach J I, Sielemann S, Xie Z, et al. Detection of the Gasoline Components Methyl tert-Butyl Ether, Benzene, Toluene, and m-Xylene Using Ion Mobility Spectrometers with a Radioactive and UV Ionization Source [J]. Anal. Chem., 2003, 75(6): 1483-1490.
[15]  Sielemann S, Baumbach J I, Schmidt H, et al. Detection of alcohols using UV-ion mobility spetrometers [J]. Anal. Chim. Acta., 2001, 431(2): 293-301.
[16]  Deng Qiaolin, Li Hu, Han Haiyan, et al. Aromatic Compounds Separation and Detection by Tandem Gas Chromatography and Photoionization Ion Mobility Spectrometry [J]. Chinese Journal of Quantum Electronics(量子电子学报), 2011, 28(1): 12-18 (in Chinese).
[17]  Meier R R. Ultraviolet spectroscopy and remote sensing of the upper atmosphere [J]. Space Sci. Rev., 1991, 58(1-2): 1-185.
[18]  Tabrizchi M, Abedi A. A novel use of negative ion mobility spectrometry for measuring electron attachment rates [J], J. Phys. Chem. A, 2004, 108 (30): 6319-6324.
[19]  http://www.cheric.org/research/kdb/hcprop/cmpsrch.php [OL].
[20]  Kanu A B, Gribb M M, Hill H H. Predicting optimal resolving power for ambient pressure ion mobility spectrometry [J]. Anal. Chem., 2008, 80(17): 6610-6619.
[21]  Rokushika S, Hatano H, Baim M A, et al. Resolution measurement for ion mobility spectrometry [J]. Anal. Chem., 1985, 57(9): 1902-1907.
[22]  Siems W F, Wu C, Tarver E E, et al. Measuring the Resolving Power of Ion Mobility Spectrometers [J]. Anal. Chem., 1994, 66(23): 4195-4201.
[23]  Spangler G E. Expanded theory for the resolving power of a linear ion mobility spectrometer [J]. Int. J. Mass Spectrom., 2002, 220(3): 399-418.