指脉脉搏波自动识别及自动采集控制方法

Keywords: 脉搏波,自动判波,自动采集

Full-Text   Cite this paper   Add to My Lib

Abstract:

提出一种指脉脉搏波自动识别及采集自动控制方法.以手指伸入传感器后脉搏波采样值的变化自动启动采集;通过对采集信号的模拟和数字滤波以及基线漂移的滤除,得到有效的脉搏波波形;利用差分阈值法提取脉搏波的起始点得到单个的脉搏波;依据单个脉搏波的异常类型对脉搏波进行自动识别;当连续3屏的脉搏波特征参数平均值绝对误差小于0.005时,自动停止脉搏波采集.实际测量实验表明:该方法能对脉搏波有效性进行判断和对脉搏波检测进行自动控制,可应用在指脉脉搏波检测中.

References

[1]  刘静纨,陈建慧,罗志昌.人体微循环弹性腔模型与模型参数估计[J].北京生物医学工程,2001,20(2):102-105.LIU Jing-wan,CHEN Jian-hui,LUO Zhi-chang.Anelastic chamber model of human microcirculation systemand its parameter estimation[J].Beijing BiomedicalEngineering,2001,20(2):102-105.(in Chinese)
[2]  张俊利,蔺嫦燕.容积脉搏波的检测方法及其在评价心血管功能方面的应用[J].北京生物医学工程,2007,26(2):220-224.ZHANG Jun-li,LIN Chang-yan.Measurement of volumepulse wave and the application in evaluation of thecardiovascular function[J].Beijing BiomedicalEngineering,2007,26(2):220-224.(in Chinese)
[3]  王秀章,陈声权.脉搏式血氧饱和度检测方法的研究[J].中国医学物理学杂志,1995,12(1):58-62.WANG Xiu-zhang,CHEN Sheng-quan.Study onmeasuring method of pulse oxygen saturation[J].ChineseJournal of Medical Physics,1995,12(1):58-62.(inChinese)
[4]  罗志昌,张松,杨益民.脉搏波的工程分析与临床应用[M].北京:科学出版社,2006:173-182.
[5]  罗晓民,李晓慧,张冰.基于指端容积血流脉动波形图的血流动力学分析[J].医学研究杂志,2008,37(8):57-59.LUO Xiao-min,LI Xiao-hui,ZHANG Bing.Hemodynamicanalysis of peripheral blood volume pulsation detectable onfinger tip with photoplethsmography[J].Journal ofMedical Research,2008,37(8):57-59.(in Chinese)
[6]  MURTHY V S,RAMAMOORTHY S,SRINIVASAN N,et al.Analysis of photoplethysmography signals ofcardiovascular patients[C]∥Proceedings of the 23rdAnnual EMBS International Conference,Istanbul,October25-28,2001:2204-2207.
[7]  OMBONI S,PARATI G,CASTIGLIONI P,et al.Estimation of blood pressure variability from 24-hourambulatory finger blood pressure[J].Hypertension,1998,32(1):52-58.
[8]  RANG S,LAPIEDRA B D,VAN MONTFRANS G A,et al.Modelflow:a new method for noninvasive assessmentof cardiac output in pregnant women[J].AmericanJournal of Obstetrics and Gynecology,2007,196(3):235-237.
[9]  虞启琏,陈卓,周鑫玲,等.智能化外周血液循环监测仪[J].中国医疗器械杂志,1996,20(6):325-328.YU Qi-lian,CHEN Zhuo,ZHOU Xin-ling,et al.Aintelligent detective instrument for perio circulation function[J].Chinese Journal of Medical Instrumentation,1996,20(6):325-328.(in Chinese)
[10]  KIM B S,YOO S K.Motionartifact reduction inphotoplethysmography using independent componentanalysis[J].IEEE Transactions on BiomedicalEngineering,2006,53(3):566-568.
[11]  王燕,张松,杨益民,等.基于容积血流脉搏波的心血管血流参数监护模块的研制[J].北京生物医学工程,2006,25(2):148-150.WANG Yan,ZHANG Song,YANG Yi-min,et al.Development of a newly monitoring module forcardiovascular hemodynamic parameters based on theblood volume pulse[J].Beijing BiomedicalEngineering,2006,25(2):148-150.(in Chinese)