工程力学  2012 

假体植入偏差对TKA后膝关节面间力学特性的影响

DOI: 10.6052/j.issn.1000-4750.2012.07.S043, PP. 226-232

Keywords: 全膝关节置换,植入偏差,胫骨假体衬垫,力学特性,有限元

Full-Text   Cite this paper   Add to My Lib

Abstract:

该文基于含假体膝关节的实体模型,构建包含股骨假体植入偏差的直立位TKA后膝关节模型,通过有限元仿真,分析了TKA术的假体植入偏差:股骨假体的内/外移偏差、前/后移偏差、内/外旋偏差及内/外翻偏差对含假体关节面间的力学特性的影响。分析结果表明:股骨假体相对胫骨的内翻偏差对胫骨假体衬垫的最大等效应力和接触应力分布的影响最显著。因此,在全膝关节置换手术中,应尽量减小股骨假体与胫骨间的内/外翻偏差,以控制假体关节面间的应力变化及应力分布变化,来降低胫骨假体衬垫的磨损,该文的分析结果能够为膝关节假体磨损及抗磨损膝假体形面设计提供参考。

References

[1]  Simpson D J, Gray H, Lima D D, et al. The effect of bearing congruency, thickness and alignment on the stresses in unicompartmental knee replacements [J]. Clinical Biomechanics, 2008, 23: 1148―1157.
[2]  郑红光, 黄荣瑛, 郑海东. 全膝关节置换中韧带的两种简化方式对胫股关节力学特性的影响[C]. 第20届全国结构工程学术会议论文集. 北京: 《工程力学》杂志社, 2011: 304―307.
[3]  Zheng Hongguang, Huang Rongying, Zheng Haidong. The effect of ligaments in two simplified ways on the biomechanics of tibiofemoral joint in TKA [C]. Proceedings of the Twentieth National Conference on Structural Engineering. Beijing: Engineering Mechanics Press, 2011: 304―307. (in Chinese)
[4]  Godest A C, Beaugonin M, Haug E, et al. Simulation of a knee joint replacement during gait cycle using explicit finite element analysis [J]. Journal of Biomechanics, 2002, 35: 267―275.
[5]  Matsuda Shuichi, Stephen E W, Victor G, et al. Contact stress analysis in meniscal bearing total knee arthroplasty [J]. The Journal of Arthroplasty, 1998, 6(13): 699―706.
[6]  刘明. 全膝关节置换术后关节线变化对髌骨关节生物力学影响的有限元分析[D]. 北京: 军医进修学院, 2009.
[7]  Liu Ming. Impact of malposition of joint line on biomechanics of patello-femoral joint after total knee replacement: a finite element analysis [D]. Beijing: PLA Postgraduate Medical School, 2009. (in Chinese)
[8]  Feng E L, Stulberg S D, Wixson R I. Progressive subluxation and polyethylene wear in total knee replacements with flat articular surfaces [J]. Clinical Orthopaedics, 1994, 299: 60―71.
[9]  Collier J P, Mayor M B, McNamara J L, Suprenant V A, Jensen R E. Analysis of the failure of 122 polyethylene inserts from uncemented tibial knee components [J]. Clinical Orthopaedics, 1991, 273: 232―242.
[10]  Cheng Cheng-Kung, Huang Chang-Hung, Liau Jiann-Jong, Huang Chun-Hsiung. The influence of surgical malalignment on the contact pressures of fixed and mobile bearing knee prostheses-a biomechanical study [J]. Clinical Biomechanics, 2003, 18(3): 231―236.
[11]  Liau Jiann-Jong, Cheng Cheng-Kung, Huang Chun-Hsiung, et al. The influence of contact alignment of the tibiofemoral joint of the prostheses in in vitro biomechanical testing [J]. Clinical Biomechanics, 1999, 14: 717―721.
[12]  Sawatari T, Tsumura H, Furushiro Y, Torisu T. Three-dimensional finite element analysis of unicompartmental knee arthroplasty-the influence of tibial component inclination [J]. Journal of Orthopaedic Research, 2005, 23(3): 549―554.
[13]  Simpson D J, Price A J, Gulati A, et al. Elevated proximal tibial strains following unicompartmental knee replacement-a possible cause of pain [J]. Medical Engineering & Physics, 2009, 31: 752―757.