Implementation and Validation of a Stride Length Estimation Algorithm, Using a Single Basic Inertial Sensor on Healthy Subjects and Patients Suffering from Parkinson’s Disease

DOI: 10.4236/health.2015.76084, PP. 704-714

Keywords: Inertial Measurement Unit (IMU), Stride Length, Gait Analysis, Parkinson’s Disease, Biologging

Full-Text   Cite this paper   Add to My Lib

Abstract:

As low cost and highly portable sensors, inertial measurements units (IMU) have become increas-ingly used in gait analysis, embodying an efficient alternative to motion capture systems. Mean-while, being able to compute reliably accurate spatial gait parameters using few sensors remains a relatively complex problematic. Providing a clinical oriented solution, our study presents a gy-rometer and accelerometer based algorithm for stride length estimation. Compared to most of the numerous existing works where only an averaged stride length is computed from several IMU, or where the use of the magnetometer is incompatible with everyday use, our challenge here has been to extract each individual stride length in an easy-to-use algorithm requiring only one inertial sensor attached to the subject shank. Our results were validated on healthy subjects and patients suffering from Parkinson’s disease (PD). Estimated stride lengths were compared to GAITRite? walkway system data: the mean error over all the strides was less than 6% for healthy group and 10.3% for PD group. This method provides a reliable portable solution for monitoring the in-stantaneous stride length and opens the way to promising applications.

References

[1]  Chelius, G., Braillon, C., Pasquier, M., Horvais, N., Gibollet, R.P., Espiau, B. and Coste, C.A. (2011) A Wearable Sensor Network for Gait Analysis: A Six-Day Experiment of Running through the Desert. IEEE/ASME Transactions on Mechatronics, 16, 878-883.
http://dx.doi.org/10.1109/TMECH.2011.2161324
[2]  Azevedo Coste, C., Jovic, J., Pissard-Gibollet, R. and Froger, J. (2014) Continuous Gait Cycle Index Estimation for Electrical Stimulation Assisted Foot Drop Correction. Journal of NeuroEngineering and Rehabilitation, 11.
http://dx.doi.org/10.1186/1743-0003-11-118
[3]  Maetzler, W., Domingos, J., Srulijes, K., Ferreira, J.J. and Bloem, B.R. (2013) Quantitative Wearable Sensors for Objective Assessment of Parkinson’s Disease. Movement Disorders: Official Journal of the Movement Disorder Society, 28, 1628-1637.
http://www.ncbi.nlm.nih.gov/pubmed/24030855
[4]  Azevedo Coste, C., Sijobert, B., Pissard-Gibollet, R., Pasquier, M., Espiau, B. and Geny, C. (2014) Detection of Freezing of Gait in Parkinson Disease: Preliminary Results. Sensors, 14, 6819-6827.
http://dx.doi.org/10.3390/s140406819
[5]  Djuric Jovicic, M., Jovicic, N., Radovanovic, S., Stankovic, I., Popovic, M. and Kostic, V. (2013) Automatic Identification and Classification of Freezing of Gait Episodes in Parkinson’s Disease Patients. IEEE Transactions on Neural Systems and Rehabilitation Engineering: A Publication of the IEEE Engineering in Medicine and Biology Society.
http://www.ncbi.nlm.nih.gov/pubmed/24235277
[6]  Cancela, J., Pastorino, M., Arredondo, M.T., Nikita, K.S., Villagra, F. and Pastor, M.A. (2013) Feasibility Study of a Wearable System Based on a Wireless Body Area Network for Gait Assessment in Parkinson’s Disease Patients. Sensors (Basel, Switzerland), 14, 4618-4633.
http://www.pubmedcentral.nih.gov/articlerender.fcgi?artid=4003960
http://dx.doi.org/10.3390/s140304618
[7]  Murodela Herran, A., Garcia-Zapirain, B. and Mendez-Zorrilla, A. (2014) Gait Analysis Methods: An Overview of Wearable and Non-Wearable Systems, Highlighting Clinical Applications. Sensors, 3362-3394.
http://dx.doi.org/10.3390/s140203362
[8]  Yang, S. and Li, Q. (2012) Inertial Sensor-Based Methods in Walking Speed Estimation: A Systematic Review. Sensors, 12, 6102-6116.
http://dx.doi.org/10.3390/s120506102
[9]  Aminian, K., Najafi, B., Leyvraz, P.F. and Robert, P. (2002) Spa-tio-Temporal Parameters of Gait Measured by an Ambulatory System Using Miniature Gyroscopes. Journal of Biomechanics, 35, 689-699.
http://dx.doi.org/10.1016/S0021-9290(02)00008-8
[10]  Trojaniello, D., Cereatti, A., Pelosin, E., Avanzino, L., Mirelman, A., Hausdorff, J.M. and Croce, U.D. (2014) Estimation of Step-by-Step Spatio-Temporal Parameters of Normal and Impaired Gait Using Shank-Mounted Magneto-Inertial Sensors: Application to Elderly, Hemiparetic, Parkinsonian and Choreicgait. Journal of Neuroengineering and Rehabilitation, 1-12.
[11]  Rampp, A., Barth, J., Schulein, S., Gasmann, K., Klucken, J. and Eskofier, B. (2015) Inertial Sensor-Based Stride Parameter Calculation From Gait Sequences in Geriatric Patients. IEEE Transactions on Biomedical Engineering, 62, 1089-1097.
http://dx.doi.org/10.1109/TBME.2014.2368211
[12]  Kose, A., Cereatti, A. and Croce, U.D. (2012) Bilateral Step Length Estimation Using a Single Inertial Measurement Unit Attached to the Pelvis. Journal of NeuroEngineering and Rehabilitation, 9, 9.
http://dx.doi.org/10.1186/1743-0003-9-9
[13]  Mariani, B., Hoskovec, C., Rochat, S., Büla, C., Penders, J. and Aminian, K. (2010) 3D Gait Assessment in Young and Elderly Subjects Using Foot-Worn Inertial Sensors. Journal of Biomechanics, 43, 2999-3006.
http://www.sciencedirect.com/science/article/pii/S0021929010003684
http://dx.doi.org/10.1016/j.jbiomech.2010.07.003
[14]  Webster, K.E., Wittwer, J.E. and Feller, J.A. (2005) Validity of the Gait Rite Walkway System for the Measurement of Averaged and Individual Step Parameters of Gait. Gait and Posture, 22, 317-321.
http://www.sciencedirect.com/science/article/pii/S0966636204002061
http://dx.doi.org/10.1016/j.gaitpost.2004.10.005
[15]  Nelson, A.J., Zwick, D., Brody, S., Doran, C., Pulver, L., Rooz, G., Sadownick, M., Nelson, R. and Rothman, J. (2002) The Validity of the Gait Rite and the Functional Ambulation Performance Scoring System in the Analysis of Parkinson Gait. Neuro Rehabilitation, 17, 255-262.
http://iospress.metapress.com/content/KVTKPTU0PPWFWU8A
[16]  Frosio, I., Pedersini, F., Borghese, N.A. and Model, A.S. (2009) Auto Calibration of MEMS Accelerometers. IEEE Transactions on Instrumentation and Measurement, 58, 2034-2041.
http://dx.doi.org/10.1109/TIM.2008.2006137
[17]  Li, Q., Young, M., Naing, V. and Donelan, J.M. (2010) Walking Speed Estimation Using a Shank-Mounted Inertial Measurement Unit. Journal of Biomechanics, 43, 1640-1643.
http://dx.doi.org/10.1016/j.jbiomech.2010.01.031
[18]  Laudanski, A., Yang, S. and Li, Q. (2011) A Concurrent Comparison of Inertia Sensor-Based Walking Speed Estimation Methods. Proceedings of the 2011 Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Boston, 30 August-3 September 2011, 3484-3487.
http://dx.doi.org/10.1109/IEMBS.2011.6090941
[19]  Moore, S.T., MacDougall, H.G., Gracies, J.M., Cohen, H.S. and Ondo, W.G. (2007) Long-Term Monitoring of Gait in Parkinson’s Disease. Gait and Posture, 26, 200-207.
http://www.ncbi.nlm.nih.gov/pubmed/17046261
http://dx.doi.org/10.1016/j.gaitpost.2006.09.011
[20]  Martinand, P. and Salau¨n, E. (2010) Design and Implementation of a Low-Cost Observer-Based Attitude and Heading Reference System. Control Engineering Practice, 18, 712-722.
http://dx.doi.org/10.1016/j.conengprac.2010.01.012
[21]  Madgwick, S.O.H., Harrison, A.J.L. and Vaidyanathan, R. (2011) Estimation of IMU and MARG Orientation Using a Gradient Descent Algorithm. Proceedings of the 2011 IEEE International Conference on Rehabilitation Robotics, Zurich, 29 June-1 July 2011, 1-7.
http://dx.doi.org/10.1109/ICORR.2011.5975346
[22]  Sijobert, B., Gibollet, R.P., Azevedo, C. and Sensbio, T.K. (2015) An Open-Source Generic Toolkit for Inertial Sensors Based Motion Analysis.
https://github.com/sensbio/sensbiotk
[23]  Sijobert, B., Azevedo, C., Denys, J. and Geny, C. (2014) IMU Based Detection of Freezing of Gait and Festination in Parkinson’s Disease. Proceedings of the 19th International Functional Electrical Stimulation Society (IFESS) Annual Conference, Kuala Lumpur, 17-19 September 2014, 1-3.
http://dx.doi.org/10.1109/ifess.2014.7036751

Full-Text

Contact Us

service@oalib.com

QQ:3279437679

WhatsApp +8615387084133