Purpose: Respiratory-gated radiation therapy (RT) using the real-time
tumor-tracking radiotherapy (RTRT) system is an effective technique for
managing tumor motion. High dosimetric and geometric accuracy is needed;
however, quality assurance (QA) for respiratory-gated RT using the RTRT system has not been reported. The purpose of
this study was to perform QA for respiratorygated RT using the RTRT
system. Materials and Methods: The RTRT system detected the position of the fiducial
marker and radiation delivery gated to the motion of the marker was performed.
The dynamic anthropomorphic thorax phantom was positioned at the isocenter
using the fiducial marker in the phantom. The phantom was irradiated only when
the fiducial marker was within a three-dimensional gating window of ±2 mm from
the planned position. First, the absolute doses were measured using anionization
chamber inserted in the phantom under the stationary, gating and non-gating
state for sinusoidal (nadir-to-peak amplitude [A]: 20 - 40 mm, breathing period
[T]: 2 - 4 s) and the basic respiratory patterns. Second, the dose profiles
were measured using Gafchromic films in the phantom under the same conditions. Differences
between dose profiles were calculated to evaluate the dosimetric and geometric
accuracy. Finally, differences between the actual and measured position of the
fiducial marker were calculated to evaluate the tracking accuracy for
sinusoidal and basic respiratory patterns. Results: For the sinusoidal patterns,
the relative doses were 0.93 for non-gating and 0.99 for gating (A = 20 mm, T =
2 s), 0.94 for non-gating and 1.00 for gating (A = 20 mm, T = 4 s), 0.55 for
non-gating and 1.00 for gating (A = 40 mm, T = 4 s), respectively. For the basic
respiratory pattern, the relative doses were 1.00 for non-gating and 1.00 for
gating, respectively. Compared to the stationary conditions, the differences in
lateral distance between the 90% dose of dose profiles were 6.23 mm for
non-gating and 0.36 mm for gating (A = 20 mm, T = 2 s), 8.79 mm for non-gating
and 1.73 mm for gating (A = 20 mm, T = 4 s), 18.37 mm for non-gating and 0.67
mm for gating (A = 40 mm, T = 4 s), respectively. For the basic respiratory
pattern, those were 5.23 mm for non-gating and 0.35 mm for gating. The root
mean square (RMS) values of the tracking error were
References
[1]
Mah, D., Hanley, J., Rosenzweig, K.E., Yorke, E., Braban, L., Ling, C.C., Leibel, S.A. and Mageras, G. (2000) Technical Aspects of the Deep Inspiration Breath-Hold Technique in the Treatment of Thoracic Cancer. International Journal of Radiation Oncology, Biology, Physics, 48, 1175-1185. http://dx.doi.org/10.1016/S0360-3016(00)00747-1
[2]
Rosenzweig, K.E., Hanley, J., Mah, D., Mageras, G., Hunt, M., Toner, S., Burman, C., Ling, C.C., Mychalczak, B., Fuks, Z. and Leibel, S.A. (2000) The Deep Inspiration Breath-Hold Technique in the Treatment of Inoperable Non-Small-Cell Lung Cancer. International Journal of Radiation Oncology, Biology, Physics, 48, 81-87. http://dx.doi.org/10.1016/S0360-3016(00)00583-6
[3]
Berson, A.M., Emery, R., Rodriguez, L., Richards, G.M., Ng, T., Sanghavi, S. and Barsa, J. (2004) Clinical Experience Using Respiratory Gated Radiation Therapy: Comparison of Free-Breathing and Breath-Hold Techniques. International Journal of Radiation Oncology, Biology, Physics, 60, 419-426. http://dx.doi.org/10.1016/j.ijrobp.2004.03.037
[4]
Ford, E.C., Mageras, G.S., Yorke, E., Rosenzweig, K.E., Wagman, R. and Ling, C.C. (2002) Evaluation of Respiratory Movement during Gated Radiotherapy Using Film and Electronic Portal Imaging. International Journal of Radiation Oncology, Biology, Physics, 52, 522-531. http://dx.doi.org/10.1016/S0360-3016(01)02681-5
[5]
Ramsey, C.R., Cordrey, I.L. and Oliver, A.L. (1999) A Comparison of Beam Characteristics for Gated and Nongated Clinical x-Ray Beams. Medical Physics, 26, 2086-2091. http://dx.doi.org/10.1118/1.598723
[6]
Shirato, H., Shimizu, S., Kunieda, T., Kitamura, K., van Herk, M., Kagei, K., Nishioka, T., Hashimoto, S., Fujita, K., Aoyama, H., Tsuchiya, K., Kudo, K. and Miyasaka, K. (2000) Physical Aspects of a Real-Time Tumor-Tracking System for Gated Radiotherapy. International Journal of Radiation Oncology, Biology, Physics, 48, 1187-1195. http://dx.doi.org/10.1016/S0360-3016(00)00748-3
[7]
Saw, C.B., Brandner, E., Selvaraj, R., Chen, H., Saiful Huq, M. and Heron, D.E. (2007) A Review on the Clinical Implementation of Respiratory-Gated Radiation Therapy. Biomedical Imaging and Intervention Journal, 3, e40. http://dx.doi.org/10.2349/biij.3.1.e40
[8]
Cardenas, A., Fontenot, J., Forster, K.M., Stevens, C.W. and Starkschall, G. (2004) Quality Assurance Evaluation of Delivery of Respiratory-Gated Treatments. Journal Applied Clinical Medical Physics, 5, 55-61. http://dx.doi.org/10.1120/jacmp.v5i3.2002
[9]
Klein, E.E., Hanley, J., Bayouth, J., Yin, F.F., Simon, W., Dresser, S., Serago, C., Aguirre, F., Ma, L., Arjomandy, B., Liu, C., Sandin, C. and Holmes, T. (2009) Task Group 142 Report: Quality Assurance of Medical Accelerators. Medical Physics, 36, 4197-4242.
Shi, C., Papanikolaou, N., Yan, Y., Weng, X. and Jiang, H. (2006) Analysis of the Source of Uncertainty for EDR2 Film-Based IMRT Quality Assurance. Journal Applied Clinical Medical Physics, 7, 1-8. http://dx.doi.org/10.1120/jacmp.v7i2.2230
[12]
Li, X., Yang, Y., Li, T., Fallon, K., Heron, D.E. and Huq, M.S. (2013) Dosimetric Effect of Respiratory Motion on Volumetric-Modulated Arc Therapy-Based Lung SBRT Treatment Delivered by True Beam Machine with Flattening Filter-Free Beam. Journal Applied Clinical Medical Physics, 14,195-204. http://dx.doi.org/10.1120/jacmp.v14i6.4370
[13]
Keall, P., Vedam, S., George, R., Bartee, C., Siebers, J., Lerma, F., Weiss, E. and Chung, T. (2006) The Clinical Implementation of Respiratoy-Gated Intensity-Modulated Radiotherapy. Medical Dosimetry, 31, 152-162. http://dx.doi.org/10.1016/j.meddos.2005.12.002
[14]
ICRU (1987) Use of Computers in External Beam Radiotherapy Procedures with High-Energy Photons and Electrons. International Commission on Radiation Units and Measurement (ICRU), Report 42.
[15]
Cardenas, A., Fontenot, J., Forster, K.M., Stevens, C.W. and Starkschall, G. (2004) Quality Assurance Evaluation of Delivery of Respiratory Gated Treatments. Journal Applied Clinical Medical Physics, 5, 55-61. http://dx.doi.org/10.1120/jacmp.v5i3.2002
[16]
Jiang, S.B., Wolfgang, J., Forster, K.M. and Mageras, G.S. (2008) Quality Assurance Challenges for Motion-Adaptive Radiation Therapy: Gating, Breathing Holding, and Four-Dimensional Computed Tomography. International Journal of Radiation Oncology, Biology, Physics, 71, S103-S107. http://dx.doi.org/ 0.1016/j.ijrobp.2007.07.2386
[17]
Giraud, P. and Houle, A. (2013) Respiratory Gating for Radiotherapy: Main Technical Aspects and Clinical Benefits. International Scholarly Research Notices, 2013, 1-13. http://dx.doi.org/10.1155/2013/519602
[18]
Keall, P.J., Mageras, G.S., Balter, J.M., Emery, R.S., Forster, K.M., Jiang, S.B., Kapatoes, J.M., Low, D.A., Murphy, M.J., Murray, B.R., Ramsey, C.R., Van Herk, M.B., Vedam, S.S., Wong, J.W. and Yorke, E.(2006) Task Group 76 Report: The Management of Respiratory Motion in Radiation Oncology. Medical Physics, 33, 3874-3900.
[19]
Jin, J.Y. and Yin, F.F. (2005) Time Delay Measurement for Linac Based Treatment Delivery in Synchronized Respiratory Gating Radiotherapy. Medical Physics, 32, 1293-1296. http://dx.doi.org/10.1118/1.1896452
[20]
Smith, W.L. and Becker, N. (2009) Time Delays and Margins in Gated Radiotherapy. Journal Applied Clinical Medical Physics, 10, 140-154. http://dx.doi.org/10.1120/jacmp.v10i3.2896
[21]
Shirato, H., Shimizu, S., Kitamura, K., Nishioka, T., Kagei, K., Hashimoto, S., Aoyama, H., Kunieda, T., Shinohara, N., Dosaka-Akita, H. and Miyasaka, K. (2000) Four-Dimensional Treatment Planning and Fluoroscopic Real-Time Tumor Tracking Radiotherapy for Moving Tumor. International Journal of Radiation Oncology, Biology, Physics, 48, 432-442. http://dx.doi.org/10.1016/S0360-3016(00)00625-8.