In the current study, a non-invasive technique for systolic blood pressure (SBP) measurement based on the detection of photoplethysmographic (PPG) pulses during pressure-cuff deflation was compared to sphygmomanometry—the Korotkoff sounds technique. The PPG pulses disappear for cuff-pressures above the SBP value and reappear when the cuff-pressure decreases below the SBP value. One hundred and twenty examinations were performed on forty subjects. In 97 examinations the two methods differed by less than 3 mmHg. In nine examinations the SBP value measured by PPG was higher than that measured by sphygmomanometry by 5 mmHg or more. In only one examination the former was lower by 5 mmHg or more than the latter. The appearance of either the PPG pulses or the Korotkoff sounds assures that the artery under the cuff is open during systolic peak pressure. In the nine examinations mentioned above the PPG pulses were observed while Korotkoff sounds were not detected, despite the open artery during systole. In these examinations, the PPG-based technique was more reliable than sphygmomanometry. The high signal-to-noise ratio of measured PPG pulses indicates that automatic measurement of the SBP by means of automatic detection of the PPG signals is feasible.
References
[1]
O'Brien, E.; Petriel, J.; Littler, W.; de Swiet, M.; Padfield, P.L.; Altman, D.G.; Brand, M.; Coats, A.; Atkins, N. The British Hypertension Society protocol for the evaluation of blood pressure measuring devices. J. Hypertens. 1993, 11, S43–S62.
[2]
Pickering, T.G.; Blank, S.G. Blood Pressure Measurement and Ambulatory Blood Pressure Monitoring. In Hypertension: Pathophysiology, Diagnosis and Management; Laragh, J.H., Brenner, B.M., Eds.; Raven Press: New York, NY, USA, 1990; p. p. 89.
[3]
Pickering, T.G.; Hall, J.E.; Appel, L.J.; Falkner, B.E.; Graves, J.; Hill, M.N.; Jones, D.W.; Kurtz, T.; Sheps, S.G.; Roccella, E.J. Recommendations for blood pressure measurement in humans and experimental animals, Part1: Blood pressure measurement in humans: A statement for professionals from the subcommittee of professional and public education of the American Heart Association council of high blood pressure research. Hypertension 2005, 45, 142–161.
Ng, K.G.; Small, C.F. Survey of automated noninvasive blood pressure monitors. J. Clin. Eng. 1994, 19, 452–475.
[6]
Drzewiecki, G. Noninvasive Assessment of Arterial Blood Pressure and Mechanics. In Biomedical Engineering Handbook; Bronzino, J.D., Ed.; CRC Press: Boca Raton, FL, USA, 1995; p. p. 73.
[7]
Liu, J.; Hahn, J.-O.; Mukkamala, R. Error mechanisms of the oscillometric fixed-ratio blood pressure measurement method. Ann. Biomed. Eng. 2013, 41, 587–597.
[8]
Kaplan, N.M.; Lieberman, E.; Neal, W. Kaplan's Clinical Hypertension, 8th ed. ed.; Lippincott Williams & Wilkins: Philadelphia, PA, USA, 2002; p. p. 2.
[9]
Ursino, M.; Cristalli, C. A mathematical study of some biomechanical factors affecting the oscillometric blood pressure measurement. IEEE Trans. Biomed. Eng. 1996, 43, 761–778.
[10]
White, W.B.; Berson, A.S.; Robins, C.; Jamieson, M.J.; Prisant, L.M.; Rocella, E.; Sheps, S.G. National standard for measurement of resting and ambulatory blood pressures with automated sphygmomanometers. Hypertension 1993, 21, 504–509.
[11]
O'Brien, E.; Waeber, B.; Parati, G.; Staessen, J.; Myers, M.G. Blood pressure measuring devices: Recommendations of the European society of Hypertension. Br. Med. J. 2001, 322, 531–536.
[12]
Asmar, R.; Zanchetti, A. Guidelines for the use of self blood pressure monitoring: A summary report of the first international consensus conference. J. Hypertens. 2000, 18, 493–508.
[13]
Stergiou, G.S.; Siontis, K.C.M.; Ioannidis, J.P.A. Home blood pressure as a cardiovascular outcome predictor. It's time to take this method seriously. Hypertension 2010, 55, 1301–1303.
[14]
Stergiou, G.S.; Parati, G. How to best assess blood pressure? The ongoing debate on the clinical value of blood pressure average and variability. Hypertension 2011, 57, 1041–1042.
[15]
Schillaci, G.; Pucci, G.; Parati, G. Blood pressure variability. An additional target for antihypertensive treatment? Hypertension 2011, 58, 133–135.
[16]
Whyte, R.K.; Elseed, A.M.; Fraser, C.B.; Shinebourne, E.A.; de Swiet, M. Assessment of Doppler ultrasound to measure systolic and diastolic blood pressures in infants and young children. Arch. Dis. Child. 1975, 50, 542–544.
[17]
Stegall, H.F.; Kardon, M.B.; Kemmerer, W.T. Indirect measurement of arterial blood pressure by Doppler ultrasonic sphygmomanometry. J. Appl. Physiol. 1968, 25, 793–798.
[18]
Sykes, D.A.; McCarty, K.; Mullkerin, E.; Fisher, D.J.; Woodcock, J.P. Correlation between Korotkoff's sounds and ultrasonics of the brachial artery in healthy and normotensive subjects. Clin. Phys. Physiol. Meas. 1991, 12, 327–331.
[19]
Talke, P.O. Measurement of systolic blood pressure using pulse oximetry during helicopter flight. Crit. Care Med. 1991, 19, 934–937.
[20]
Langbaum, M.; Eyal, F.G. A practical and reliable method of measuring blood pressure in the neonate by pulse oximetry. J. Pediatr. 1994, 125, 591–595.
[21]
Fronek, A.; Blazek, V.; Curran, B. Toe pressure determination by audiophotoplethysmography. J. Vasc. Surg. 1994, 20, 267–270.
[22]
Laurent, C.; Jonsson, B.; Vegfors, M.; Lindberg, L.G. Non-invasive measurement of systolic blood pressure on the arm utilizing photoplethysmography: Development of the methodology. Med. Biol. Eng. Comput. 2005, 43, 131–135.
[23]
Verrij, E.; van Montfrans, G.; Bos, J.-W. Reintroduction of Riva-Rocci measurements to determine systolic blood pressure? Neth. J. Med. 2008, 66, 480–482.
[24]
Nitzan, M.; Patron, A.; Glik, Z.; Weiss, A.T. Automatic noninvasive measurement of systolic blood pressure using photoplethysmography. Bio. Med. Eng. OnLine 2009, 8, doi:10.1186/1475-925X-8-28.
[25]
Siennicki-Lantz, A.; Elmst?hl, S. Phenomenon of declining blood pressure in elderly-high systolic levels are undervalued with Korotkoff method. BMC Geriatr. 2011, 11, doi:10.1186/1471-2318-11-57.
[26]
Nitzan, M.; Babchenko, A.; Khanokh, B.; Landau, D. The variability of the photoplethysmographic signal—a potential method for the evaluation of the autonomic nervous system. Physiol. Meas. 1998, 19, 93–102.
[27]
Nitzan, M.; Rosenfeld, C.; Weiss, A.T.; Grossman, E.; Patron, A.; Murray, A. Effects of external pressure on arteries distal to the cuff during sphygmomanometry. IEEE Trans. Biomed. Eng. 2005, 52, 1120–1127.
[28]
Spence, J.D. Pseudohypertension. In Hypertension: Pathophysiology, Diagnosis and Management; Laragh, J.H., Brenner, B.M., Eds.; Raven Press: New York, NY, USA, 1990; p. p. 87.
[29]
Chungchareon, D. Genesis of Korotkoff sounds. Am. J. Physiol. 1964, 207, 190–194.
[30]
Ur, A.; Gordon, M. Origin of Korotkoff sounds. Am. J. Physiol. 1970, 218, 524–529.
[31]
Cunningham, T. Korotkoff sounds. Stud. BMJ 2003, 11, 219–262.
[32]
Hunyor, S.N.; Flynn, J.M.; Cochineas, C. Comparison of performance of various sphygmomanometers with intra-arterial blood-pressure readings. Br. Med. J. 1978, 2, 159–162.
[33]
Avolio, A.P.; Butlin, M.; Walsh, A. Arterial blood pressure measurement and pulse wave—their role in enhancing cardiovascular assessment. Physiol. Meas. 2010, 31, R1–R47.
[34]
Ghione, S.; Franchi, D.; Ripoli, A.; Palagi, G.; Mezzasalma, L.; Nardi, G.C.; Belardinelli, A.; Berti, S.; Bedini, R. On the contribution of biomedical engineering and technology to the understanding and the management of arterial hypertension. J. Med. Eng. Technol. 1998, 22, 31–36.