Sensors  2013 

Design and Evaluation of a Low-Cost Smartphone Pulse Oximeter

DOI: 10.3390/s131216882

Keywords: pulse oximetry, mHealth, telemonitoring, pneumonia diagnostics

Full-Text   Cite this paper   Add to My Lib

Abstract:

Infectious diseases such as pneumonia take the lives of millions of children in low- and middle-income countries every year. Many of these deaths could be prevented with the availability of robust and low-cost diagnostic tools using integrated sensor technology. Pulse oximetry in particular, offers a unique non-invasive and specific test for an increase in the severity of many infectious diseases such as pneumonia. If pulse oximetry could be delivered on widely available mobile phones, it could become a compelling solution to global health challenges. Many lives could be saved if this technology was disseminated effectively in the affected regions of the world to rescue patients from the fatal consequences of these infectious diseases. We describe the implementation of such an oximeter that interfaces a conventional clinical oximeter finger sensor with a smartphone through the headset jack audio interface, and present a simulator-based systematic verification system to be used for automated validation of the sensor interface on different smartphones and media players. An excellent agreement was found between the simulator and the audio oximeter for both oxygen saturation and heart rate over a wide range of optical transmission levels on 4th and 5th generations of the iPod Touch TM and iPhone TM devices.

References

[1]  Wiens, M.O.; Kumbakumba, E.; Kissoon, N.; Ansermino, J.M.; Ndamira, A.; Larson, C.P. Pediatric sepsis in the developing world: Challenges in defining sepsis and issues in post-discharge mortality. J. Clin. Epidemiol. 2012, 4, 319–325.
[2]  Eichhorn, J.H.; Cooper, J.B.; Cullen, D.J.; Maier, W.R.; Philip, J.H.; Seeman, R.G. Standards for patient monitoring during anesthesia at Harvard Medical School. J. Am. Med. Assoc. 1986, 256, 1017–1020.
[3]  Gibbs, N.; Rodoreda, P. Anaesthetic mortality rates in Western Australia 1980–2002. Anaesth. Intensiv. Care 2005, 33, 616–622.
[4]  Funk, L.M.; Weiser, T.G.; Berry, W.R.; Lipsitz, S.R.; Merry, A.F.; Enright, A.C.; Wilson, I.H.; Dziekan, G.; Gawande, A.A. Global operating theatre distribution and pulse oximetry supply: An estimation from reported data. Lancet 2010, 376, 1055–1061.
[5]  Walker, I.A.; Merry, A.F.; Wilson, I.H.; McHugh, G.A.; O'Sullivan, E.; Thoms, G.M.; Nuevo, F.; Whitaker, D.K. GO Project teams. Global oximetry: An international anaesthesia quality improvement project. Anaesthesia 2009, 64, 1051–1060.
[6]  World Health Organization. Patient Safety Pulse Oximetry Project 2013. Available online: http://who.int/patientsafety/safesurgery/pulse_oximetry/en/index.html (accessed on 4 December 2013).
[7]  Wilson, I.H.; Walker, I.A. Lifebox: Give a little, help a lot. BMJ 2012, doi:10.1136/bmj.e313.
[8]  Neuman, M.I.; Monuteaux, M.C.; Scully, K.J.; Bachur, R.G. Prediction of pneumonia in a pediatric emergency department. Pediatrics 2011, 128, 246–253.
[9]  Thangaratinam, S.; Brown, K.; Zamora, J.; Khan, K.S.; Ewer, A.K. Pulse oximetry screening for critical congenital heart defects in asymptomatic newborn babies: A systematic review and meta-analysis. Lancet 2012, 379, 2459–2464.
[10]  Duke, T.; Blaschke, A.J.; Sialis, S.; Bonkowsky, J.L. Hypoxaemia in acute respiratory and non-respiratory illnesses in neonates and children in a developing country. Arch. Dis. Child. 2002, 86, 108–112.
[11]  Subhi, R.; Adamson, M.; Campbell, H.; Weber, M.; Smith, K.; Duke, T. The prevalence of hypoxaemia among ill children in developing countries: A systematic review. Lancet Infect. Dis. 2009, 9, 219–227.
[12]  Madico, G.; Gilman, R.H.; Jabra, A.; Rojas, L.; Hernández, H.; Fukuda, J.; Bern, C.; Steinhoff, M. The role of pulse oximetry. Its use as an indicator of severe respiratory disease in peruvian children living at sea level. Arch. Pediatr. Adolesc. Med. 1995, 149, 1259–1263.
[13]  Rosen, L.M.; Yamamoto, L.G.; Wiebe, R.A. Pulse oximetry to identify a high-risk group of children with wheezing. Am. J. Emerg. Med. 1989, 7, 567–570.
[14]  Fu, L.Y.; Ruthazer, R.; Wilson, I.; Patel, A.; Fox, L.M.; Tuan, T.A.; Jeena, P.; Chisaka, N.; Hassan, M.; Lozano, J.; et al. Brief hospitalization and pulse oximetry for predicting amoxicillin treatment failure in children with severe pneumonia. Pediatrics 2006, 118, e1822–e1830.
[15]  Majumdar, S.R.; Eurich, D.T.; Gamble, J.-M.; Senthilselvan, A.; Marrie, T.J. Oxygen saturations less than 92% are associated with major adverse events in outpatients with pneumonia: A population-based cohort study. Clin. Infect. Dis. Off. Publ. Infect. Dis. Soc. Am. 2011, 52, 325–331.
[16]  Hansen, D.; Gausi, S.C.; Merikebu, M. Anaesthesia in Malawi: Complications and deaths. Trop. Dr. 2000, 30, 146–149.
[17]  Ouro-Bang'na Maman, A.F.; Tomta, K.; Ahouangbévi, S.; Chobli, M. Deaths associated with anaesthesia in Togo, West Africa. Trop. Dr. 2005, 35, 220–222.
[18]  Heywood, A.J.; Wilson, I.H.; Sinclair, J.R. Perioperative mortality in Zambia. Ann. R. Coll. Surg. Engl. 1989, 71, 354–358.
[19]  Karlen, W.; Hudson, J.; Petersen, C.; Anand, R.; Dumont, G.A.; Ansermino, J.M. The Phone Oximeter. Proceedings of the IEEE Engineering in Medicine and Biology Conference, Boston, MA, USA, 30 August–3 September, 2011.
[20]  The United Nations Foundation, and Vodafone Partnership. mHealth for Development: The Opportunity of Mobile Technology for Healthcare in the Developing World. 2009.
[21]  Kelly, T.; Halewood, N.; Blackman, C. The Transformational Use of Information and Communication Technologies in Africa; The Executive Summary and Overview. Available online: http://go.worldbank.org/CXS4GFJDE0 (accessed on 4 December 2013).
[22]  NDP DisplaySearch. Smartphone Quarterly; Santa Clara, CA, USA, 2013.
[23]  Hudson, J.; Nguku, S.M.; Sleiman, J.; Karlen, W.; Dumont, G.A.; Petersen, C.L.; Warriner, C.B.; Ansermino, J.M. Usability testing of a prototype Phone Oximeter with healthcare providers in high- and low-medical resource environments. Anesthesia 2012, 67, 957–967.
[24]  Rusch, T.L.; Sankar, R.; Scharf, J.E. Signal processing methods for pulse oximetry. Comput. Biol. Med. 1996, 26, 143–159.
[25]  Petersen, C.L.; Gan, H.; MacInnis, M.J.; Dumont, G.A.; Ansermino, J.M. Ultra-Low-Cost Clinical Pulse Oximetry. Proceedings of the 35th Annual International Conference of the IEEE EMBS, Osaka, Japan, 3–7 July 2013.
[26]  ISO 80601-2-61 Medical electrical equipment—Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment; Geneva, Switzerland, 2011.
[27]  Petersen, C.L.; Gan, H.; Gorges, M.; Ansermino, J.M.; Dumont, G.A. Systematic Evaluation of Mobile Phone Pulse Oximetry Performance. Proceedings of the Annual Meeting of the American Society of Anesthesiologists, San Francisco, CA, USA, 12–16 October 2013.