Impact of Acute Changes of Left Ventricular Contractility on the Transvalvular Impedance: Validation Study by Pressure-Volume Loop Analysis in Healthy Pigs
Background The real-time and continuous assessment of left ventricular (LV) myocardial contractility through an implanted device is a clinically relevant goal. Transvalvular impedance (TVI) is an impedentiometric signal detected in the right cardiac chambers that changes during stroke volume fluctuations in patients. However, the relationship between TVI signals and LV contractility has not been proven. We investigated whether TVI signals predict changes of LV inotropic state during clinically relevant loading and inotropic conditions in swine normal heart. Methods The assessment of RVTVI signals was performed in anesthetized adult healthy anesthetized pigs (n = 6) instrumented for measurement of aortic and LV pressure, dP/dtmax and LV volumes. Myocardial contractility was assessed with the slope (Ees) of the LV end systolic pressure-volume relationship. Effective arterial elastance (Ea) and stroke work (SW) were determined from the LV pressure-volume loops. Pigs were studied at rest (baseline), after transient mechanical preload reduction and afterload increase, after 10-min of low dose dobutamine infusion (LDDS, 10 ug/kg/min, i.v), and esmolol administration (ESMO, bolus of 500 μg and continuous infusion of 100 μg·kg?1·min?1). Results We detected a significant relationship between ESTVI and dP/dtmax during LDDS and ESMO administration. In addition, the fluctuations of ESTVI were significantly related to changes of the Ees during afterload increase, LDDS and ESMO infusion. Conclusions ESTVI signal detected in right cardiac chamber is significantly affected by acute changes in cardiac mechanical activity and is able to predict acute changes of LV inotropic state in normal heart.
References
[1]
Kass DA, Maughan WL, Guo ZM, Kono A, Sunagawa K, et al. (1987) Comparative influence of load versus inotropic states on indices of ventricular contractility: experimental and theoretical analysis based on pressure–volume relationships Circulation. 76: 1422–1426.
[2]
Kass DA, Yamazaki T, Burkhoff D, Maughan WL, Sagawa K (1987) Determination of left ventricular end-systolic pressure-volume relationships by the conductance (volume) catheter technique. Circulation 76: 1422–1436.
[3]
Burkhoff D, Mirsky I, Suga H (2005) Assessment of systolic and diastolic ventricular properties via pressure-volume analysis: a guide for clinical, translational, and basic researchers. Am J Physiol Heart Circ Physiol 289: H501–H512.
[4]
Post H, d'Agostino C, Lionetti V, Castellari M, Kang EY, et al. (2003) Reduced left ventricular compliance and mechanical efficiency after prolonged inhibition of NO synthesis in conscious dogs. J Physiol 552: 233–239.
[5]
Lin HY, Freed D, Lee TW, Arora RC, Ali A, et al. (2011) Quantitative assessment of cardiac output and left ventricular function by noninvasive phase-contrast and cine MRI: validation study with invasive pressure-volume loop analysis in a swine model. J Magn Reson Imaging 34: 203–210.
[6]
Padeletti L, Pieragnoli P, Ricciardi G, Perrotta L, Perini AP, et al. (2012) Acute hemodynamic effect of left ventricular endocardial pacing in cardiac resynchronization therapy: assessment by pressure-volume loops. Circ Arrhythm Electrophysiol 5: 460–467.
[7]
Delnoy PP, Ottervanger JP, Luttikhuis HO, Vos DH, Elvan A, et al. (2009) Pressure-volume loop analysis during implantation of biventricular pacemaker/cardiac resynchronization therapy device to optimize right and left ventricular pacing sites. Eur Heart J 30: 797–804.
[8]
Padeletti L, Paoletti Perini A, Gronda E (2012) Cardiac resynchronization therapy: the issue of non-response. Heart Fail Rev 17: 97–105.
[9]
Merchant FM, Dec GW, Singh JP (2010) Implantable sensors for heart failure. Circ Arrhythm Electrophysiol 3: 657–667.
[10]
Braunschweig F (2007) Therapeutic and diagnostic role of electrical devices in acute heart failure. Heart Fail Rev 12: 157–166.
[11]
Chirife R (1988) Physiological principles of a new method for rate responsive pacing using the pre-ejection interval. Pacing Clin Electrophysiol 11: 1545–1554.
[12]
Bennett T, Sharma A, Sutton R, Camm AJ, Erickson M, et al. (1992) Development of a rate adaptive pacemaker based on the maximum rate-of-rise of right ventricular pressure (RV dP/dtmax). Pacing Clin Electrophysiol 15: 219–234.
[13]
Pichlmaier AM, Braile D, Ebner E, Greco OT, Hutten H, et al. (1992) Autonomic nervous system controlled closed loop cardiac pacing. Pacing Clin Electrophysiol 15: 1787–1791.
[14]
Clementy J, Kobeissi A, Garrigue S, Jais P, Le Métayer P, et al. (2001) Validation by serial standardized testing of a new rate-responsive pacemaker sensor based on variations in myocardial contractility. Europace 3: 124–131.
[15]
Gasparini G, Curnis A, Gulizia M, Occhetta E, Corrado A, et al. (2005) Rate-responsive pacing regulated by cardiac haemodynamics. Europace 7: 234–241.
[16]
Padeletti L, Porciani MC, Ritter P, Michelucci A, Colella A, et al. (2000) Atrioventricular interval optimization in the right atrial appendage and interatrial septum pacing: a comparison between echo and peak endocardial acceleration measurements. Pacing Clin Electrophysiol 23: 1618–1622.
[17]
Bocchiardo M, Meyer zu Vilsendorf D, Militello C, Lippert M, et al. (2010) Resynchronization therapy optimization by intracardiac impedance. Europace 12: 1589–1595.
[18]
Ritter P, Delnoy PP, Padeletti L, Lunati M, Naegele H, et al. (2012) A randomized pilot study of optimization of cardiac resynchronization therapy in sinus rhythm patients using a peak endocardial acceleration sensor vs. standard methods. Europace 14: 1324–1333.
[19]
Stahl C, Walker T, Straub A, Kettering K, Knubben K, et al. (2009) Assessing acute ventricular volume changes by intracardiac impedance in a chronic heart failure animal model. . Pacing Clin Electrophysiol. 32: 1395–401.
[20]
Bocchiardo M, Meyer zu Vilsendorf D, Militello C, Lippert M, Czygan G, et al. (2010) Intracardiac impedance monitors stroke volume in resynchronization therapy patients. . Europace. 12: 702–707.
[21]
Chirife R, Ortega DF, Salazar A (1993) Feasibility of measuring relative right ventricular volumes and ejection fraction with implantable rhythm control devices. Pacing Clin Electrophysiol 16: 1673–1683.
[22]
Schaldach M (1990) Automatic adjustment of pacing parameters based on intracardiac impedance measurements. Pacing Clin Electrophysiol 13: 1702–1710.
[23]
Valzania C, Eriksson MJ, Holmstr?m N, J?rverud K, Gadler F (2009) Multiple vector impedance measurements during biventricular pacing: feasibility and possible implications for hemodynamic monitoring. Pacing Clin Electrophysiol 32: 1492–1500.
[24]
Ginks MR, Sciaraffia E, Karlsson A, Gustafsson J, Hamid S, et al. (2011) Relationship between intracardiac impedance and left ventricular contractility in patients undergoing cardiac resynchronization therapy. Europace 13: 984–991.
[25]
Di Gregorio F, Morra A, Finesso M, Bongiorni MG (1996) Transvalvular impedance (TVI) recording under electrical and pharmacological cardiac stimulation. Pacing Clin Electrophysiol 19: 1689–1693.
[26]
Taborsky M, Kupec J, Vopalka R, Barbetta A, Di Gregorio F (2010) Left-ventricular mechanical activity detected by impedance recording. Europace 12: 534–539.
[27]
Taborsky M, Fedorco M, Skala T, Kocianova E, Pastucha D, et al.. (2013) Acute effects of right ventricular pacing on cardiac haemodynamics and transvalvular impedance. Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. In press.
[28]
Gemignani V, Bianchini E, Faita F, Lionetti V, Campan M, et al. (2010) Transthoracic sensor for noninvasive assessment of left ventricular contractility: validation in a minipig model of chronic heart failure. Pacing Clin Electrophysiol 33: 795–803.
[29]
Cheung MM, Smallhorn JF, Redington AN, Vogel M (2004) The effects of changes in loading conditions and modulation of inotropic state on the myocardial performance index: comparison with conductance catheter measurements. Eur Heart J 25: 2238–2242.
[30]
Borow KM, Neumann A, Marcus RH, Sareli P, Lang RM (1992) Effects of simultaneous alterations in preload and afterload on measurements of left ventricular contractility in patients with dilated cardiomyopathy: comparisons of ejection phase, isovolumetric and end-systolic force-velocity indexes. J Am Coll Cardiol 20: 787–795.
[31]
Nozawa T, Yasumura Y, Futaki S, Tanaka N, Uenishi M, et al. (1988) Efficiency of energy transfer from pressure–volume area to external mechanical work increases with contractile state and decreases with afterload in the left ventricle of the anesthetized closed-chest dog. Circulation 77: 1116–1124.
[32]
Sunagawa K, Maughan WL, Burkhoff D, Sagawa K (1983) Left ventricular interaction with arterial load studied in isolated canine ventricle. Am J Physiol Heart Circ Physiol 245: H773–H780.
[33]
Suga H (1990) Ventricular energetics. Physiol Rev 70: 247–277.
[34]
Teitel DF, Klautz RJ, Cassidy SC, Steendijk P, van der Velde ET, et al.. (1992) The end-systolic pressure-volume relationship in young animals using the conductance technique. Eur Heart J. 13 Suppl E:40–46.
[35]
Lionetti V, Aquaro GD, Simioniuc A, Di Cristofano C, Forini F, et al. (2009) Severe mechanical dyssynchrony causes regional hibernation-like changes in pigs with nonischemic heart failure. J Card Fail 15: 920–928.
[36]
Altman RK, McCarty D, Chen-Tournoux AA, Tournoux FB, Riedl L, et al. (2011) Usefulness of low-dose dobutamine echocardiography to predict response and outcome in patients undergoing cardiac resynchronization therapy. Am J Cardiol 108: 252–257.
[37]
Pingitore A, Aquaro GD, Lorenzoni V, Gallotta M, De Marchi D, et al. (2013) Influence of preload and afterload on stroke volume response to low-dose dobutamine stress in patients with non-ischemic heart failure: A cardiac MR study. Int J Cardiol 166: 475–481.
[38]
Cannesson M, Jacques D, Pinsky MR, Gorcsan J 3rd (2006) Effects of modulation of left ventricular contractile state and loading conditions on tissue Doppler myocardial performance index. Am J Physiol Heart Circ Physiol 290: H1952–H1959.
[39]
Osswald S, Cron T, Gr?del C, Hilti P, Lippert M, et al. (2000) Closed-loop stimulation using intracardiac impedance as a sensor principle: correlation of right ventricular dP/dtmax and intracardiac impedance during dobutamine stress test. Pacing Clin Electrophysiol 23: 1502–1508.
[40]
Bombardini T, Gaggini G, Marcelli E, Parlapiano M, Plicchi G (2000) Peak endocardial acceleration reflects heart contractility also in atrial fibrillation. Pacing Clin Electrophysiol 23: 1381–1385.
[41]
Dzwonczyk R, del Rio C, McSweeney TD, Zhang X, Howie MB (2009) Myocardial electrical activity does not affect myocardial electrical impedance measurements. J Clin Monit Comput 23: 217–222.
[42]
Dongaonkar RM, Stewart RH, Geissler HJ, Laine GA (2010) Myocardial microvascular permeability, interstitial oedema, and compromised cardiac function. Cardiovasc Res 87: 331–339.
[43]
Fleischhauer J, Lehman L, Kléber A (1995) Electrical resistances of interstitial and microvascular space as determinants of the extracellular electrical field and velocity of propagation in ventricular myocardium. Circulation 92: 587–59.