Background: Seven
patients at a hospital in Houston,TX, were diagnosed during a two-week period in 2009 with joint space
infection of pansusceptible P. aeruginosa following arthroscopic procedures of the knee or shoulder. Tosh et al. (2011), who investigated and
published the principal report discussing this bacterial outbreak, conclude
that its most likely cause was the improper reprocessing of certain reusable,
physically-complex, heat-stable arthroscopic instrumentsused during these arthroscopic
procedures. These reusable instruments reportedly remained contaminated with
remnant tissue,despite
diligent efforts by the hospital to clean their internal structures.This retained bioburden presumably
shielded the outbreak’s strain of embedded P.
aeruginosa from contact with the pressurized steam, reportedly resulting in
ineffective sterilization of these arthroscopic instruments and bacterial
transmission.Objectives: First, to clarify which specific
sterilization methods, in addition to steam sterilization, Methodist Hospital
employed to process its reusable arthroscopic instrumentation at the time of
its outbreak, in 2009; second, to evaluate Tosh et al.’s (2011) conclusion that ineffective steam sterilization due
to inadequate cleaning was the most likely cause of this hospital’s outbreak;
third, to consider whether any other hitherto unrecognized factors could have
plausibly contributed to this outbreak; and, fourth, to assess whether any
additional recommendations might be warranted to prevent disease transmission
following arthroscopic procedures. Methods: The medical literature was reviewed; some of the principles of quality
assurance, engineering and a root-cause analysis were employed; and Tosh et al.’s (2011) findings and conclusions
were reviewed and compared with those of other published reports that evaluated
the risk of disease transmission associated with the steam sterilization of
physically-complex, heat-stable, soiled surgical instruments. Results and Conclusion: Reports
documenting outbreaks of P. aeruginosa or
another vegetative bacterium associated with the steam sterilization
of inadequatelycleaned
surgical or arthroscopic instruments are scant. This
References
[1]
P. K. Tosh, et al., “Outbreak of Pseudomonas aeruginosa Surgical Site Infections after Arthroscopic Procedures: Texas, 2009,” Infection Control and Hospital Epidemiology, Vol. 32, No. 12, 2011, pp. 179-186.
doi:10.1086/662712
[2]
J. Eaton, “Filthy Surgical Instruments: The Hidden Threat in America’s Operating Rooms. How Dirty Medical Devices Expose Patients to Infection,” The Center for Public Integrity, Washington DC, 2012.
[3]
Food and Drug Administration (FDA), “Ongoing Safety Review of Arthroscopic Shavers,” Safety Communications, 2009.
[4]
Food and Drug Administration (FDA), “Formula 180 Shaver Handpeice. Manufacturer: Stryker Endoscopy,” Manufacturer and User Facility Device Experience Event, 2009.
The NBC Nightly News, “Dirty Surgical Tools Put Patients at Risk,” 2012.
[7]
Fox News, “Report: Dirty Surgical Tools in Hospitals Putting Patients at Risk,” 2012.
[8]
Food and Drug Administration (FDA), “Sterrad NX Sterilizer (Manufacturer: Advanced Sterilization Products),” Manufacturer and User Facility Device Experience (MAUDE) Event, 2010.
[9]
M. Sorin, et al., “Nosocomial Transmission of Imipenem-Resistant Pseudomonas aeruginosa Following Bronchoscopy Associated with Improper Connection to the Steris System 1 Processor,” Infection Control and Hospital Epidemiology, Vol. 22, No. 7, 2001, pp. 409-413.
doi:10.1086/501925
[10]
L. F. Muscarella, “Leading a Horse to Water: Are Crucial Lessons in Endoscopy and Outbreak Investigations Being Learned?” Infection Control and Hospital Epidemiology, Vol. 23, No. 7, 2002, pp. 358-360. doi:10.1086/503469
[11]
P. Davies, “Germ Watch: Clinic Infections Put a Sterilizer of Lab Devices under Microscope,” The Wall Street Journal, 2004.
[12]
C. J. Alvarado, S. M. Stolz and D. G. Maki, “Nosocomial Infections from Contaminated Endoscopes: A Flawed Automated Endoscope Washer. An Investigation Using Molecular Epidemiology,” The American Journal of Medicine, Vol. 91, No. 3, 1991, pp. 272S-280S.
doi:10.1016/0002-9343(91)90381-7
[13]
Centers for Disease Control and Prevention (CDC), “Pseudomonas aeruginosa Infections Associated with Transrectal Ultrasound-Guided Prostate Biopsies—Georgia, 2005,” Morbidity and Mortality Weekly Report, Vol. 55, No. 28, 2006, pp. 776-777.
[14]
L. F. Muscarella, “Medical Errors, Infection-Control Breaches and the Use of Adulterated and Misbranded Medical Devices,” World Journal of Clinical Infectious Diseases, Vol. 2, No., 2, 2012, pp. 13-27.
[15]
M. J. Struelen, F. Rost, A. Deplano, et al., “Pseudomonas aeruginosa and Enterobacteriaceae bacteremia after Biliary Endoscopy: An Outbreak Investigation Using DNA Macrorestriction Analysis,” The American Journal of Medicine, Vol. 95, No. 5, 1993, pp. 489-498.
doi:10.1016/0002-9343(93)90331-I
[16]
H. J. Kolmos, et al., “Pseudo-Outbreak of Pseudomonas aeruginosa in HIV-Infected Patients Undergoing Fiberoptic Bronchoscopy,” Scandinavian Journal of Infectious Diseases, Vol. 26, No. 6, 1994, pp. 653-657.
doi:10.3109/00365549409008632
[17]
L. F. Muscarella, “Contribution of Tap Water and Environmental Surfaces to Nosocomial Transmission of Antibiotic-Resistant Pseudomonas aeruginosa,” Infection Control and Hospital Epidemiology, Vol. 25, No. 4, 2004, pp. 342-345. doi:10.1086/502402
[18]
Centers for Disease Control and Prevention (CDC), “Guideline for Disinfection and Sterilization in Healthcare Facilities,” 2008, pp. 1-158.
[19]
Association of Perioperative Registered Nurses (AORN), “Perioperative Standards and Recommended Practices,” Denver, 2005.
[20]
C. R. Voyles, et al., “Steam Sterilization of Laparoscopic Instruments,” Surgical Laparoscopy Endoscopy, Vol. 5, No. 2, 1995, 139-141.
[21]
W. A. Rutala, M. F. Gergen and D. J. Weber, “Impact of an Oil-Based Lubricant on the Effectiveness of the Sterilization Processes,” Infection Control and Hospital Epidemiology, Vol. 29, No. 1, 2008, pp. 69-72.
doi:10.1086/524326
[22]
F. T. Blevins, et al., “Septic Arthritis Following Arthroscopic Meniscus Repair: A Cluster of three Cases,” Arthroscopy, Vol. 15, No. 1, 1999, pp. 35-40.
doi:10.1053/ar.1999.v15.015003
[23]
S. A. McNeil, et al., “Outbreak of Sternal Surgical Site Infections Due to Pseudomonas aeruginosa Traced to a Scrub Nurse with Onychomycosis,” Clinical Infectious Diseases, Vol. 33, No. 3, 2001, pp. 317-323.
doi:10.1086/321890
[24]
R. L. Moolenaar, et al., “A Prolonged Outbreak of Pseudomonas aeruginosa in a Neonatal Intensive Care Unit: Did Staff Fingernails Play a Role in Disease Transmission?” Infection Control and Hospital Epidemiology, Vol. 21, No. 2, 2000, pp. 80-85. doi:10.1086/501739
[25]
A. F. Widmer, et al., “Outbreak of Pseudomonas aeruginosa Infections in a Surgical Intensive Care Unit: Probable Transmission via Hands of a Health Care Worker,” Clinical Infectious Diseases, Vol. 16, No. 3, 1993, pp. 372-376. doi:10.1093/clind/16.3.372
[26]
X. Bertrand, et al., “Large Outbreak in a Surgical Intensive Care Unit of Colonization or Infection with Pseudomonas aeruginosa That Overexpressed an Active Efflux Pump,” Clinical Infectious Diseases, Vol. 31, No. 4, 2000, pp. E9-E14. doi:10.1086/318117
[27]
A. M. Rogues, et al., “Contribution of Tap Water to Patient Colonisation with Pseudomonas aeruginosa in a Medical Intensive Care Unit,” Journal of Hospital Infection, Vol. 67, No. 1, 2007, pp. 72-78.
doi:10.1016/j.jhin.2007.06.019
[28]
D. Mayank et al., “Nosocomial Cross-Transmission of Pseudomonas aeruginosa between Patients in a Tertiary Intensive Care Unit,” Indian Journal of Pathology and Microbiologyl, Vol. 52, No. 4, 2009, pp. 509-513.
doi:10.4103/0377-4929.56143
[29]
S. Hota, et al., “Outbreak of Multidrug-Resistant Pseudomonas aeruginosa Colonization and Infection Secondary to Imperfect Intensive Care Unit Room Design,” Infection Control and Hospital Epidemiology, Vol. 30, No. 1, 2009, pp. 25-33. doi:10.1086/592700
[30]
C. Lowe, et al., “Outbreak of Extended-Spectrum β-Lactamase-Producing Klebsiella oxytoca Infections Associated with Contaminated Handwashing Sinks,” Emerging Infectious Diseases, Vol. 18, No. 8, 2012, pp. 1242-1247.
doi:10.3201/eid1808.111268
[31]
D. M. Dudzinski, et al., “The Disclosure Dilemma— Large-Scale Adverse Events,” The New England Journal of Medicine, Vol. 363, No. 10, 2010, pp. 978-986.
doi:10.1056/NEJMhle1003134
[32]
Food and Drug Administration, “Establishment Inspection Report: Stryker Endoscopy,” Inspection, San Jose, 2009.
[33]
L. F. Muscarella, “Instrument Design and Cross-Infection,” AORN Journal, Vol. 67, No. 3, 1998, pp. 552-553, 556. doi:10.1016/S0001-2092(06)62824-X
[34]
L. F. Muscarella, “Are All Sterilization Processes Alike?” AORN Journal, Vol. 67, No. 5, 1998, pp. 966-970, 973-976. doi:10.1016/S0001-2092(06)62622-7
[35]
L. F. Muscarella, “The Importance of Bronchoscope Reprocessing Guidelines: Raising the Standard of Care,” Chest, Vol. 126, No. 3, 2004, pp. 1001-2; reply, pp. 1002-1003.
[36]
L. F. Muscarella, “Application of Environmental Sampling to Flexible Endoscope Reprocessing: The Importance of Monitoring the Rinse Water,” Infection Control and Hospital Epidemiology, Vol. 23, No. 5, 2002, pp. 285-289. doi:10.1086/502053