全部 标题 作者
关键词 摘要

OALib Journal期刊
ISSN: 2333-9721
费用:99美元

查看量下载量

相关文章

更多...

Assay Validation for KIM-1: human urinary renal dysfunction biomarker

Full-Text   Cite this paper   Add to My Lib

Abstract:

Urinary kidney injury molecule (KIM-1) is a sensitive quantitative biomarker for early detection of kidney tubular injury. The objective of the present work was to analytically validate this urinary renal injury biomarker. Duo-set reagents from R&D were used to develop the ELISA and validate the assay's linearity, intra-run precision, inter-run precision, lower limit of quantification, recovery, dilutional verification, reference range, stability, and length of run. The reference range data suggests that the healthy population falls within the assay range (59 - 2146 pg/mL) and upper limit of quantitation for this assay is 17168 pg/mL for the patient population. This is a robust assay to detect urinary levels of KIM-1, which serves as a non-invasive sensitive, reproducible, and potentially high-throughput method to detect early kidney injury in drug development studies.

Full-Text

Contact Us

service@oalib.com

QQ:3279437679