Recurrent syncope in a hypertensive subject with vascular cognitive impairment and permanent pacemaker

DOI: 10.4081/ni.2012.e12

Keywords: synsope , neurology , hypertension , pacemaker

Full-Text   Cite this paper   Add to My Lib

Abstract:

Syncope following permanent pacemaker (PM) implantation is a nightmare for electrophysiologists. We describe a case of daily recurrent syncope in an 84-year-old man having a dual-chamber pacemaker implanted for complete atrio-ventricular block occurred 4 years before the admission to our department. He had a history of arterial hypertension, parossistic atrial fibrillation, chronic obstructive pulmonary disease, stage-III chronic renal failure, mild vascular cognitive impairment and glaucoma. The initial work-up including electrocardiogram (ECG), repeated PM inter- rogations, Holter electrocardiogram, blood pressure measurement in orthostatic position, complete blood count, serum glycaemia, electrolytes and thyroid function tests showed normal findings. Syncope occurred in lying position and during 90° left clockwise neck rotation and was associated to pallor, sweating, tonic-clonic seizures and transient self-limited loss of consciousness lasting a few seconds. Electroencephalogram was normal. During continuous ECG monitoring, the right rotation of the head determined a ventricular asystolic pause lasting 9 seconds associated with loss of consciousness. Restoration of sinus rhythm was observed after bringing back the head in axis. The PM interrogation, performed during pacing failure, recorded low impedance of bipolar ventricular lead, suggesting a damage in lead insulation. It is likely that lead movements during clockwise neck rotation produced an intermittent short circuit that prevented sufficient energy delivery to the myocardium with a consequence of sudden loss of capture.

Full-Text

Contact Us

service@oalib.com

QQ:3279437679

WhatsApp +8615387084133