Endotracheal tube cuff pressure assessment maneuver induces drop of expired tidal volume in the postoperative of coronary artery bypass grafting

DOI: 10.1186/1749-8090-7-53

Keywords: Coronary artery bypass grafting, Endotracheal tube cuff pressure, Air leakage, Expired tidal volume, Mechanical ventilation

Full-Text   Cite this paper   Add to My Lib

Abstract:

A total of 488 patients were analyzed. After the operation, the lungs were ventilated in pressure-assist-control mode and the same ventilatory settings were adjusted for all patients. After intensive care unit arrival, the cuff was fully deflated and then progressively inflated by air injection, to promote a minimal volume to occlude the trachea. To assist the cuff inflation and the air leakage identification, the graphical monitoring of the volume-time curve was adopted. After 20 minutes a first cuff pressure evaluation was performed (P1) and a second measurement (P2) was taken after 20 minutes with an analog manometer. ETV was obtained always pre and post P1 measurement.The CP assessment maneuver promoted a significant drop of P2 in relation to P1 when the manometer was attached to the pilot balloon (p?<?0.0001). When compared the moments, pre-P1 versus post-P1, a significant drop of the ETV was also observed (p?<?0.0001).The CP assessment maneuver promoted a significant decrease in CP values and occurrence of air leakage with reduction of ETV in the early postoperative of CABG.