色谱  2000 

Trouble Shooting of Splitter Apparatus of Shimadzu GC-1 4A Gas Chromatograph
GC-14A气相色谱仪分流装置的故障排除

Keywords: gas chromatography,split/splitless sample injector,trouble shooting
故障排除
,GC-14A型,气相色谱仪

Full-Text   Cite this paper   Add to My Lib

Abstract:

This paper describes the abnormal phenomena caused by almost complete block of connective tube of SPL 14 split/splitless sample injector of Shimadzu GC 14A gas chromatograph. It has been pointed out that if oil compounds with high boiling point are analyzed, periodical clean up should be done to remove blocking materials for experimental repeatability and accuracy.