色谱 1997
Determination of Sulpride in Human Plasma by High Performance Liquid Chromatography
|
Abstract:
This paper describes a reliable method for the pharmacokinetic study of Sulpride in human plasma by reversed-phase high performance liquid chromatography. To compensate the loss of Sulpride during the extraction procedure we used an internal standard very similar in chemical structure and UV absorbance to those of Sulpride. The mobile phase was methanol-water-acetic acid (60:30:1) with a flow rate of 1.2 mL/min. A UV detector was used at 290 nm. The linear range was 5-100 mg/L and the detectable limit was 1.0 mg/L. The recovery and RSD were 97.95%-99.96% and 2.6%-5.1% respectively. The results showed that this method is a sensitive and accurate one which makes the pharmacokinetic study of Sulpride possible. If the concentration was too low to be detected by UV monitor, a fluorescence detector could be used with the excitation wavelength at 299 nm and emission at 342 nm. We analyzed the plasma samples from 30 day-treated psychotic patients and got the satisfactory results.