色谱 2006
Determination of Aromatics in Light Petroleum Products by Comprehensive Two-Dimensional Gas Chromatography
|
Abstract:
In recent years, comprehensive two-dimensional gas chromatography (GC x GC) have been used widely, and the applications of this technique to many fields have already been reported. In the standard method of oil analysis, the concentrations of aromatics and naphthalene hydrocarbons in light petroleum products must be detected by more than two methods. Mono-aromatics, di-aromatics etc. in light petroleum products were detected only by comprehensive two-dimensional gas chromatography. After the proper selection of column system and optimization of chromatographic conditions, the method can achieve the group separations of paraffins, olefins, naphthenes, aromatics with 1 to 2 rings and some target components in light petroleum products with good reproducibility and good precision. The recoveries of standard compounds were 89.5% - 106.1%, and the relative standard deviations of repeatedly detecting the components were all lower than 5.8%. It took only 30 min to finish a determination.