Sensors  2008 

Influence of Fluid Cell Design on the Frequency Response of AFM Microcantilevers in Liquid Media

DOI: 10.3390/s8095927

Keywords: AFM, microcantilever, frequency response, fluid cell, liquid

Full-Text   Cite this paper   Add to My Lib

Abstract:

A study of the frequency response of AFM microcantilevers in liquid media contained in a commercial fluid cell is presented. Such systems exhibit complicated dynamics which are often not well described by available theories. Their dynamic behavior has a direct effect on the use of the AFM in dynamic mode while imaging in liquid or while extracting the rheological properties of the fluid. We explore the issues related to the design of the cantilever holder/fluid cell and propose an approach for evaluating, minimizing and recognizing the ultimate limitations of commercial cantilever holders. A technique for estimating the frequency response spectrum of the fluid cell itself from experimental data is presented. This spectrum can then be used to evaluate whether or not the fluid cell is suited for the desired purpose.