冶金分析 2012
Study on the improvement of analysis precision for laser induced breakdown spectrometry
|
Abstract:
The analytical precision of laser induced breakdown spectrometry(LIBS) was influenced by severe signal fluctuation,strong continuous emission interference complex spectrum,laser energy fluctuation and bad repeatability.In this study,based on the self-developed laser spectrometer,the spatial resolution and data stabilization was taken in to account with laser beam adjusting technique;the analytical sensitivity and lower limit of detection was improved with gating integration techniques;the reliability of data was improved by data screening technology.Thus,the analytical precision of LIBS was enhanced.Fifteen elements including C,Si,Mn,P,S,B,Mo,Ni,Cr,V,Cu,Ti,Al,W and Fe in amiddle-low alloy steel were determined by LIBS.Except for P and B,the relative standard deviations of determination results were all less than 10 %.The analytical precision was better than those of LIBS reported in references.