Characterization of the Hamamatsu R11410-10 3-Inch Photomultiplier Tube for Liquid Xenon Dark Matter Direct Detection Experiments

DOI: 10.1016/j.nima.2012.08.052

Full-Text   Cite this paper   Add to My Lib

Abstract:

To satisfy the requirements of the next generation of dark matter detectors based on the dual phase TPC, Hamamatsu, in close collaboration with UCLA, has developed the R11410-10 photomultipler tube. In this work, we present the detailed tests performed on this device. High QE (>30%) accompanied by a low dark count rate (50 Hz at 0.3 PE) and high gain (10^7) with good single PE resolution have been observed. A comprehensive screening measurement campaign is ongoing while the manufacturer quotes a radioactivity of 20 mBq/PMT. These characteristics show the R11410-10 to be particularly suitable for the forthcoming zero background liquid xenon detectors.