垃圾焚烧傅里叶红外气体分析仪的研制及应用
Development and Application of Fourier Transform Infrared Gas Analyzer for Garbage Incineration

DOI: 10.12677/APP.2023.1311052, PP. 500-508

Keywords: 傅里叶红外光谱仪,气体监测,测量
Fourier Transform Infrared Spectrometer
, Gas Monitoring, Measure

Full-Text   Cite this paper   Add to My Lib

Abstract:

为了满足垃圾焚烧污染源多组分气体在线监测的需要,研制了一款基于傅里叶红外光谱分析技术的气体分析仪,本文对分析仪的硬件、软件进行了描述。在实验室测量了该仪器对SO2和NO的重复性和示值误差,结果表明分析仪的重复性优于1%,示值相对误差小于2%,具备良好的准确性;同时,模拟测量了排放现场不同浓度水气干扰下的结果,结果表明SO2和NO的示值绝对偏差低于1 ppm,相对偏差小于2%,具备一定的现场应用能力。
In order to meet the need for online monitoring of multi-component gases from waste incinera-tion pollution sources, a gas analyzer based on Fourier transform infrared spectroscopy analysis technology has been developed. This article describes the hardware and software of the analyzer. The repeatability and indication error of the instrument for SO2 and NO were measured in the laboratory, and the results showed that the repeatability of the analyzer was better than 1%, and the relative error of indication was less than 2%, indicating good accuracy; At the same time, simulation measurements were conducted on the results of different concentrations of water vapor interference at the emission site. The results showed that the absolute deviation of SO2 and NO readings was less than 1 ppm, and the relative deviation was less than 2%, indicating certain on-site application capabilities.

References

[1]  GB 18485-2014生活垃圾焚烧污染控制标准[EB/OL]. https://openstd.samr.gov.cn/bzgk/gb/newGbInfo?hcno=9FB5FEAE9E7FA4678998D8045BFCB075, 2014-05-16.
[2]  HJC-ZY80-2017生活垃圾焚烧固定源烟气排放连续监测系统技术要求及检测方法[EB/OL]. https://www.doc88.com/p-68961203873207.html, 2023-07-28.
[3]  穆永吉, 陈远, 余俊, 等. 一种多次反射怀特池的研究与设计[J]. 光学技术, 2014, 40(4): 345-348. https://doi.org/10.13741/j.cnki.11-1879/o4.2014.04.013