Use of weak ion association in the separation of inorganic anions by capillary electrophoresis with specific application to simultaneous-trace determination of bromate and iodate in drinking waterKeywords: Weak ion association , Selectivity , Capillary electrophoresis , Tetrabutylammonium ion , Bromate , Iodate , Drinking water Abstract: In this work weak ion association was used to effect selectivity and detection of inorganic anions with environmental or health significance by capillary electropheresis, CE. Tetrabutylammonium ion was used as a pairing anion to separate mixtures containing closely or co-migrating inorganic anions at pHs 3.8 and pH 7. Despite taking a longer analysis time, better resolution is achieved at pH 7 than 3.8. Trace level of bromate and iodate present in drinking water were determined after online pre-concentration by field enhanced sample injection (FESI) technique. Consequently an internal standard (SCN-) was employed, which entailed the use of tetrabutylammonium ion as a pairing cation to resolve the internal standard from a co-migrating broad peak. The LODs (S/N = 3) were 7.8 x 10-10 M (10 ppb) and 1.2 x 10-9 M (0.21 ppb) for bromate and iodate, respectively. The method was subsequently used to determine bromate and iodate levels in drinking water.
|