%0 Journal Article %T Multicenter validation of the flow measurement of classical monocyte fraction for chronic myelomonocytic leukemia diagnosis %J - %D 2018 %R https://doi.org/10.1038/s41408-018-0146-8 %X Peripheral blood monocytes include three subsets defined by CD14 and CD16 surface markers. An increase in the CD14++CD16£¿ classical monocyte fraction£¿¡Ý£¿94% of the total monocytes was proposed to rapidly and efficiently distinguish chronic myelomonocytic leukemia from reactive monocytosis. The robustness of this assay required a multicenter validation. The flow cytometry assay designed to quantify peripheral blood monocyte subsets was implemented by multiple diagnosis laboratories in France. A nationwide survey was performed to evaluate its performance. All the 48 French laboratories answered the questionnaire, revealing that 63% use this assay routinely. Central blind reanalysis of 329 cytometry files collected from five laboratories demonstrated an excellent correlation in classical monocyte fraction measurement (r£¿=£¿0.93; p£¿<£¿0.0001). The cutoff value of 94% classical monocytes being the critical readout for diagnosis, we then compared 115 patients with classical monocytes£¿¡Ý£¿94% and 214 patients with a fraction£¿<£¿94% between initial analysis and reanalysis. An agreement was obtained in 311 files. Finally, an overt diagnosis, available for 86 files, confirmed a good sensitivity (93.6%) and specificity (89.7%). This survey demonstrates the robustness of the flow assay with limited variability of classical monocyte percentage between centers, validates the 94% cutoff value, and confirms its sensitivity and specificity %U https://www.nature.com/articles/s41408-018-0146-8