%0 Journal Article %T Correlation between different methods of intraabdominal pressure monitoring in varying intraabdominal hypertension models %A F.M. Sanchez Margallo %A G Castellanos %A L Correa-Martin %A M.L.N.G. Malbrain %A P Becker %A R.D. Wise %A R.N. Rodseth %J - %D 2017 %X Background. Advances in intra-abdominal pressure (IAP) measurement have enabled better monitoring and physiological manipulation of patients with intra-abdominal hypertension or abdominal compartment syndrome. This study aimed to determine the correlation between transvesical (TV), transgastric (TG) and direct transperitoneal (TP) IAP monitoring at different IAPs in porcine models. %K Objectives. To assess the statistical agreement between TV %K TG and TP pressure monitoring in a pneumoperitoneum and an intestinal obstruction intra-abdominal hypertension model at different IAPs. %U https://www.ajol.info/index.php/sajcc/article/view/159673